What is signal to noise ratio?
Ratio between signal detected and amplitude of noise
What are five factors afffecting the SNR?
Magnetic field strength
Coil types and position
TR, TE and flip angle
Number of signal averages
Voxel column and slice thickness
How does increasing slice thickness affect the signal?
Increases number of hydrogen atoms that can be used to produce a signal, producing more information/signal
What are the disadvantages of increasing Slice thickness?
Loss of spatial resolution
What is the matrix size?
Number of pixels in the image
How does increasing the matrix size affect the signal?
Decreases the signal to noise ratio
What are the disadvantages of increasing matrix size?
Increase the time
What is the average number of excitations?
Number of times the data is acquired repeatedly to form the same image
What is the disadvantage of increasing the number of excitations?
Increases scan time
How does average number of excitations increase snr?
Superimposes multiple pictures and removes moving artefacts
What are the four types of artefacts in MRI?
Motion
Radiofrequency
Image processing
Magnetic susceptibility
What is phase mismapping?
Replication of anatomy in the phase encoding direction caused by motion
What are four ways to compensate for motion artefacts?
Phase swap
Pre-saturation bands
Respiratory navigator
Patient comfort
What is phase swap?
Swapping phase encoding direction to avoid creating artefact in area of interest
What are the disadvantages of phase swap?
May cause wrap around artefact
What is wrap around artefact?
Signal from anatomy outside of field of view appears flipped on the opposite side of the image
What are presaturation bands?
Null the signal from the area the band is placed on
What is a respiratory navigator?
Monitors signal when anatomy is motionless and data acquired outside these boundaries
What are the disadvantages of respiratory navigator?
Can be inconsistent if patients respiratory rate is inconsistent
What is a zipper artefact?
Interference of external radiofrequency waves disrupting the magnetic field
What is the cause of wrap around artefacts?
Anatomy close to the receiver coil produces a signal
What are the three remedies for wrap around artefacts?
Increase field of view
Saturation bands
Change phase encoding direction
What are the two values determining field of view?
Field of view
Phase field of view
What is field of view?
Total area being imaged
What is the phase field of view?
Portion of field of view used in phase encoding direction
How does changing phase encoding direction reduce wrap artefact?
Avoids anatomy involved in the artefact
How do magnetic artefacts affect images?
Produce signal void
How does magnetic interference work?
Metal objects become magnetised and causes inhomogenities
Dephase hydrogen in surrounding tissues, causing signal loss
What are the two remedies to reduce magnetic susceptibility?
Spin echo sequences
Metal Artefact Reduction Sequences
What are the benefits of paramagnetic properties of deoxygenated blood?
Identify haemorrhages in gradient echo pulses