1/21
Looks like no tags are added yet.
Name | Mastery | Learn | Test | Matching | Spaced |
---|
No study sessions yet.
Define Quality Assurance (QA)
Deals with people
Define Quality Control (QC)
Deals with instrumentation and equipment
What does Quality Assurance Monitors
-Proper patient scheduling
-Reception
-Preparation
-Appropriateness of the exams
-Whether proper instruction is given
-Whether patients condition agree with radiologist's diagnosis(
outcome analysis)
-Was the radiology report available for the ordering physician in a timely manner
What does Quality Control Monitors
-Equipment conditions
-Image processing facilities
-Image analysis for artifacts and deficiencies
-Repeat rate
-Radiation dose to the patient
How often should Monitoring of Equipment performace
Annually unless a major component is replaced 1) Replacing X-Ray Tube 2) Replacing the entire machine
Who is responsible for Quality Control (QC)?
-Principally the responsibility of the medical physicist
-Requires a team effort
-QC program focuses on various characteristics and strength of the image
QC program focuses on various characteristics and strengths of the image which is
-Spatial resolution
-Contrast resolution
-Temporal resolution
-Signal to noise ratio
-Artifacts
Define Spatial Resolution
Is the ability to image small objects that have high subject contrast
Contrast Resolution
Is the ability to image two separate objects and visually distinguish one from the other
Temporal Resolution
The time needed to acquire the data to generate an image
Signal to Nosie Ratio
Is calculated from the ratio of the mean signal and the standard deviation of the noise within a region of interest
Artifacts
Is any feature which appears in an image which is not present in the original imaged object
Who developed the guidelines for QC of Screen Film Radiography
American College of Radiology (ACR)
American Association of Physicists in Medicine
Filtration
Ensures patient protection
2.5mm Al required by the state
Measured in terms of HVL
Evaluated annually or after a change in x-ray tube or tube housing
Collimation
Misalignment should not exceed 2% of the SID
Distance indicator must be accurate to 2% of SID
Centering indicator must be accurate to 1% of the SID
Semi annually
Focal Spot Size
-Determines the spatial resolution
-Pinhole camera
-Star pattern
-Slit camera
-Checked annually or when new equipment or replacement x-ray tube is installed
Alternate to focal spot size measurement
- Line pair test tool
- Provide spatial frequency
-Higher spatial frequency indicate better spatial resolution
kVP Calibration
Annually or whenever generator components are changed
4% change in kVp is required for a visible change in OD
Measured kVp should be within 10% of the indicated kVp
Exposure Timer Accuracy
Accuracy of 5% of the indicated exposure time for the exposures greater than 10 ms
Accuracy of 20% for exposure time less than 10 ms
Annually or when components are changed
Automatic Exposure Control (AEC)
Backup timer should terminate the exposure at 6s or 600 mAs whichever occurs first
Or 1.5 times the actual exposure time
Exposure Linearity
Exposure linearity must be within 10% of the adjacent mA stations
mR/mAs value must within 10% between adjacent mA stations
Exposure Reproducibility
Radiation exposure should be reproducible to within 5%
Output radiation intensity should not vary more than 5%
Annually