1/172
Looks like no tags are added yet.
Name | Mastery | Learn | Test | Matching | Spaced | Call with Kai |
|---|
No study sessions yet.
Transmission,
goes through the body
Absorption
, high subject density, whites on image bones, hardware, good for image quality, photoelectric effect
Scatter
Unwanted noise, useless to image
Compton scatter
Contributes most to scatter effects image quality, increases patient dose, increases tech dose, decrease image quality
Compton scatter
Incident x-ray photon interacts with outer shell electron, the photon scatters and has less energy, recoil electron
What does the effect scatter have on image?
decreases contrast, low contrast image, has fog or noise on image, more shades of grey
3 factors that impact scatter production?
Patient thickness
kVp
field size/beam size
Part thickness
Increasing the size of patient or the body part
increases scatter
increases tissue density
- increases scatter More matter more scatter
Field size
Increasing the field size exposed
increased scatter
kVp
Increasing kVp= decrease in photoelectric and Compton interactions
there are more Compton interactions than photoelectric interactions.
Scatter photons has higher remaining energy when higher kVp is used
Increase kVp=Increased scatter
Transmission occurs but more scatter
Mass effect on scatter?
No affect
kVp
Lower contrast images do not have differences in shades more grey
Scatter control
Beam restricting devices:
affects scatter production
decrease the x-ray beam field size and the amount of tissue irradiated, reducing the amount of scatter radiation produced in the patient
Radiographic grids-
Affects scatter clean up after production
Used to improve radiographic image quality by absorbing scatter radiation that exists the patient, reducing the amount of scatter reaching the image receptor
Beam restriction
aslo know as Collimation
Beam restriction two main purposes
Limit patient exposure
reduce scatter production installed outside of the x-ray tube housing changes the shape and size of the primary beam
increase collimation= decrease field size
decrease collimation= increase field size
LOCATED BELOW THE TUBE HOUSING, change the size to what we want
Increase collimation
decrease field size
decrease collimation
Increase field size
Increasing beam restriction
decrease patient dose
decrease the scatter produced in the patient
decreases scatter reaching the IR
increased radiographic contrast
Increase collimation
Increase intensity of scatter radiation
Significant increase collimation
Quantum noise increases as the number of photons reaching the IR decreases
mAs settings must be increased to provide sufficient photon quantity
Not adjusting KVP because it would impact image contrast
Types of beam restriction devices
aperture diaphragm
collimator
cylinder
cones
lead mask/shield
Automatic Collimators (PBL)
Aperture Diaphragm
A flat piece of lead (diaphragm) that has a hole (aperture) in it and is placed directly below the x-ray tube window
very easy to use
can be cut out of lead
DOWNFALLS- field size cannot be adjusted
because of the proximity to the radiation source their is a large area of unsharpness that surrounds the image, because the device is so close to the tube
Cones and Cylinders
an aperture diaphragm with an extended flange attached to it
slide onto the tube directly below the window
The flange can be made to telescope to increase its total length
typically fit to a specific equipment easy to use
Cones and Cylinders
Limit the unsharpness surrounding the image
come in limited sizes
May not be interchangeable among the tube housings
Cylinders are more useful than cones
LIMITED IN SIZE AND SHAPE THAT CAN BE PRODUCED
NOT INTERCHANGABLE TO CERTAIN TUBE HOUSINGS
CYLINDERS ARE MORE USEFUL than cones
Cones and Cylinders
Produce a circular projected field
Collimators Located?
Located immediately below the tube window
Collimators
2 sets of lead shutters
First lead shutter
entrance shutters
located directly below tube
limit the x-ray beam just as much as a aperture diaphragm would
Second lead shutter
Adjustable shutters
located 3-7 inches below the tube
consists of both longitudinal and lagitunal leads or blades, adjustable
limits off focus radiation
off focus radiation
adds noise to the image
x-rays not aligned with regular beam
x-rays created anywhere outside of focal spot
Collimators equipped with
A white light source and a mirror to project a light field onto the patient which indicated the exposure area
An x-ray field measurement guide in case of light failure
A plastic template with crosshairs to indicate centering
Collimation should match what?
field guide
light bulb and mirror could?
come out of alignment due to rough handling/ off centering. Can put plastic template to show where crosshairs should be
Automatic collimators
Positive beam limiting (PBL) devices
PBL
Automatically limits the size and the shape of the primary beam to the size and shape of the image receptor
Override mechanism
overrides the positive beam limitation
Lead blocker/ Masks
Similar to an aperture diaphragm
limited to that shape and size
cut out the shape of body part
can be placed behind someone for thicker body parts/preventing scatter from reaching the image receptor
Lead blocker
can make it for the patient/ cut out/ or a piece of rubber to be placed in the patients back.
what does a Lead blocker do?
