Dark Room

A lightproof space designed for developing x-ray films into visible diagnostic images using chemical solutions or an automatic processor.

2
New cards

Latensification

An increase in sensitivity, by as much as 2-8 times, in film that has already been exposed to light or ionizing radiation compared to unexposed film.

3
New cards

Film Sensitivities

Film can be affected by excess heat, humidity, static electricity, pressure, and chemical fumes.

4
New cards

OSHA Classification

The Occupational Safety and Health Administration classifies a darkroom as a laboratory.

5
New cards

Minimum Darkroom Guidelines

The space should be clean, well-ventilated, well-organized, a non-smoking zone, and static-free.

6
New cards

Ventilation Functions

Prevents the build-up of heat and humidity and removes excessive fumes from processing solutions that may sensitize film emulsions.

7
New cards

Ventilation Placement

A ventilator duct should be placed near the floor, in either the lower portion of the entrance door or the wall.

8
New cards

Humidity Range

The proper range is 30% to 60% of relative humidity.

9
New cards

Psychrometer

A device used to monitor humidity in the darkroom.

10
New cards

Static Electricity

Proper humidity minimizes the induction of static electricity.

11
New cards

Temperature Range

Should be maintained between 65° to 75°F (18° to 24°C).

12
New cards

Processing Fumes

Considered toxic, corrosive, and potentially carcinogenic.

13
New cards

OSHA PEL for Acetic Acid

25 mg/m³ (10 parts per million [ppm]).

14
New cards

OSHA PEL for Ammonium Thiosulfate

35 mg/m³ (as ammonia) (50 ppm).

15
New cards

OSHA PEL for Hydroquinone

2 mg/m³ (0.44 ppm).

16
New cards

OSHA PEL for Phenol

19 mg/m³ (5 ppm).

17
New cards

OSHA PEL for Sulfur Dioxide

13 mg/m³ (5 ppm).

18
New cards

OSHA PEL for Glutaraldehyde

0.7 mg/m³ (0.2 ppm).

19
New cards

OSHA PEL for Silver

0.01 mg/m³.

20
New cards

Darkroom Disease

A condition where radiographers or darkroom technicians have hypersensitivity to darkroom chemicals.

21
New cards

Darkroom Disease Symptoms

Hives, severe fatigue, and impairment of the immune system.

22
New cards

Indications of Proper Ventilation

Humidity and temperature are within range and chemical exposure is well below PEL.

23
New cards

Drainage Capacity

Drains must handle 2.5 times the maximum outflow of the processor when all drains are open.

24
New cards

Darkroom Walls

Should use enamel or epoxy paints in pastel and light colors with a matte finish and have shielding equivalent to adjacent radiation areas.

25
New cards

Storage Methodology

Uses a "first in, first out" methodology with separate areas for chemicals and films.

26
New cards

Film Storage Position

Films must be stored in an upright position.

27
New cards

Passbox

A film transfer system that prohibits white light from entering.

28
New cards

Static Reduction Methods

Grounding countertops/mats, proper film handling, wearing natural fibers like cotton, and regular screen cleaning with antistatic cleaner.

29
New cards

Darkroom Lighting Types

Consists of overhead lights and safelights.

30
New cards

Overhead Lighting

Standard white light necessary for cleaning, maintenance, and emergencies.

31
New cards

Overhead Light Standards

Standard fluorescent fixtures with 2-4 48-inch tubes per 8 square feet, interlocked with the film bin.

32
New cards

Safelight

A light source emitting wavelengths to which particular films are not sensitive, placed 3-4 feet from work areas.

33
New cards

Average-sized Darkroom Safelight

Fixture with a 7.5W or 15W bulb covered by a filter like Kodak Wratten 6B or Kodak Mor-Lite.

34
New cards

Large Darkroom Safelight

Sodium vapor lamp mounted at least 6 feet above counters for bright conditions.

35
New cards

Blue-Violet-Sensitive Film

Sensitive to blue, indigo, and violet; normally uses an amber-colored safelight.

36
New cards

Orthochromatic Film

Mainly sensitive to the green spectrum plus blue-violet; uses red or magenta dye and filters like Kodak GBX or Kodak 2.

37
New cards

Panchromatic Film

Sensitive to all visible colors; uses a dark green filter like Kodak Number 7.

38
New cards

Light and Leakage Testing

Includes safelight testing and darkroom light leakage testing.

39
New cards

Safelight Unsafe Conditions

Caused by cracks/pinholes in filters, wrong bulb wattage, or sodium vapor lamp doors open too far.

40
New cards

Safelight Testing Frequency

Done semiannually or more frequently if problems are discovered.

41
New cards

Filter Replacement Schedule

Annually if turned on 24/7 or every 2 years if on 12/7.

42
New cards

Penetrometer Procedure

Involves exposing a fresh film at a 40-inch SID with 70 kVp and 5 mAs to achieve an OD of approximately 1.

43
New cards

Maximum Acceptable Safelight Fog

Less than 0.05 Optical Density units for dual-emulsion films.

44
New cards

Sensitometer Procedure Limits

Maximum density difference should not exceed 0.05 OD for dual-emulsion or 0.02 OD for single-emulsion films.

45
New cards

Leakage Testing Procedure

Turn on outside lights, turn off darkroom lights, adapt for 5 minutes, and check for white light leaks.

46
New cards

Processor Area Check

Inspect countertops for objects, dampness, and sharp edges to prevent artifacts.

47
New cards

Film Storage Conditions

Temperature between 55-75°F (14-24°C) and humidity between 30-60%.

48
New cards

Chemical Storage Conditions

Well-ventilated area, temperature between 40-85°F (5-30°C), not exceeding 70°F for extended periods.

49
New cards

Intensifying Screen Development

Thomas Alva Edison developed IS in 1896; Michael Pupin first used the film/screen combination a year later.

50
New cards

Luminescence

Light emission from phosphor crystals energized by x-rays.

51
New cards

Fluorescence

Light emitted within 10⁻⁸ seconds after crystals are exposed.

52
New cards

Phosphorescence

Light emitted sometime after 10⁻⁸ seconds after crystals are exposed.

53
New cards

Intensifying Screen Speed

The amount of light emitted by the screen for a given amount of x-ray exposure.

54
New cards

Intensification Factor (IF)

Exposure without screen divided by exposure with screen.

55
New cards

Relative Speed

Most common designation method for screen speed, typically a multiple of 100.

56
New cards

New mAs Formula

(Old mAs × Old Relative Speed) / New Relative Speed.

57
New cards

Screen Speed Names

Range from Ultrahigh/Hi-plus (300) to Ultradetail (25).

58
New cards

Speed Uniformity Standard

Optical density readings should not vary by more than ±0.05.

59
New cards

Spectral Matching

Requirement that intensifying screen phosphors (blue or green) match the color sensitivity of the film