Radiographic Exposures - Beam Limitation and Filtration

0.0(0)
studied byStudied by 0 people
0.0(0)
full-widthCall Kai
learnLearn
examPractice Test
spaced repetitionSpaced Repetition
heart puzzleMatch
flashcardsFlashcards
GameKnowt Play
Card Sorting

1/79

encourage image

There's no tags or description

Looks like no tags are added yet.

Study Analytics
Name
Mastery
Learn
Test
Matching
Spaced

No study sessions yet.

80 Terms

1
New cards

The equipment is designed with my devices that radiologists and technologist can use to do what?

- optimize the quality of the image

- reduce radiation exposure for patients undergoing various imaging procedures

2
New cards

Why have multiple features been built into new x-ray producing machines by their manufacturers?

to ensure radiation safety

3
New cards

Why have some extra characteristics have also been included into the equipment?

to meet enhanced federal regulations

4
New cards

What are the accessories available to do?

lower radiation dose for the patient

5
New cards

Measures must be taken to ensure that radiographic equipment operates safely in order to protect whom?

patients and all personnel

6
New cards

What 2 things does every diagnostic imaging system need to have?

- protective tube housing

- correctly functioning control panel

7
New cards

What is a radiographic examination table, as well as other devices and accessories, designed to do?

reduce the patient's radiation dose

8
New cards

What is a lead-lined metal diagnostic-type protective tube housing required to do?

protect the patient and imaging personnel from off-focus, or leakage radiation by restricting the emission of x-rays to the area of the useful, or primary, beam

9
New cards

How should the x-ray tube housing be constructed?

so that if radiation leaks through any portion of the housing, it leaks away from the useful beam

- measured at a distance of 1-m from the x-ray source

- does not exceed 1 mGya/hr (100 mR/hr) when the tube is operated at its highest voltage at the highest current that allows continuous operation

10
New cards

The x-ray tube is constructed in such a way that when the tube is operated at its highest voltage at the highest current that allows continuous operation it does not exceed what?

1 mGya/hr (100 mR/hr)

11
New cards

Where is the control panel or console located?

behind a suitable protective barrier that has a radiation-absorbent window that permits observation of the patient during any procedure

12
New cards

What must the control panel or console indicate?

the conditions of exposure and provide a positive indication when the x-ray tube is energized

13
New cards

What does the control panel or console have that permits the operator to assess exposure conditions?

visible mA and kVp digital readouts

14
New cards

What happens when the exposure begins?

a tone is emitted, and the sound stops when the exposure terminates

15
New cards

What does the audible sound indicate?

that the x-ray tube is energized and ionizing radiation is being emitted

16
New cards

How strong must the radiographic examination table be?

strong enough to adequately support patients whose weight is in excess of 300 pounds

17
New cards

What does a radiographic examination table frequently have? Why?

a floating table top; makes it easier to maneuver the patient during an imaging procedure

18
New cards

What should the radiographic examination table be in regards to thickness?

uniform thickness

19
New cards

Why must the radiographic examination table be as radiolucent as possible?

to absorb only a minimal amount of radiation, thereby reducing the patient's radiation dose

20
New cards

What is the radiographic examination table, frequently made of?

carbon fiber material

21
New cards

SID

Source to Image Receptor distance

22
New cards

What does the Source-to-Image Receptor Distance (SID) Indicator do?

provides a way to measure the distance from the anode focal spot to the image receptor (IR) to ensure that the correct SID is maintained

23
New cards

Is there a manual way to measure the SID?

yes, frequently a simple device such as a tape measure is attached to the collimator or tube housing so that the radiographer can manually measure the SID

24
New cards

What is another way to measure the SID?

lasers

25
New cards

How accurate does the distance and centering indicators have to be?

within 2% and 1% of the SID

26
New cards

How should the primary x-ray beam be adequately collimated?

so that it is no larger than the size of the IR being used for the examination

27
New cards

What is the light-localizing variable-aperture rectangular collimator used for?

to adjust the size and shape of the x-ray beam either automatically or manually

28
New cards

What is currently the predominant x-ray beam limitation device?

collimator

29
New cards

What do the x-ray beam limitation devices do?

confine the useful, or primary beam before it enters the area of clinical interest, and thereby limits the surface area of body tissue irradiated.

30
New cards

What is another thing the x-ray beam limitation devices do?

reduce the amount of scatter radiation in the tissue and prevent unnecessary exposure to tissues not under examination

31
New cards

What are the benefits of restricting x-ray field size to include only the anatomic structures of clinical interest?

- significant reduction in patient dose because less scatter radiation is produced

- improves the overall quality of the radiographic image

32
New cards

What are the light-localizing variable-aperture rectangular collimators?

