1/79
Looks like no tags are added yet.
Name | Mastery | Learn | Test | Matching | Spaced |
---|
No study sessions yet.
Dynamic
visualizing the motion and function of internal structures
Static
captures still images of the body
Dark adaptation
required to permit the RODS of the eye to become activated
Visual acuity
controlled by the CONES of the eye
Where are cones located in the eye?
Center of the retina
What are cones used for?
Used for perception of small objects and bright/high light level conditions
Scotopic
vision in dim light
Photopic
vision in daylight or other bright light
Photopic visual acuity is ____times greater than scotopic acuity
10
Typical basic fluoroscopic image chain and equipment used
Fluoro x-ray tube and IR are mounted on a C-arm to maintain alignment
What does the C-arm permit?
- Permits the IR to be raised and lowered to vary the beam geometry for max resolution while the x ray tube remains in position
- Permits scanning the length and width of the table
Carriage
the arm that supports the equipment suspended over the table
what does the carriage include?
- Image intensifier
- X-ray tube
- Controls for power to drive
- Spot image selection
- Tube shutters (collimation)
- on/off dead man switch
- Magnification selection
- Reverse image search
- Brightness
What must be done for an exposure be taken when using the carriage?
Exposure cannot be taken until the carriage is returned to a full beam intercept position
Fluoro xray tubes : mA range
0.5 - 5.0 mA
The tube target is fixed to prevent a SOD of...
less than 15 inches or 38 cm
Tube can be operated by a ...
foot pedal (dead man type)
What do the tube have that permit maintenance of close collimation?
electronically controlled shutters
How much can image intensifiers increase image brightness?
500- 8000 times
Brightness of fluoro depends on
anatomic part
kVp
mA
What kVp and mA is preferred?
high kVp
low mA
Size distortion is caused by
OID
Shape distortion is caused by
geometric problem in the shape of the image intensification tube
How does image intensifier work?
1. Beam exits patient
2. strikes input screen
3. input screen absorbs xray photons and emits light photons
4. light photons interacts with photocathode and absorbs the light photons and emits electrons (photoemission)
5. Electrons are then focused and accelerated from the photocathode toward the anode and output screen
6. Output screen absorbs these electrons and emits light photons
What is the input screen composed of?
cesium iodide
How much of the incident beam do Csl phosphors absorb?
66%
What shape is the input screen and why?
concave shape b/c helps minimize image distortion
What is the photocathode made of?
cesium and antimony compounds
Why is the photocathode in contact with the input screen?
to prevent divergence
How are the electrons focused and accelerated from the photocathode toward the anode and output screen?
potential difference (25 kV) that exists between each side
What are electrostatic lenses?
series of charged electrodes used to accelerates and focus electrons to the anode and output screen
What is the output screen made of?
Zinc cadmium sulfide
each photon electron that hits the output screen results in _____________more light photons than before
50 - 75 x
What part of the image intensifier output screen has better resolution?
center
(brighter image and less geometric distortion )
What is flux gain?
measurement of the increase in light photons due to the conversion efficiency of the output screen
Flux gain formula
output screen (light photons) / input screen (light photons)
What is minification gain?
result of the same number of electrons that were produced at the large input screen being compressed into the area on the small output screen
Minification gain formula
input screen diameter^2 / output screen diameter^2
What is brightness control?
used to automatically maintain satisfactory fluoroscopic image brightness and contrast
What does ABC stand for?
Automatic brightness control
What does ABC do?
Maintains the brightness of the image by automatically adjusting the exposure as necessary according to subject density and contrast
- has a slow response time
What maintains optimum fluoroscopic image brightness during fluoroscopic
operation - Very important in maintaining image quality?
Electronic feedback loop between image intensifier and xray tube
What is vignetting?
a fall off in brightness at the periphery of an image
Csl image intensifier capable of...
2lp/mm
Digital flouro capable of...
3lp/mm or better
What is interrogation time?
length of time required for the generator to come on and acheive the necessary kVp and mAs levels
What extinction time?
time required to shut the generator down in preparation for the next pulse
continuous fluoro
allows for real time imaging but delivers large dose for patient
pulsed fluoro
delivers bursts of radiation at set intervals, reducing fluoro times
duty cycle
the fraction of time between the interrogation and the extinction times (xray tube is energized)
thin film transistors (TFT)
combined with pulsed technology
replaces the image intensifier
most common viewing system
flat panel TFT system
Flat panel TFT system includes...
a charged coupled device (CCD) that stores from light photons then transmitting the image to a monitor for viewing
200 - 400 microns DEL size (2-3 lp/mm)
Pixel
smallest element of digital image
matrix
2d series of square boxes composed of pixels
Digital pixel size?
1024 x 1024 pixels
pros of digital flouro
- Post processing (results in enhanced contrast resolution)
- Speed of acquisition
- 1024 x 1024 image matrix (high resolution)
- Reduces patient dose with both dynamic and static images recorded at reduced doses
what are the two designs of digital flat panel fluoro systems?
Fluoro tube below patient
Fluoro tube above patient
Multiple detector flat-panel systems:
- Tube is under the table and detector is in carriage above table
- 2 additional detectors located to perform both upright and recumbent
- Can switch electronically between fluoro and static
Single Detector flat- panel system:
- Used for all imaging
- More cost effective
- Requires single x-ray tube to perform both static and dynamic images
- Tube is mounted above the table top
What is a feature of digital radiation protection that retains the last image?
Last-image hold feature
What are some post processing features in digital radiation protection?
Window level, window width, filtering
How much can digital radiation protection reduce patient exposure compared to conventional methods?
Up to 90%
Radiation protection for patient:
Use largest field of view to area of interest
Highest kVp
Low dose modes
Pulsed fluoro
Tabletop exposure rate should not exceed...
10R/min
(At that rate skin redding can happen after 20 minutes ( 10 minutes under boost mode))
Mobile minimum SOD
12 inches or 30 cm
Fixed minimum SOD
15 inches or 38 cm
Magnification mode __________ dose with image intensifier. (2x dose increase per mag field)
increases
Lead apron pb/eq
at least 0.5 mm
If hands must be placed in the primary beam, lead gloves of atleast ____________Pb/eq must be worn
0.5 mm pb/eq
Tech should keep what part of their body with the lead apron facing ______?
front of their body, facing the patient and tube
highest scatter happens @ _________ degree angle incident beam
90 degree angle
bucky slot cover
0.25 mm Pb/eq
Total brightness gain
measurement of the increase in image intensity achieved by a image intensification tube
total brightness gain formula
minification gain x flux gain
The greater the voltage, the greater the acceleration, what happens to the focal point?
closer the focal point moves toward the input screen
triple field intensifiers
14, 7, 5 inch fields
dual fields
9, 6 inch field
Patient exposure _________ when a __________ viewing mode is selected
increases, smaller
magnification formula
Input Screen Diameter / diameter of input screen used during magnification