1/37
Looks like no tags are added yet.
Name | Mastery | Learn | Test | Matching | Spaced |
---|
No study sessions yet.
quality assurance
the routine, periodic evaluation of an ultrasound system to guarantee optimal image quality
how often should quality assurance evaluations be performed?
periodically and routinely
what are the 4 requirements for a quality assurance program?
assessment of system components
repairs
preventive maintenance
record keeping
perfomance testing
scheduled activities that evaluate the accuracy of the system’s imaging capabilities
proper methods for quality assurance programs include:
test under known, defined conditions
use constant instrument settings
use a phantom with measurable characteristics
image in an identical environment
devices that measure the acoustic output of the instrument:
this group considers only the beam former and the transducer acting together as a source of ultrasound
devices that test the operation of the instrument (anatomic imaging and flow evaluation performance):
this group takes into account the operation of the entire instrument
imaging and doppler performance are important in evaluating the instrument as a diagnostic tool
the acoustic output of an instrument is important when considering bioeffects, risk and safety
what are the 2 types of commercially available devices that test imaging performance?
tissue-equivalent phantoms
test objects
tissue-equivalent phantoms
simulate tissue properties, allowing assessment of detail and contrast resolutions, penetrations, dynamic range, and time-gain-compensation operation
what are tissue-equivalent phantoms made out of?
graphite-filled aqueous gels or urethane rubber materials
graphite particles act as ultrasound scatterers in materials
test objects
DO NOT simulate tissue characteristics but do provide some specific measure of instrument performance
the beam-profile slice-thickness test object contains a thin, scattering layer in an echo-free material
the object can be used to show beam width in the scan plane or perpendicular to it (section thickness)
doppler testing phantom
evaluates the doppler capabilities of imaging systems
utilize physical moving structure, such as a vibrating string, a moving belt, or a circulation pump that moves blood-simulating fluid through the phantom
what can be evaluated using a doppler testing phantom?
doppler sensitivity
depth resolution
volume flow
velocity accuracy
slice thickness phantom
evaluates the elevational plane to the thickness portion of the sound beam perpendicular to the imaging plane
contains a diffuse scattering plane that is at an angle to the incident sound beam
mimics soft tissue
sensitivity
refers to the ability of a system to display low-level echoes
what are the two forms of sensitivity?
normal
maximum
normal sensitivity
with these settings, all the pins, solid masses, and cystic structures in the test phantom are accurately displayed
what is adjusted to establish normal sensitivity?
output power
TGC
amplification
all subsequent quality assurance and performance measurements are made at what form of sensitivity?
normal sensitivity
maximum sensitivity
with these settings, a tissue equivalent phantom is imaged and the depth of tissue-like texture on the display is measured
what is maximum sensitivity evaluated with?
evaluated with the output power and amplification of the system set to the maximum practical levels
dead zone
the region close to the transducer where images are inaccurate
the dead zone extends from the transducer to the:
shallowest depth from which meaningful reflections appear
information within the dead zone is:
unreliable and may not be used in the diagnostic setting
the dead zone is assessed with:
the shallowest series of pins in the test object
in a tissue equivalent phantom, the dead zone is the:
shallowest depth at which uniform tissue texture appears
what type of transducer frequency has a thinner dead zone?
higher frequency transducers
registration accuracy
the ability of the system to place reflections in proper positions while imaging from different orientations
range accuracy/vertical depth calibration
describes the system’s accuracy in placing reflectors at correct depths located parallel to the sound beam
if differences appear between the ultrasound scan and the actual pin locations in the test object, the error may be caused by:
system malfunction
the speed of sound in the phantom is different than 1,540 m/s
horizontal calibration
the system’s ability to place echoes in their correct position when the reflectors are perpendicular to the sound beam
distance measurement accuracy
components of the ultrasound system used to measure distances require periodic evaluation
digital calipers should be checked in both vertical and horizontal directions
focal zone
surround the focus
lateral resolution is excellent because the beam is narrow
the focus of phased array transducers must be carefully evaluated
axial resolution
evaluated by scanning sets of successively closer spaced pins within the phantom
lateral resolution
evaluated by measuring the width of reflections on the display created by the point targets in the phantom
compensation operation or uniformity
describes the system’s ability to display similar reflectors in the phantom with echoes of equal brightness
with proper TGC, identical reflectors should have the same appearance on the monitor regardless of their depth
mock cysts and solid masses
tissue-equivalent phantoms are used to evaluate dimension, texture and fill-in of cysts
the system should accurately display hollow structures as anechoic and solid structures as hyperechoic
display, hardcopy output and gray scale dynamic range
adjustments to the output power and amplification of the system should alter the appearance of the image on the display and all output devices
adjustments on a single display device (such as the brightness or contrast of a monitor) alter the image on that particular device only