CLEANS UP SCATTER
why is the Lead blocker is placed behind the thick body part
cleans up scatter
Scatter produced where?
in the patient
radiographic grids do what with scatter?
clean up scatter
Who invented the first grid
Gustav bucky
the first grid was a
Cross hatch grid, most effective means of limiting scatter reaching image receptor
Hollis potter grids
LINEAR STRIPS ONLY
thinner strips
Potter Bucky diaphragm- moved grid during exposure to blur grid lines
moved during exposure to blur grid lines
thinner stripes less you would see grid lines
lead does what?
absorbs radiation
Radiographic grids are?
◦Very thin lead strips absorb scatter from the patient before it can reach the IR
◦Placed between patient and IR
◦Typical grid use:
◦Part >10 cm
◦kVp above 60
Radiographic grids are placed?
between patient and image receptor
Grids do what
improve contrast
Grid setback?
adjust technique, increase mAs,
Grids are used?
10 cm or greater
Grid kVp that is required to use?
kVp above 60
Grid size to use for a chest x-ray?
30 cm or greater
Grids why we increase mass?
reduce the transmitted photons and still let some scatter photons pass through. They are not perfect thats why we increase mass
what happens if you do not use a grid?
scattered photons reach the IR and add unwanted exposure
Can grids be removed?
yes most remove in and out of tables.
Grids size?
mostly 14 by 17 inches,
Thickness of a grid?
a quarter of an inch thick
Grids range in sizes!
8 ×10 or 17×19 size
The patients body is the what?
SOURCE OF SCATTER!!!
Grids are used to?
clean up and not put any unnessarsary information on image
when using film it is
important to use grids! shows more detail less grey!
What are the two most common material in grids?
aluminum and plastic!
they are the spacing
Strips are made of?
lead
The interspace material can be made of what?
aluminum or plastic
Why does lead absorb radiation
high atomic number
easy to shape less expensive
Aluminum or plastic
radiolucent, easy to use and durable
Front and back of grid are covered by?
aluminum to protect them
what are the grid dementions?
height (H)
thickness lead strip (T)
distance between strips (D)
What is the T
the thickness of the lead stip how thick or wide
what is D
measures the amount of space between one stip of lead to the other, represents the thickness between the interspaces material
What is H
the height or how tall
What is the height of the grid or the distance between the lead strips?
Grid ratio
What is grid ratio related to?
efficeincy of the grid
Grid construction is described by?
grid ratio
grid frequency
grid frequency
number of grid lines between a certain distance, inch or centimeter
Grid ratio
compares height of the distance to the grid directly related to the efficiency of the grid
Grid ratio also known as
GR
Grid ratio
Height of lead strips (H)
Distance between strips (D)
◦Ranges: 5:1 to 16:1
◦As the ratio increases, so does the efficiency of scatter cleanup
◦
Ex. The lead strips of a grid are 2mm high and distance between is 0.2mm. What is the grid ratio?
2mm (H) = 10
0.2 mm (W)
This grid has a 10:1 grid ratio.
Grid ratio is determined by
many factors of the patient
Grid ratio lowest?
5:1 less clean up
Grid ratio higher?
16:1 more clean up reduces scatter more
we can determine grid radio by what?
the equation!
H/W=GR
height divided by width equal to the grid ratio
The more lead in grid the more what?
more clean up will happen
how is the grid ratio changed?
Changing the height of strips or the distance between them changes the grid ratio - the leeway for scatter changes.
Increasing the GR does what?
Increase scatter cleanup, imporve image quality.contrast
As we increase ratio it will?
decrease the angle of the openings and be absorbed
decrease ratio
more opening to be less absrobed
What describes the number of lead lines per unit length?
Grid frequency
Grid frequency is?
Number of lead lines per unit length (cm/mm/ inches)
◦Ranges: 60 – 110 lines/ inch (25-45 lines/cm)
◦Most common: 85 -103 LPI
Grid frequency varies in ranges of?
60-110 lines/inch (25-45 lines/cm)
Most common lines per inch for grid frequency?
85-103
Grid frequency math
get the thickness of one line strip, then space between and divide by 1 inch which gets the line pairs or lead strips in one inch of the grid.
Thicker strips
Greater frequency
Relationship between grid ratio, frequency, and about of lead content in grid?
increase grid ratio to same frequency the lead content will increase with more scatter production.
However Relationship between grid ratio, frequency, and about of lead content in grid?
If grid frequency is increased with same grid ratio, their will be less lead content because the width of the inner space or thickness have been decreased. DECREASES LEAD CONTENT AND SCATTER PRODUCTION.
thicker strips
more scatter clean up
Linear grid pattern?
lead lines wun in only one direction
allows angulation of entral ray along length of lead strips because only absorbs scatter in one direction
What is the most popular grid pattern?
linear
Crossed grid pattern is?
lead lines run at right angles to one another