Flashcards created from the lecture notes covering Digital Radiography (DR), Computed Radiography (CR), image processing, histograms, detectors, and DICOM standards.
Name | Mastery | Learn | Test | Matching | Spaced | Call with Kai | Chat |
|---|
No analytics yet
Send a link to your students to track their progress
What does the deviation index compare?
Acquired exposure to target exposure
What did the creation of electronic data sets allow in medical imaging?
Computer manipulation of imaging data
Digital radiography receptors are commonly referred to as:
Flat-panel detectors
The primary function of a look-up table (LUT) is to adjust:
Image contrast
High-pass filtering is most closely associated with:
Edge enhancement
Pixel pitch is dependent on which two factors?
Matrix size and image receptor size
Early digital computerization was first applied to which imaging modalities?
CT and ultrasound
Higher DQE is associated with:
Lower patient dose for the same image quality
Which process begins grayscale point processing?
Creation of a histogram
Histogram rescaling involves comparing the acquired histogram to a:
Reference histogram
Histogram equalization is recognized as a standard procedure by which organization?
American College of Radiology (ACR)
One responsibility of the technologist when using EI is to:
Ensure ALARA compliance
Major developments in digital image processing occurred primarily during which decades?
1970s through 1990s
A histogram is best described as:
A graphical representation of signal values
Standardized EI and DI values are especially important because:
New and upgraded digital equipment continues to be introduced
The field of view (FOV) determines the:
Area of anatomy included in the image
Gray scale bit depth in digital imaging typically ranges between:
8 and 32 bits
How many bits are required to form one byte?
Eight
Which exposure condition cannot be corrected by histogram rescaling?
Excessive overexposure
Point processing operations primarily involve:
Adjusting pixel values one at a time
Why should overexposure be avoided in CR imaging despite wide exposure latitude?
To reduce unnecessary patient dose
Typical CR plate throughput ranges between:
30 and 200 plates per hour
What allows liberated electrons in the imaging plate to move within the crystal structure?
Laser stimulation
Which step in CR imaging involves formation of a latent image?
Image acquisition
CR image acquisition can be performed using which setup?
Tabletop or with a grid
Which two methods are used to read CR imaging plates?
Point-by-point and line-by-line readout
Computed radiography systems use which type of image receptor?
Photostimulable imaging plates
Low band-pass filtering is primarily used to:
Suppress visible image noise
Computed radiography was first developed during which decade?
Early 1980s
In computed radiography, data manipulation is also known as:
Post-processing
Which factor can falsely trigger collimator edge identification algorithms?
Prosthetic devices or implants
Proper positioning in CR imaging is important because it helps prevent:
Histogram analysis errors
Which method may be used to collect the electrical signal produced in an indirect DR detector?
Thin-film transistor or charge-coupled device
In a TFT array, pixel size is directly related to the size of the:
Detector element (DEL)
Which action helps prevent fluid-related damage to DR panels during portable exams?
Enclosing the panel in a protective bag
Direct conversion detectors use which material as a photoconductor?
Amorphous selenium
Exceeding weight load limits on a DR panel may result in:
Permanent mechanical damage
A limitation of traditional generator-dependent DR panels is that they:
Use vendor-specific interfaces
Which scintillation materials are commonly used in indirect flat-panel detectors?
Cesium iodide and gadolinium oxysulfide
Flat-panel detectors were first introduced into clinical radiography in approximately:
1995
In digital radiography, the latent image is best described as:
Electrical signal values stored in detector elements
A characteristic limitation associated with indirect conversion detectors is:
Light spread that may reduce spatial resolution
Thin-film transistors are used in which detector types?
Both indirect and direct detectors
Which component is common to both direct and indirect flat-panel detector systems for signal readout?
Thin-film transistor (TFT) array
Panels that detect x-radiation without a direct generator interface are commonly called:
Trigger panels
In auto-detection technology, the TFT array begins collecting signal when:
X-radiation is detected by the panel
Which configuration describes a DR panel that is permanently mounted within the imaging system?
Integral to equipment design
The primary distinction between direct and indirect DR panels is based on:
The method of x-ray energy conversion
TFT fill factor is defined as the percentage of a detector element that consists of:
Active sensing material
Responsibility for understanding manufacturer-specific DR panel restrictions belongs to the:
Medical physicist
What does DICOM stand for?
Digital Imaging and Communications in Medicine
Analog-to-digital conversion consists of which two steps?
Sampling and quantification
LUT adjustments are most similar to changing which film characteristic?
D-log-E curve
What is the purpose of applying a look-up table (LUT)?
To map pixel values to display brightness
What process converts analog signals into digital data?
Analog-to-digital conversion
The volume of interest (VOI) is identified during which step?
Histogram analysis
Point processing in digital imaging affects:
Individual pixels
What is a key advantage of having a quantifiable exposure index?
It allows objective assessment of image quality and ALARA compliance
Why were healthcare reimbursement structures modified in favor of DR systems?
To encourage migration from CR to DR
A DQE value of 1 represents:
100 percent efficiency with no information loss
The primary purpose of the DICOM standard is to:
Ensure consistent storage, display, and sharing of images
Why is there no universal exposure index system in digital radiography?
Different manufacturers use different exposure indicator systems
Density resolution increases when:
Gray scale bit depth increases
LUT values are assigned to which data points?
Pixel values in the histogram
Sampling frequency is commonly expressed as:
Pixels per millimeter
Where are electrons trapped during latent image formation?
Europium sites
The CR readout process occurs in which manner?
Line by line only
Photostimulable imaging plates are best described as:
Reusable, multilayer rigid sheets
Which post-processing technique enhances cortical margins and trabecular detail?
High band-pass filtering
Which interaction is primarily responsible for energy absorption in CR phosphors?
Photoelectric interaction
Suboptimum mAs exposures in CR imaging are most likely to result in:
Increased quantum noise
Pixel pitch refers to the:
Distance between adjacent pixels
What type of laser is used to scan CR imaging plates?
Neon–helium laser
During digitization, the analog signal is converted into:
A limited set of discrete values
Spatial resolution in a TFT-based detector is indirectly related to:
DEL dimensions
Why does amorphous silicon require the use of a scintillator in indirect detectors?
Its atomic number is too low to efficiently absorb x-ray energy
Flat-panel detectors are most accurately described as part of which imaging modality?
Digital radiography (DR)
Which feature distinguishes direct detectors from indirect detectors?
Absence of a scintillation layer
A detector with a 95 percent fill factor contains approximately what percentage of non-sensing electronics?
5 percent
In digital radiography, how can the latent image also be described?
A stored phosphor image
Which detector class converts x-ray energy directly into an electrical signal without producing light?
Direct conversion
Which image processing step identifies relevant anatomy and excludes non-image areas?
Exposure field recognition
Which term is most commonly used in clinical practice to refer to a flat-panel detector?
Panel