Which montage primarily localizes activity by phase reversals?
Bipolar
Common mode rejection (CMR) refers to the amplifier’s ability to:
Reject identical signals at both inputs
A differential amplifier measures:
The difference between input 1 and input 2
If a negative signal enters Input 1( input 1 more negative than input 2), the deflection typically moves:
Upward
Sensitivity is measured in:
μV/mm
Increasing the sensitivity setting (μV/mm) makes waveforms appear:
Attenuated
In a referential montage, localization is based mainly on:
Amplitude differences
Muscle (EMG) artifact is most often prominent in
Temporal/frontal regions
The 60 Hz (notch) filter primarily targets
Electrical interference
A clear phase reversal in a bipolar chain usually indicates:
A focal localization point
The time constant is part of which filter system?
Low frequency filter
Impedance measure
Resistance at an electrode site
Channel polarity is determined by:
Input 1-input 2
A low frequency filter (LFF) attenuates:
Slow activity
A high frequency (HFF) attenuates:
Fast activity
Filters affect the amplitude of EEG waves, not the:
Frequency
When ‘end of chain’ localization is suspected in a bipolar montage, the best first step is to:
Change montage to Hat
The human eye behaves electrically as a:
Dipole
The LFF is commonly used to reduce:
Slow baseline drift/sweat components
The HFF is commonly used to reduce:
Fast activity/EMG components
In a longitudinal (double banana) montage, a focal spike is most confidently localized at the site of the:
Phase reversal
In a referential montage, a focal spike is most confidently localized at the electrode with:
Highest amplitude to reference
Which montage best highlights interhemispheric asymmetries across the midline (F7-F8 vs T3-74 etc)?
Transverse (coronal) bipolar
Which filter is best to reduce sweat artifact without eliminating alpha?
Increase LFF (raise from 0.1 to 1 Hz)
Raising sensitivity from 7 uv/mm to 10 uv/mm will make a 70 uv spike appear:
Attenuated
Which rhythm is most reactive to eye opening in awake adults?
Alpha
When blocking (clipping) occurs due to very high amplitudes, the quickest adjustment prevent further blocking is to:
Increase sensitivity
Which artifact produces steep, irregular, high frequency activity (>20Hz) predominantly in frontal-temporal channels?
EMG (muscle)
Lowering the high-frequency filter form 70 Hz to 35 Hz will likely:
Reduce EMG contamination
The notch filter should be used:
Only when necessary due to potential waveform distortion
In average reference montage, a unilateral focal spike may appear smaller if:
Many electrodes are involved by the same field