Ballisticsrdiography (BCG)
Principal measurement range of parameter:
• 0-7mg
• 0-100 um
Signal frequency range (Hz)
• dc-40
• dc-40
Standard sensor of method
• Accelerometer, strain gage
• Displacement linear variable differential transformer (LVDT)
Bladder pressure
Principal measurement range of parameter:
• 1-100 cm H2O
Signal frequency range (Hz)
• dc-10
Standard sensor of method
• Strain gage manometer
Blood flow
Principal measurement range of parameter:
• 1-300 ml/s
Signal frequency range (Hz)
• dc-20
Standard sensor of method
• Flowmeter (electromagnetic or ultrasonic)
Direct arterial blood pressure
Principal measurement range of parameter:
• 10-400mm Hg
Signal frequency range (Hz)
• dc-50
Standard sensor of method
• strain-gage manometer
Indirect arterial blood pressure
Principal measurement range of parameter:
• 25-400 mm Hg
Signal frequency range (Hz)
• dc-60
Standard sensor of method
• cuff, auscultation
Blood pressure, venous
Principal measurement range of parameter:
• 0-50 mm Hg
Signal frequency range (Hz)
• dc-50
Standard sensor of method
• strain gage
Blood gas PO2
Principal measurement range of parameter:
• 30-100 mm Hg
Signal frequency range (Hz)
• dc-2
Standard sensor of method
• specific electrode, volumetric or manometric
Blood gas PCO2
Principal measurement range of parameter:
• 40-100 mm Hg
Signal frequency range (Hz)
• dc-2
Standard sensor of method
• specific electrode, volumetric or manometric
Blood gas PN2
Principal measurement range of parameter:
• 1-3mm Hg
Signal frequency range (Hz)
• dc-2
Standard sensor of method
• specific electrode, volumetric or manometric
Blood gas PCO
Principal measurement range of parameter:
• 0.1-0.4 mm Hg
Signal frequency range (Hz)
• dc-2
Standard sensor of method
• specific electrode, volumetric or manometric
Blood pH
Principal measurement range of parameter:
• 6.8-7.8 pH
Signal frequency range (Hz)
• dc-2
Standard sensor of method
• specific electrode
Cardiac output
Principal measurement range of parameter:
• 4-25 liter/min
Signal frequency range (Hz)
• dc-20
Standard sensor of method
• dye dilution, fick
Electrocardiography (ECG)
Principal measurement range of parameter:
• 0.5-4 mV
Signal frequency range (Hz)
• 0.01-250
Standard sensor of method
• skin electrodes
Electroencephalography (EEG)
Principal measurement range of parameter:
• 5-300 micro V
Signal frequency range (Hz)
• dc-150
Standard sensor of method
• scalp electrodes
Electrocorticigraphy and brain depth
Principal measurement range of parameter:
• 10-5000 micro V
Signal frequency range (Hz)
• dc-150
Standard sensor of method
• brain surface or depth electrodes
Electrogastrography
Principal measurement range of parameter:
• 10-1000 micro V
• 0.5-80 mV
Signal frequency range (Hz)
• dc-1
• dc-1
Standard sensor of method
• skin surface electrodes
• stomach surface electrodes
Electromyography (EMG)
Principal measurement range of parameter:
• 0.1-5 mV
Signal frequency range (Hz)
• dc-10,000
Standard sensor of method
• needle electrodes
Electro-oculogram (EOG)
Principal measurement range of parameter:
• 50-3500 micro V
Signal frequency range (Hz)
• dc-50
Standard sensor of method
• contact electrodes
Electroretinogram (ERG)
Principal measurement range of parameter:
• 0-900 micro V
Signal frequency range (Hz)
• dc-50
Standard sensor of method
• contact electrodes
Galvanic skin response (GSR)
Principal measurement range of parameter:
• 1-500 k ohms
Signal frequency range (Hz)
• 0.01-1
Standard sensor of method
• skin electrodes
Gastric pH
Principal measurement range of parameter:
• 3-13 pH units
Signal frequency range (Hz)
• dc-1
Standard sensor of method
• pH electrode antimony electrode
gastrointestinal pressure
Principal measurement range of parameter:
• 0-100 cm H2O
Signal frequency range (Hz)
• dc-10
Standard sensor of method
• strain-gage manometer
gastrointestinal forces
Principal measurement range of parameter:
• 1-50g
Signal frequency range (Hz)
• dc-1
Standard sensor of method
• displacement system, LVDT
nerve potentials
Principal measurement range of parameter:
• 0.01-3mV
Signal frequency range (Hz)
• dc-10,000
Standard sensor of method
• surface or needle electrodes
phonocardiography
Principal measurement range of parameter:
• dynamic range 8- dB, threshold about 100 microPa
Signal frequency range (Hz)
• 5-2000
Standard sensor of method
• microphone
plethysmography (volume change)
Principal measurement range of parameter:
• varies w organ
Signal frequency range (Hz)
• dc-30
Standard sensor of method
• displacement chamber or impedance change
circulatory
Principal measurement range of parameter:
• 0-30 ml
Signal frequency range (Hz)
• dc-30
Standard sensor of method
• displacement chamber or impedance change
Respiratory functions Pneumotachograph (flow rate)
Principal measurement range of parameter:
• 0-600 liter/min
Signal frequency range (Hz)
• dc-40
Standard sensor of method
• pneumotachograph head and differential pressure
Respiratory rate
Principal measurement range of parameter:
• 2-50 breaths/min
Signal frequency range (Hz)
• 0.1-10
Standard sensor of method
• strain gage on chest, impedance, nasal thermistor
Tidal volume
Principal measurement range of parameter:
• 50-1000ml/breath
Signal frequency range (Hz)
• 0.1-10
Temp of body
Principal measurement range of parameter:
•32-40 C
• 90-104 F
Signal frequency range (Hz)
• dc-0.1
Standard sensor of method
• themistor
• thermocouple