Advanced Image Acquisition- unit 1- chapter 23 technical considerations in digital imaging

0.0(0)
Studied by 0 people
call kaiCall Kai
Locked
learnLearn
examPractice Test
spaced repetitionSpaced Repetition
heart puzzleMatch
flashcardsFlashcards
GameKnowt Play
Card Sorting

1/92

encourage image

There's no tags or description

Looks like no tags are added yet.

Last updated 3:50 AM on 9/21/26
Name
Mastery
Learn
Test
Matching
Spaced
Call with Kai
Chat

No analytics yet

Send a link to your students to track their progress

93 Terms

1
New cards

what type of technical factors should be used for digital imaging?

high kVp and low mAs

2
New cards

why should high kVp be used in digital imaging?

high kVp ensures adequate penetrability of the body part

3
New cards

why should low mAs be used in digital imaging?

low mAs minimizes radiation exposure

4
New cards

why should a short exposure time be used?

a short exposure time minimizes motion

5
New cards

what is the total amount of exposure reaching the detector called?

the signal

6
New cards

what happens if mAs is too low?

quantum mottle may be introduced

7
New cards

do most under exposures demonstrate a significant amount of noise?

no most underexposures do not

8
New cards

what does the computer do when a digital image is underexposed?

it attempts to adjust the image to make the anatomy visible

9
New cards

what does visible quantum mottle indicate?

an insufficient amount of x-rays have reached the IR

10
New cards

what does insufficient signal mean?

there is insufficient remnant radiation reaching the image receptor

11
New cards

what happens to quantum mottle when adequate exposure reached the image receptor?

the mottle does not actually go away but it is overwhelmed by the good signal

12
New cards

what does sufficient exposure ensure?

a high signal to noise ratio

13
New cards

what can excessive mAs produce?

detector saturation also called data drop

14
New cards

do most digital overexposures demonstrate saturation?

no

15
New cards

what does detector saturation indicate?

multiple times more exposure than necessary was used

16
New cards

how can detector saturation be compared to buckets filled with water?

like full buckets spilling water into neighboring buckets excessive electrical charge overwhelms the receptors storage capacity and spill across an area of the detector plate

17
New cards

what happens to pixel values during detector saturation?

the readouts reach their maximum values, so there is no longer a distinction between pixels and therefore no image to process

18
New cards

approx how much exposure may be required to cause saturation?

about 8-10 timed the normal exposure, some require more

19
New cards

is detector saturation common?

no

20
New cards

why can most digital overexposures still provide acceptable images?

the computer system can usually compensate for the overexposure so the final image has acceptable image quality

21
New cards

what is used to reference image quality and level of exposure?

the exposure indicator number EI#

22
New cards

are EI calculations the same for every manufacturer?

no

23
New cards

what should be established as a standard care for exposure indicators?

target EI numbers

24
New cards

who often provides target EI numbers?

equipment manufacturer

25
New cards

what are the 2 historically proposed exposure technique systems?

fixed kVp with variable mAs and variable kVp with fixed mAs

26
New cards

which technique system is best suited for digital receptors?

the fixed kVp system

27
New cards

how does fixed kVp system work?

it uses an optimum kVp for various body parts

28
New cards

why does the fixed kVp system use an optimum kVp for each body part?

to ensure proper beam penetration

29
New cards

when using a fixed kVp system, by how much should mAs generally be changed?

in increments of at least 30%

30
New cards

what is the suggested optimal kVp range for infant extremities?

50-60 kVp

31
New cards

what is the suggested optimal kVp range for adult extremities?

65-75 kVp

32
New cards

what is the suggested optimal kVp range for bucky extremities?

75-90 kVp

33
New cards

what is the suggested optimal kVp range for an AP spine?

85-95 kVp

34
New cards

what is the suggested optimal kVp range for a lateral cervical spine?

85-100 kVp

35
New cards

what is the suggest optimal kVp range for a lateral lumbar spine?

85-100 kVp

36
New cards

what is the suggested optimal kVp range for chest?

110-130 kVp

37
New cards

what us the suggested optimal kVp range for skull?

80-90 kVp

38
New cards

how does the variable kVp system work?

it establishes a threshold minimum kVp and increases kVp for each 2 cm change in part thickness

39
New cards

what is the typical threshold for minimum kVp in a variable kVp system?

30 kVp

40
New cards

when using the variable kVp system, when is kVp increased?

for each 2 cm change in part thickness

41
New cards

what is the formally for calculating kVp in the variable kVp system?

30 kVp + (2 kVp x part thickenss in cm) = new kVp

42
New cards

why is digital post processing tolerant of overexposure?

saturation is generally only seen at extreme levels of overexposure

43
New cards

does underexposure of overexposure have less exposure latitude?

underexposure

44
New cards

what is digital image dose creep?

the tendency for exposure techniques to gradually increase because digital systems tolerate overexposure while underexposure can produce quantum mottle

45
New cards

why can dose creep occur?

techs may increase exposure techniques to avoid quantum mottle because overexposure is often compensated for by digital system

46
New cards

According to ASRT best practices, what technical factors should technologists use?

