A lightproof space designed for developing x-ray films into visible diagnostic images using chemical solutions or an automatic processor.
Latensification
An increase in sensitivity, by as much as 2-8 times, in film that has already been exposed to light or ionizing radiation compared to unexposed film.
Film Sensitivities
Film can be affected by excess heat, humidity, static electricity, pressure, and chemical fumes.
OSHA Classification
The Occupational Safety and Health Administration classifies a darkroom as a laboratory.
Minimum Darkroom Guidelines
The space should be clean, well-ventilated, well-organized, a non-smoking zone, and static-free.
Ventilation Functions
Prevents the build-up of heat and humidity and removes excessive fumes from processing solutions that may sensitize film emulsions.
Ventilation Placement
A ventilator duct should be placed near the floor, in either the lower portion of the entrance door or the wall.
Humidity Range
The proper range is 30% to 60% of relative humidity.
Psychrometer
A device used to monitor humidity in the darkroom.
Static Electricity
Proper humidity minimizes the induction of static electricity.
Temperature Range
Should be maintained between 65° to 75°F (18° to 24°C).
Processing Fumes
Considered toxic, corrosive, and potentially carcinogenic.
OSHA PEL for Acetic Acid
25 mg/m³ (10 parts per million [ppm]).
OSHA PEL for Ammonium Thiosulfate
35 mg/m³ (as ammonia) (50 ppm).
OSHA PEL for Hydroquinone
2 mg/m³ (0.44 ppm).
OSHA PEL for Phenol
19 mg/m³ (5 ppm).
OSHA PEL for Sulfur Dioxide
13 mg/m³ (5 ppm).
OSHA PEL for Glutaraldehyde
0.7 mg/m³ (0.2 ppm).
OSHA PEL for Silver
0.01 mg/m³.
Darkroom Disease
A condition where radiographers or darkroom technicians have hypersensitivity to darkroom chemicals.
Darkroom Disease Symptoms
Hives, severe fatigue, and impairment of the immune system.
Indications of Proper Ventilation
Humidity and temperature are within range and chemical exposure is well below PEL.
Drainage Capacity
Drains must handle 2.5 times the maximum outflow of the processor when all drains are open.
Darkroom Walls
Should use enamel or epoxy paints in pastel and light colors with a matte finish and have shielding equivalent to adjacent radiation areas.
Storage Methodology
Uses a "first in, first out" methodology with separate areas for chemicals and films.
Film Storage Position
Films must be stored in an upright position.
Passbox
A film transfer system that prohibits white light from entering.
Static Reduction Methods
Grounding countertops/mats, proper film handling, wearing natural fibers like cotton, and regular screen cleaning with antistatic cleaner.
Darkroom Lighting Types
Consists of overhead lights and safelights.
Overhead Lighting
Standard white light necessary for cleaning, maintenance, and emergencies.
Overhead Light Standards
Standard fluorescent fixtures with 2-4 48-inch tubes per 8 square feet, interlocked with the film bin.
Safelight
A light source emitting wavelengths to which particular films are not sensitive, placed 3-4 feet from work areas.
Average-sized Darkroom Safelight
Fixture with a 7.5W or 15W bulb covered by a filter like Kodak Wratten 6B or Kodak Mor-Lite.
Large Darkroom Safelight
Sodium vapor lamp mounted at least 6 feet above counters for bright conditions.
Blue-Violet-Sensitive Film
Sensitive to blue, indigo, and violet; normally uses an amber-colored safelight.
Orthochromatic Film
Mainly sensitive to the green spectrum plus blue-violet; uses red or magenta dye and filters like Kodak GBX or Kodak 2.
Panchromatic Film
Sensitive to all visible colors; uses a dark green filter like Kodak Number 7.
Light and Leakage Testing
Includes safelight testing and darkroom light leakage testing.
Safelight Unsafe Conditions
Caused by cracks/pinholes in filters, wrong bulb wattage, or sodium vapor lamp doors open too far.
Safelight Testing Frequency
Done semiannually or more frequently if problems are discovered.
Filter Replacement Schedule
Annually if turned on 24/7 or every 2 years if on 12/7.
Penetrometer Procedure
Involves exposing a fresh film at a 40-inch SID with 70 kVp and 5 mAs to achieve an OD of approximately 1.
Maximum Acceptable Safelight Fog
Less than 0.05 Optical Density units for dual-emulsion films.
Sensitometer Procedure Limits
Maximum density difference should not exceed 0.05 OD for dual-emulsion or 0.02 OD for single-emulsion films.
Leakage Testing Procedure
Turn on outside lights, turn off darkroom lights, adapt for 5 minutes, and check for white light leaks.
Processor Area Check
Inspect countertops for objects, dampness, and sharp edges to prevent artifacts.
Film Storage Conditions
Temperature between 55-75°F (14-24°C) and humidity between 30-60%.
Chemical Storage Conditions
Well-ventilated area, temperature between 40-85°F (5-30°C), not exceeding 70°F for extended periods.
Intensifying Screen Development
Thomas Alva Edison developed IS in 1896; Michael Pupin first used the film/screen combination a year later.
Luminescence
Light emission from phosphor crystals energized by x-rays.
Fluorescence
Light emitted within 10⁻⁸ seconds after crystals are exposed.
Phosphorescence
Light emitted sometime after 10⁻⁸ seconds after crystals are exposed.
Intensifying Screen Speed
The amount of light emitted by the screen for a given amount of x-ray exposure.
Intensification Factor (IF)
Exposure without screen divided by exposure with screen.
Relative Speed
Most common designation method for screen speed, typically a multiple of 100.
New mAs Formula
(Old mAs × Old Relative Speed) / New Relative Speed.
Screen Speed Names
Range from Ultrahigh/Hi-plus (300) to Ultradetail (25).
Speed Uniformity Standard
Optical density readings should not vary by more than ±0.05.
Spectral Matching
Requirement that intensifying screen phosphors (blue or green) match the color sensitivity of the film