1/90
Looks like no tags are added yet.
Name | Mastery | Learn | Test | Matching | Spaced | Call with Kai | Chat |
|---|
No analytics yet
Send a link to your students to track their progress
Type a
X-rays interact with the patient and
are scattered away from the image
receptor.
Type b
X- rays that interact with the patient
and are absorbed.
Type c
X- rays that arrive at the image
receptor without patient interaction.
Type d
X- rays that is scattered by the
patient reaching the image receptor.
type c and d
are called Image forming X-rays
remnant x-rays
x- rays that exit from the patient are called
Image-Forming X-rays
x rays that exit and interact with the image receptor are called
kVp
As x-ray energy is increased, the absolute number of
Compton interactions decreases , but the number of
Photoelectric interactions decreases much more rapidly .
Therefore, the relative number of x-rays that undergo
Compton scattering increases.
1%
approximately __ of x-rays incident on the patient reach the image receptor.
directly related
kVp and scatter, inverse or direct?
scatter
high kVp =
acceptable scatter and px dose
optimal kVp =
high patient dose (alara)
low kVp =
5%
will result in a noticeable (but small) change in Image Receptor (IR) Exposure
has the same effect on image receptor as a 30% change in mAs .
This rule is used to make small exposure adjustments.
30% change in mAs
5% change in kVp =
reciprocity law
Optical Density should remain the same for a given
mAs , regardless of the combination of mA and time
used.
direct
IR exposure = mas
direct or inverse?
underexposure
Radiograph: too light or low OD
Rationale: too little x-rays reaches the image receptor
Can cause quantum mottle
overexposure
✓ Radiograph: too dark or High OD
✓ Rationale: too much x-rays reaches the image receptor
high contrast
few shades of gray
increase contrast
short-scale contrast
short (narrow) dynamic range
narrow window width
low contrast
many shades of gray
decreased contrast
long-scale contrast
large (wide) dynamic range
wide window width
inverse
collimation = field size
direct or inverse?
direct
field size = scatter
direct or inverse?
inverse
collimation = scatter
direct or inverse?
direct
collimation = contrast
increased collimation
patient dose decreases
scatter radiation decrease
radiographic contrast increases
exposure to image receptor decreases
increased field size
patient dose increases
scatter radiation increases
radiographic contrast decreases
exposure to image receptor increases
larger body parts
have more tissue to interact with the photons, resulting in greater scatter production .
direct
thickness = scatter
direct or inverse?
compression of anatomy
improves spatial resolution and contrast resolution and lowers the patient radiation dose.
by using a compression device
thickness decreases
oid decreases
spatial resolution increases
patient dose decreases
contrast resolution increases
contrast
the degree of difference in OD between areas of a radiographic image
contrast resolution
the ability to image and distinguish soft tissues
reduced image contrast
results from scattered x-rays
Aperture Diaphragm
» The simplest of all beam -restricting devices .
» It is basically a lead or lead lined metal diaphragm that is attached to the x -ray tube head.
» The opening in the diaphragm usually is designed to cover just less than the size of the image receptor used.
»The main disadvantage of the diaphragm is that is does not provide a sharp demarcation of
the edge of the x -ray beam
»With a diaphragm, Off - focus is not controlled, and there is a large area of penumbra at the edge of the exposed film area.
Cones & Cylinders
» Radiographic extension cones and cylinders are considered modifications of the aperture
diaphragm
» The useful beam produced by an extension cone or cylinder is usually circular .
» The position and size of the distal end act as an aperture and determine field size.
Cone-cutting
It is impossible for the circular cross section
of the x-ray beam to be fitted to a
rectangular film.
aperture diaphragm, cones and cylinders, variable aperture collimator
types of beam restrictors
collimation
reduces the patient radiation dose and improves contrast resolution
variable-aperture collimator
is the most commonly used beam-restricting device in radiography
gustave buckey, 1913
who invented the radiographic grid and continues to be the most effective means for limiting the amount of scatter radiation that reaches the IR
grid
is a device that has very thin lead strips with radiolucent interspaces, intended to absorb scatter radiation emitted from the patient
10 cm (4 inches) or greater in thickness, and more than 60 kvp
grids are typically used only when the anatomic part is ____?