- construction

- components

- reduction of off-focus, or stem, radiation

- confinement of the radiographic beam

- skin sparing

- minimizing skin exposure to electrons produced by photon interaction with the collimator

- luminescence

- coincidence between the radiographic beam and the localizing light beam

- positive beam limitation (PBL)

- alignment of the x-ray beam

33
New cards

Light-Localizing Variable-Aperture Rectangular Collimators: construction

box shaped and contains beam defining system

34
New cards

Light-Localizing Variable-Aperture Rectangular Collimators: components

- 2 sets of adjustable lead shutters mounted with in the device at different levels

- a light source to illuminate the x-ray field

- a mirror to deflect the light beam toward the patient to be imaged

35
New cards

Where is the first set of adjustable lead shutters?

near the tube window

36
New cards

What does the first set of adjustable lead shutters do?

eliminate off focus radiation

37
New cards

What is off focus radiation?

x-rays emitted from parts of the tube other than the focal spot and at various angles from the x-ray tube window

38
New cards

Where is the second set of collimator shutters mounted?

below the light source and mirror and further confine beam to area of interest

39
New cards

What is skin sparing?

reduce skin exposure to electrons produced by photon interaction with collimation

40
New cards

What should the SSD be for skin sparing?

at least 15 cm below the collimator for fixed radiographic equipment

41
New cards

SSD

Source to Skin Distance

42
New cards

What should the collimator light provide?

enough luminescence to adequately outline the margins of the radiographic beam

43
New cards

What should the light field be?

within the 2% SID with the corresponding beam

44
New cards

PBL

positive beam limitation

45
New cards

What is positive beam limitation?

automatic collimators that restrict the beam to be no larger than the size of the image receptor

46
New cards

What should the PBL be?

within 2% of the actual exposed field

47
New cards

What must the radiographer ensure with the PBL?

that collimation is adequate by collimating the radiographic beam so that it is no larger than the IR

48
New cards

What is the x-ray beam limitation device called a cone used for?

radiographic examinations of specific areas such as the head?

49
New cards

What does the cone look like?

flared metal tubes and straight cylinders

50
New cards

When could the beam-defining cone be used?

dental radiography

51
New cards

What is filtration?

the removal of low energy photons from the beam

52
New cards

Types of filtration

inherent and added

53
New cards

What is the purpose of filtration?

- increase the mean (average) energy of the beam

- fewer photoelectric interactions

- markedly reduces patient dose

54
New cards

What is the downside to filtration?

fewer photoelectric interactions

55
New cards

What is inherent filtration?

always present

56
New cards

What are the components of inherent filtration?

- glass envelope encasing the x-ray tube

- insulating oil surrounding the tube

- glass window in the tube housing

- equivalent to .5 mm aluminum

57
New cards

What are the components of added filtration?

- thin sheets of aluminum (or equivalent)

- added outside of glass window above the shutters

58
New cards

What is the minimum aluminum equivalent required for total filtration for fixed radiographic units operating at above 70 kVp?

2.5 mm aluminum equivalent

59
New cards

Why is aluminum selected as a filter material for filtration for general diagnostic radiology?

it effectively removes low-energy x-rays from a heterogeneous x-ray beam without severely decreasing the x-ray beam intensity

60
New cards

In filtration for general diagnostic radiology, what must a diagnostic x-ray beam always be?

adequately filtered

61
New cards

HVL

half-value layer

62
New cards

What is the half-value layer?

the thickness of a designated absorber required to decrease the intensity of the primary beam by 50% of its initial value

63
New cards

Why must the HVL be measured?

to verify that the x-ray beam is adequately filtered

64
New cards

How often should a radiologic physicist obtain the HVL measurement?

at least once a year and also after an x-ray tube is replaced or repairs have been made on the diagnostic x-ray tube housing or collimation system

65
New cards

How is HVL expressed?

millimeter of aluminum

66
New cards

HVL is a measure of ________ quality, or effective energy of the x-ray beam.

beam

67
New cards

A minimal HVL is required at a given ______.

kVp

68
New cards

What does increasing the amount of tube filtration do to the x-ray beam? Why?

increases the x-ray beam quality because there is a greater percentage of x-rays that have high energy than low energy, and it decreases the quantity of x-rays or x-ray emission

69
New cards

What are the compensating filters made of?

aluminum, lead-acrylic, or other suitable materials

70
New cards

What are the compensating filters used for?

to accomplish dose reduction and uniform imaging of body parts that vary considerably in thickness or tissue composition

71
New cards

How do compensating filters work?

they partially attenuate x-rays that are directed toward the thinner, or less dense area while permitting more x-radiation to strike the thicker, or more dense, area

72
New cards

Types of compensating filters

- wedge filter

- trough, or bilateral wedge filter

73
New cards

Wedge compensating filter

knowt flashcard image
74
New cards

Trough, or bilateral wedge compensating filter

knowt flashcard image
75
New cards

What is exposure reproducibility?

consistency in output in radiation intensity for identical generator settings from one individual exposure to subsequent exposures

76
New cards

What should the variance be for exposure reproducibility?

5% or less is acceptable

77
New cards

How might exposure reproducibility be verified?

by using the same technical exposure factors to make a series of repeated exposures and then, observing with a calibrated ion chamber, how radiation intensity varies

78
New cards

What is exposure linearity?

consistency in output radiation intensity at selected kVp settings when generator settings are changed from one mA and time combination to another

79
New cards

What is linearity?

the ratio of the difference in mSv/mAs or mR/mAs values between two successive generator stations to the sum of these mSv/mAs or mR/mAs values; must be less than 0.1

80
New cards

How much can the linearity vary when the settings are changed from one mA to a neighboring mA station?

the most that linearity can vary is 10%