The highest kVp within the optimal range for the position and part, coupled with the lowest mAs needed to provide adequate exposure to the image receptor.

47
New cards

Why is keeping exposure near the Target Exposure Index (EIT) important?

Digital systems have wide exposure latitude, so staying near the EIT helps prevent unknowingly using more radiation than necessary.

48
New cards

Why was the Deviation Index (DI) created?

Because most Exposure Index values are manufacturer-specific, so a universal standard was needed.

49
New cards

Who created the Deviation Index (DI)?

The American Association of Physicists in Medicine (AAPM).

50
New cards

What are DI values based on?

The Target Exposure Index (EIT).

51
New cards

Where can EI and DI values be permanently stored?

In the DICOM header of the image.

52
New cards

When should the Deviation Index be displayed?

Immediately after every exposure.

53
New cards

What is the formula for Deviation Index?

DI = 10 log₁₀(EI/EIT).

54
New cards

What does EI represent in the DI formula?

The Exposure Indicator readout.

55
New cards

What does EIT represent in the DI formula?

The target EI for the ideal exposure level.

56
New cards

How does DI change with approximately a 25% increase in overexposure?

The DI increases by +1.0.

57
New cards

How does DI change with approximately a 20% decrease in exposure?

The DI decreases by -1.0.

58
New cards

Are changes in DI additive or multiplicative?

Multiplicative

59
New cards

what does positive DI indicate?

overexposure

60
New cards

what does negative DI indicate?

underexposure

61
New cards

What is the target DI range?

-0.5 to +0.5

62
New cards

What exposure range does a DI of -0.5 to +0.5 represent?

Approximately -20% to +25% exposure.

63
New cards

What does a DI greater than +3.0 indicate?

Greater than 100% overexposure and excessive patient exposure.

64
New cards

What should be done with an image having a DI greater than +3.0?

Do not repeat unless gross saturation occurred. Management follow-up may be appropriate.

65
New cards

What does a DI of +1.0 to +3.0 indicate?

Approximately 25% to 100% overexposure.

66
New cards

What should be done with an image having a DI between +1.0 and +3.0?

Repeat only if saturation occurred and make efforts to properly expose in the future.

67
New cards

What does a DI of -1.0 to -3.0 indicate?

The image received approximately 50% to 80% of the target exposure and is underexposed.

68
New cards

What should be done with an image having a DI between -1.0 and -3.0?

Consider repeating if appropriate and consult the radiologist regarding image quality.

69
New cards

What does a DI less than -3.0 indicate?

Less than 50% of the target exposure and excessive underexposure.

70
New cards

What should be done with an image having a DI less than -3.0?

Repeat because quantum mottle is certain.

71
New cards

Should an overexposed digital image automatically be repeated?

No. Overexposures should not be repeated unless saturation occurred.

72
New cards

Should an exposure that received less than 80% of the EIT automatically be repeated?

No. It should not be repeated unless a radiologist determines the level of mottle is unacceptable.

73
New cards

At what exposure level should an image be assumed to be repeatable?

When the exposure is less than 50% of the EIT.

74
New cards

Should a dark digital image automatically be repeated?

No. A dark digital image should not be repeated regardless of EI unless saturation occurred.

75
New cards

Why is a DI greater than +3 concerning even if image quality is acceptable?

It is an ALARA violation and indicates excessive patient exposure.

76
New cards

What is automatic electronic masking?

An automatic process that places a black overlay over areas of the image receptor that were not exposed.

77
New cards

What does automatic electronic masking remove?

Nonclinical data from the image file.

78
New cards

Why is automatic electronic masking used?

To eliminate extraneous brightness from the image edge.

79
New cards

What does electronic masking reduce?

veil glare

80
New cards

What is veil glare?

Veil glare occurs when intense light from a light source floods the eye directly, overstimulates photoreceptors in the retina, and reduces contrast perception.

81
New cards

Can electronic masking be performed manually?

yes

82
New cards

When is manual electronic masking useful?

When automatic masking incorrectly identifies the collimated field.

83
New cards

What are other names for manual electronic masking?

cropping or shuttering

84
New cards

Is electronic masking a substitute for proper collimation?

no

85
New cards

How can electronic masking affect older digital systems?

It can impact the accuracy of the EI number.

86
New cards

Why can cropping anatomy from an image be dangerous?

Incidental findings could be missed.

87
New cards

What is the ASRT position on electronic masking?

A digital image should not be cropped or masked so that it eliminates areas of exposure from the image presented for interpretation.

88
New cards

What is electronic annotation also called?

digital annotation

89
New cards

What is electronic annotation used for?

To indicate image acquisition conditions.

90
New cards

What are examples of information added using electronic annotation?

Upright, decubitus, inspiration, expiration, standing, etc.

91
New cards

Can electronic annotation replace lead markers for laterality?

no

92
New cards

What should be used to indicate patient laterality?

lead R or L an atomic side markers

93
New cards

What is the ASRT best-practice recommendation for anatomic side markers?

Consistently use lead anatomic side markers captured on the original image during the x-ray exposure.