What is the standard stimulus presentation method for Brainstem Auditory Evoked Potentials (BAEP)?

Monaural

2
New cards

Which two transducer types are recommended for BAEP stimulus delivery?

Headphones or ear inserts

3
New cards

In BAEP, masking is provided to the non-test ear using _____.

White noise at 60 dB SPL

4
New cards

What is the system bandpass for recording VEP?

1-100 Hz

5
New cards

Where is the cathode placed for Median nerve stimulation?

Between palmaris longus tendons and flexor carpi radialis, 2 cm proximal to the wrist crease

6
New cards

For the VEP montage, Channel 1 (Ch1) is _____.

LO-MF

7
New cards

List the three polarities commonly used for BAEP stimulus presentation.

Rarefaction, condensation, and alternating

8
New cards

What is the standard system bandpass for SSEP recordings?

30-3,000 Hz

9
New cards

What is the standard analysis time (time window) for VEP?

250 ms

10
New cards

In the Median SSEP montage, C5S-REF is found on which channel?

Channel 2 (Ch2)

11
New cards

What is the maximum allowable electrode impedance for VEP?

< 5 kOhms

12
New cards

For PTN SSEP, Channel 3 (Ch3) in the montage corresponds to which derivation?

CPz-Fpz

13
New cards

What is the standard system bandpass for recording BAEP?

10-30 to 2,500-3,000 Hz

14
New cards

Where is the anode placed for PTN stimulation?

3 cm distal to the cathode

15
New cards

For PTN SSEP, Channel 2 (Ch2) in the montage corresponds to which derivation?

Fpz-C5S

16
New cards

For the VEP montage, Channel 4 (Ch4) is _____.

MF-A1

17
New cards

How are Visual Evoked Potentials (VEP) presented to the patient?

Monocularly

18
New cards

SSEP waveform labeling is based on _____ and _____.

Polarity and latency

19
New cards

What is the standard analysis time for Median SSEP?

40 ms

20
New cards

For Median SSEP, Channel 3 (Ch3) in the montage corresponds to which derivation?

CPi-REF

21
New cards

What is the maximum allowable electrode impedance for BAEP?

< 5 kOhms

22
New cards

For Median SSEP, Channel 1 (Ch1) in the montage corresponds to which derivation?

EPi-REF

23
New cards

Which electrode placement system is used for BAEP recordings?

International 10-20 System

24
New cards

What is the specific stimulus pulse width (duration) for BAEP?

100 usec

25
New cards

What is the maximum stimulus rate recommended for VEP?

4/s

26
New cards

What is the range for repetitions (averages) required for VEP?

100-200

27
New cards

What is the maximum electrode impedance permitted for SSEP recordings?

5 kOhms

28
New cards

Where is the cathode placed for Posterior Tibial Nerve (PTN) stimulation?

Between the medial border of the Achilles tendon and the posterior border of the medial malleolus

29
New cards

Where is the anode placed for Median nerve stimulation?

2-3 cm distal to the cathode, or on the dorsum of the wrist

30
New cards

What is the minimum distance allowed between the patient and the VEP stimulus screen?

70 cm

31
New cards

What specific electrode type is used for SSEP?

Standard EEG disk electrodes

32
New cards

The stimulus type utilized for BAEP is _____ clicks.

Broadband

33
New cards

In the BAEP montage, what is the electrode derivation for Channel 1?

Cz-Ai

34
New cards

For the VEP montage, Channel 3 (Ch3) is _____.

RO-MF

35
New cards

What is the standard analysis time for PTN SSEP?

60 ms

36
New cards

What is the pulse width (duration) range for SSEP stimuli?

100-300 usec

37
New cards

What is the mid-size field size for VEP stimulation?

6-12°

38
New cards

BAEP waveforms are labeled in order sequence using _____.

Roman numerals

39
New cards

VEP waveform labeling follows the standard of _____ and _____.

Polarity and latency

40
New cards

Where should the patient maintain visual fixation during a VEP test?

Center of the stimulus screen

41
New cards

The number of SSEP repetitions typically ranges from _____ to _____.

Several hundred; several thousand

42
New cards

What is the standard stimulus rate for BAEP according to ACNS guidelines?

8-10/s

43
New cards

For PTN SSEP, Channel 1 (Ch1) in the montage corresponds to which derivation?

T12S-REF

44
New cards

What is the mid-size check size for VEP stimulation?

24-32'

45
New cards

What electrode type is standard for EP procedures?

Standard EEG disk electrodes

46
New cards

For the VEP montage, Channel 2 (Ch2) is _____.

MO-MF

47
New cards

For Median SSEP, Channel 4 (Ch4) in the montage corresponds to which derivation?

CPc-CPi

48
New cards

What is the recommended stimulus rate for SSEP?

3-5/s (or 3-5 Hz)

49
New cards

According to the ACNS table, what is the minimum number of replications required for BAEP?

At least 2

50
New cards

What type of electrical pulse is used for Somatosensory Evoked Potential (SSEP) stimulation?

Monophasic rectangular pulses using constant current or constant voltage

51
New cards

The standard analysis time (time window) for BAEP is _____.

10-15 ms

52
New cards

How many repetitions (averages/trials) are typically required for a BAEP study?

1,000-4,000

53
New cards

What is the specific stimulus type and contrast requirement for VEP?

