Practicability
Method is easily repeated
Reliability
Maintain accuracy and precision
Intralab/Interlab QC
Daily monitoring of accuracy and precision
Interlab/External QC
Proficiency testing (Reference lab)
Long-term accuracy
*Difference of > or = 2 = NOT IN AGREEMENT WITH OTHER LAB
QC materials
available for minimum of 1 year
Bovine Control Materials
Preferred (Human: Biohazard)
NOT for immunochem, dye-binding, and bilirubin
Matrix effect
Due to improper product manufacturing
Unpurified analyte
Altered protein
Precision study
First step in method evaluation
Nonlab. personnel
29% of errors (lab results)
SD (Standard Deviation)
Dispersion of values from the mean
CV (Coefficient of Variation)
Index of precision
Relative magnitude of variability (%)
Variance
SD squared
Measure of variability
Inferential statistics
Compare means or SD of 2 groups of data
T-test
Means of 2 groups of data
F-test
SD of 2 groups of data
Cumulative Sum Graph (CUSUM)
V-mask
Earliest indication of systematic errors (trend)
Youden/Twin Plot
Compare results obtained from different lab
Shewhart Levey-Jennings Chart
Graphic representation of the acceptable limits of variation
Trend
Gradual loss of reliability
Cause: Deterioration of reagents (Systematic error)
Shift
Values: one side or either side of the mean
Cause: Improper calibration (Systematic error)
Outliers
Values: far from the main set of values
Highly deviating values
Random or systematic errors
Kurtosis
Degree of flatness or sharpness
Precision
Reproducibility of a value
Random error
Accuracy
Closeness of value to true value
Systematic error
Random error (Imprecision; Indeterminate)
Causes:
-Mislabeling
-Pipetting error
-Improper mixing of sample and reagents
-Voltage/Temperature fluctuation
-Dirty optics
Parameters: SD and CV
Systematic error (Inaccuracy/Determinate)
Causes:
-Improper calibration
-Deterioration of reagents
-Contaminated solution
-Sample instability/unstable reagent blanks
-Diminishing lamp power
-Incorrect sample and reagent volume
Parameter: Mean
Multirule Shewhart procedure
Control rules + Control Chart
Test method
involves at least 40 samples (Westgard)
Reference method
preferably 100 samples (Westgard
Analytical Run
Control and patient specimens assayed, evaluated, and report together
Physiologic Limit
Referred to as absurd value
POCT (Point of Care Testing)
Performed by nonlab personnel
Quality Assurance
Tripod:
Program development
Assessment and monitoring
Quality improvement
Quality Patient Care
Test Request forms, clear instruction for patient preparation, proper specimen handling, etc.
Reference Range/Interval Range/Reference values
At least 120 individuals should be tested in each age and sex category