Term used to describe deterioration of radiographic image
Radiographic noise
Random background info that is detected but does not contribute to image quality
Radiographic noise Radio
Radiographic Noise is caused by what?
# of x-rays used to expose the PT or mAs
The limited absorption efficiency of x-rays
The randomness of x-ray to light conversion
Is the term given to nonuniform or nonhomogeneous appearance of the recorded image more commonly associated w/ high speed recording
Radiographic Mottle
Attributed to random absorption of the x-ray quanta absorbed by the IR
Quantum mottle
What is the function of AEC ?
Eliminate the need for radiographers to set exposure time
What is the only factor that AEC controls?
Time
Radiographers must know the exact position of the ionization chambers for what reason?
Positioning purposesZthe
The x-ray must pass through the PT and tabletop before or after striking the ionization chamber?
Before
Using poor positioning will result in what when using AEC?
Increase repeat rate
Exposure controls should not be used to compensate for PT part thickness by doing what?
Changing kVp
Why should tight collimation be avoided near ionization chambers in AEC?
B/c if primary beam is collimated from an active chamber, the chamber operates as if the tissue is extremely dense.
Wider collimation will produce scatter radiation that will do what ?
Undercut the PT
Undercutting will cause AEC to terminate exposure early which will have what cause on the image?
Cause some areas of the image to be too light
Is the length of time necessary for the AEC to respond to the ionization and signals to terminate exposure
Minimum response time
What it the minimum response time for modern AEC?
0.001 seconds
Back-up times should be set… of the expected manual technique
150%
What is the function of the back-up timer?
Protect the unit
US public law 90-602 requires that generators automatically terminate AEC exposures at?
600 mAs or 60 kilowatt seconds
This operate by spring action similarly to a hand-wound alarm clock. These timers are not very accurate, and they are not used on modern equipment
Mechanical Timers
These timers use a synchronous motor that operates at 60 rps. Their shortest possible exposure is 1/60 s
synchronous timer
This type of timer operates on an electronic resistive-capacitive circuit based on the time required to charge a capacitor. These timers are accurate, allow exposures as short as 1 millisecond and can be used for serial radiography
Electronic timers
These timers are electronic timers that monitor the product of mA and time, and terminate the exposure when proper mAs is reached
mAs timer
What test is used to check timers and rectifiers?
Spinning top test
These devices w/ very accurate internal clock based on a quartz-crystal oscillator are used to check modern day timing equipment
Solid state radiation detector
Sums the voltages received from each cell and divides by the number of cells that have been activated. When the appropriate voltage for a diagnostic quality exposure is reached, it terminates the exposure
Operations amplifier
When should the density controls be used?
Should be used when the configuration of the ionization chamber cells cannot be adapted to the necessary positioning
How can changes in subject density result in AEC problems?
AEC problems with subject density occur whenever an unexpected density is present or when an expected density is lacking.
What do ionization chambers measure the exposure to?
IR
The automatic exposure control (AEC) will produce a diagnostic exposure of tissue that is placed where on the chamber?
Over the chamber
What does the radiographer need in order for the automatic exposure control (AEC) to produce quality results?
Good positioning Skills
When the correct voltage for a diagnostic quality exposure is reached it is terminated by what?
Operation amplifier
What automatic exposure device is rarely used in present-day radiography?
Phototimer
When there is a problem with density and contrast in an automatic exposure system what is usually the cause?
unexpected density