High contrast (>50%) black and white checkerboard pattern

54
New cards

For PTN SSEP, Channel 4 (Ch4) in the montage corresponds to which derivation?

CPi-Fpz

55
New cards

How is stimulus intensity determined for SSEP?

The threshold that produces a consistent but adequately tolerated muscle twitch

56
New cards

In the PTN SSEP montage, CPz-Fpz is recorded on which channel?

Channel 3 (Ch3)

57
New cards

Which specific electrode placement system is used for VEP?

Queen Square System

58
New cards

What is the recommended stimulus intensity range for BAEP?

40-120 dB pe SPL

59
New cards

In the BAEP montage, what is the electrode derivation for Channel 2?

Cz-Ac

60
New cards

For Median SSEP, Channel 2 (Ch2) in the montage corresponds to which derivation?

C5S-REF

61
New cards
What is the minimum requirement for the differential input impedance of an EP amplifier?

At least 100 MΩ

62
New cards
The common mode rejection ratio (CMRR) of an EP amplifier must be at least _____ at its highest sensitivity.
80 dB (or 10,000:1)
63
New cards

With filters wide open, what is the required bandpass for an EP amplifier measured at the -3 dB points?

0.1-5,000 Hz

64
New cards
The noise level of an EP amplifier must not exceed _____ with a bandpass of 0.1-5,000 Hz.
2 uV rms
65
New cards
What is the required time (horizontal) resolution for an EP averager system?

20 us/data point or less

66
New cards
While 8-bit A-D converters are adequate, what bit-depth is preferred for its broader dynamic range in EP systems?
12-bit converters
67
New cards
An EP averager system should have a minimum of how many addresses of memory available for each channel?
At least 500 addresses.
68
New cards
What is the minimum number of channels required for performing Pattern Visual Evoked Potential (PVEP) and Somatosensory Evoked Potential (SSEP) studies?
A minimum of four channels.
69
New cards
What is the minimum number of channels required for Brainstem Auditory Evoked Potential (BAEP) studies?
A minimum of two channels.
70
New cards
What is the standard limit for chassis leakage current with the ground open for EP equipment?
Less than 300 uA rms
71
New cards
In the OR or ICU, what is the maximum allowable leakage current through patient leads to ground with 120 { VAC} applied?
10 uA rms
72
New cards
What is the effect of excessive low-frequency filtering on the timing of an EP peak or valley?
It causes the peak or valley to appear earlier.
73
New cards
How does excessive high-frequency filtering influence the recorded latency of an EP component?
It delays or prolongs the peak or valley.
74
New cards
To minimize filter-induced phase shifts, what should be the minimum ratio of high-to-low frequency filter settings?
A ratio of at least 100:1
75
New cards
Why is the use of a 60 Hz notch filter generally discouraged in EP recording?
It may 'ring' when activated by a stimulus artifact, creating a burst of 60 Hz activity that contaminates the response.
76
New cards
According to standard polarity convention, what direction is a negative event typically displayed for SSEP and PVEP studies?
Upward.
77
New cards
In BAEP studies, positive events are usually displayed in which direction?
Upward.
78
New cards
What is the typical amplitude range for calibration pulses used in EP systems?

0.5-100 uV

79
New cards
What is the primary purpose of obtaining two or more temporally independent averages (replication)?
To demonstrate that the responses are of neural and not artifactual origin.
80
New cards
EP latency replication is usually considered successful if it is within _____ of the total sweep time.
1%
81
New cards
What is the standard percentage for achieving amplitude replication between successive EP averages?
Within 15% of the peak-to-peak amplitude.
82
New cards
When establishing a new EP laboratory using reference norms, what is the minimum number of normal subjects that should be studied to verify accuracy?
At least 20 normal subjects.
83
New cards
How should age-specific norms be grouped for normative studies of EP in infants?
By month.
84
New cards
How should age-specific norms be grouped for normative studies of EP in adults?
By decade.
85
New cards
The standard gain adjustment for an EP amplifier should be in steps of no more than _____.
2.5 to 1
86
New cards
The EP amplifier must maintain specifications even in the presence of a sustained _____ differential offset.
300 mV
87
New cards
An EP system must allow for the averaging of at least how many trials?
At least 4,000 trials.
88
New cards
What essential averager mechanism allows for the exclusion of trials where the signal exceeds the limits of the A-D converter?
Artifact rejection.
89
New cards
For safety, a direct connection from the patient 'ground' to the equipment chassis should be prevented by using an _____.
Isolation amplifier (or solid-state current limiter).
90
New cards
What is the maximum allowable leakage current at the patient end of the cable when the power line ground is disconnected in clinical areas?
50 uA rms
91
New cards
What general term describes algorithms like 'smoothing' that affect amplitude/morphology but usually not latency?
Digital filtering.
92
New cards
In an EP display, what is the meaning of the abbreviation 'Input terminal 1' relative to 'Input terminal 2' regarding upward deflection?
Terminal 1 is negative relative to terminal 2 for an upward deflection (if negative is up).
93
New cards
In the documentation of EP results, what does 'derivation' refer to?
The abbreviated designations of electrode locations connected to input terminals 1 and 2.
94
New cards
What is the minimum number of years required for direct training and experience for an entry-level evoked potential technologist?
Not less than 12 months.
95
New cards
What term is used to describe the consistency of latency and amplitude measures across successive trials?
Replication.