dr. Hollis potter
from Chicago made the bucky grid practical by moving it during the radiographic exposure, a procedure that blurred the grid line out of the image
the grid has since been known as the Potter-Bucky diaphragm, which is also the correct NEMA term for the device
high ratio grids
are more effective in reducing scatter radiation than are low-ratio grids
5;1, 16;1
higher ratio grids are used most often in high-kVp
85%
5;1 grid reduces approximately __ of the scatter radiation
97%
16;1 grid may reduce as much as __
grid frequency
the number of grid strips per centimeter
aluminum or plastic fiber
the interspace material of most grids consists of
nonhydroscopic
does not absorb moisture as plastic fiber does
aluminum
easier to form and roll into sheets of precise thickness
lead
easy to shape and is relatively inexpensive
high atomic number and high mass density
high atomic number and high mass density
grid strip
it should be infinitely thin and should have high absorption properties
bucky factor (B) or grid factor
when a grid is used, the radiographic technique must be increase to produce the same image Receptor signal
the amount of this increase is given by the
gustave bucky
the bucky factor is named for ______, the inventor of the grid
it is an attempt to measure the penetration of primary and scatter radiation through the grid
air gap technique
alternative to the use of radiographic grids
reducing scatter radiation, thereby enhancing image contrast
10-15 cm
when the air-gap technique is used, the image receptor is moved ___ from the patient
linear/parallel grid
easiest to manufacture, but it has some properties that are clinically undesirable
LESS COMMONLY EMPLOYED
- because the strips do not try to coincide with the divergence of the x-ray beam
- some grid cutoff will occur along the lateral edges, especially when the grid is employed at short SID
grid cutoff
the undesirable absorption of primary x-rays by the gird
most common with parallel grids
most pronounced when
1. short SID
2. large-area image receptor
parallel grid
best employed at long SID
because the beam will be straighter and perpendicular
cross-hatch/ cross grid
two sets of lead strips superimposed and running and 90* to one another
this design is used only in stationary grids
generally, a stationary grid is used for high kV work and no x-ray tube angulation
beam must be aligned with the center of grid! there should be no angulation
more efficient than parallel grids in cleaning up scatter radiation
has a higher contrast improvement factor than a parallel grids of twice the gird ratio
ex: 6:1 cross grid will clean up more scatter radiation than a 12:1 parallel grid
disadvantages of cross grids
positioning the grid critical
- central ray of the x-ray beam must coincide with the center of the grid
tilt-table techniques
- possible only if the x-ray tube and table are properly aligned
exposure technique
- requires a substantial increase with resulting higher patient radiation dose
grid cutoff
the main disadvantage of parallel and crossed grids
focused grid
lead strips are inclined inward
focused on a pre-determined point above the grid (focal range)
designed to minimize grid cutoff
coincide with the divergence of the x-ray beam
match the angle of divergence of the primary beam
non-focused grid
a parallel line grid
relationship of lead strips is uniform to one another
focused grid disadvantages
more difficult to manufacture than parallel grids
they are characterized by all the properties of parallel grids except when properly positioned, they exhibit no grid cutoff
should be used carefully for there are geometric limitations (SID)
every focused grid is marked with its intended focal distance and the side of the grid that should face the x-ray tube
grid lines
an obvious and annoying shortcoming of the grids
the presence of grid lines can be demonstrated simply by radiographing a grid
grid lines are the images made when primary x-rays are absorbed within the grid strips. EVEN THOUGH THE GRID STRIPS ARE VERY SMALL, their image is still observable
high-frew grids present less obvious grid lines compared to low-freq grids
visibility of grid lines = width of grid strips
moving grids
placed in a holding mechanism that begins moving just before x-ray exposure and continues moving after the exposure ends.
2 basic types of moving grid mechanisms:
- reciprocating grid
- oscillating grid
reciprocating grid
is a moving grid that is motor driven back and forth several times during x-ray exposure. the total distance of drive is approximately 2cm
oscillating grid
is positioned within a frame with a 2- to 3-cm tolerance on all sides between the frame and the grid. delicate, springlike devices located in the four corners hold the grid centered within the frame. a powerful electromagnet pulls the grid to one side and releases it at the beginning of the exposure. Thereafter, the grid oscillates in a circular fashion around the grid frame, coming to rest after 20 to 30 seconds.
3 possible motions of moving grids
single stroke (one way)
reciprocating (forward & backward)
catapult
off level
usually produced with an improperly positioned x-ray tube, not an improperly positioned grid
occurs when the grid tilts during horizontal beam radiography or during mobile radiography
off center-off axis
problem usually seen in focused grids
any lateral shift results in grid cutoff across the entire radiograph, producing lower OD
this error in positioning is called LATERAL DECENTERING
off- focused grid
problems in using focused grid is when using SIDs unspecified for that grid or outide its focal range.
the farther the grid from its focal range, the more severe the grid cutoff
cutoff is more severe at edges
upside-down grid
shows a severe cutoff on either side of the central ray
mammography
low-ratio grids are commonly used during?
grid ratios up to 8:1
grid represents a good compromise between desired levels of scatter radiation reduction and patient radiation dose .
are satisfactory at tube potential below 90 kVp
- are used when kVp exceeds 90 kVp
grid selection
» Grids absorb scatter
» Scatter adds exposure to the image receptor
» The more efficient a grid is at absorbing scatter, the less image receptor exposure
» Therefore, compensations must be made to increase this Exposure!
» This compensation is generally accomplished by INCREASING mAs !
» Increasing mAs results to increase in patient dose .
» The better the grid cleans up scatter, the greater will be
the dose to the patient in achieving proper IR exposure!
grid conversion factor (gcf)
the amount of mas needed can be calculated using the:
post-reduction
those taken after manipulation and casting have occured—are commonly performed to confirm alignment of the bone.
fiberglass/plaster mix
increase mAs 50% or 8% kVp
soft tissue technique
decrease kVp 20%, leave mAs on the same on
35% increase in technique
for pooling of blood in head, thorax, and abdomen on postmortem (dead person)
expiration chest technique
post mortem- an
increase 35% in mAs
post-mortem chest x-ray for expiration chests immediately after death
increase 50% in mAs
is required on post-mortem chest x-ray for expiration chests (if ½ HR or more has elapsed)
commonly used for board question
conversion factors for cylindrical cones
extension cylinder collapsed- increase mAs by 40% or 5 kVp
extension cylinder extended increase 60% mAs or 10 kVp