1/116
Looks like no tags are added yet.
Name | Mastery | Learn | Test | Matching | Spaced | Call with Kai |
|---|
No study sessions yet.
regulatory agencies
department of health and human services
national center for devices and radiological health
food and drug administration
occupational safety and health administration
environmental protection agency
nuclear regulatory commission
2 impactful government actions
safe medical devices act (SMDA)
mammography quality standards act (MQSA)
ACR
american college of radiologists
AAPM
american association of physicists in medicine
DICOM sets standards for
images
HL-7 sets standards for
transmission of written data
goal of quality control
to calibrate and monitor technical equipment to maintain image quality and ensure radiation exposure as low as reasonably achievable
follow standards set by ACR and AAPM
ACR recommendations for digital QC
must document policies and procedures for monitoring and evaluating equipment
QC program should be designed to maximize quality and accessibility of diagnostic information
components of QC
acceptance testing
diagnosis of deviation from acceptable equipment tolerances
documentation of corrections made
acceptance testing
completed upon installation
confirms equipment is functioning according to specifications set by the vendor
establishes a baseline for performance
people involved in acceptance testing
service engineer & bio med
IS and PACS administrator
medical physicist
manager
performance monitoring
evaluates performance of generator and tube
simple to complex equipment can be used
ACR requires medical physicist to perform tests
most testing is done annually
what is used for monitoring equipment performance
ion chamber dosimeters
gas filled chambers
newer solid-state dosimeters
small and accurate
measures the intensity of xr beam accurately
FDA mandates on the operating console
control booth walls
min of 7.5 sq ft with walls 7 inches high
exposure switch
must be fixed and a minimum of 30 inches from booth wall
control booth window
1.5 mm of lead equivalent
no exposure with the door open
xr must scatter 2 times before entering the control booth
tube support mechanism
strong and counterbalanced with immobilization locks
metal housing
leakage less than 100 mR/hour at 1 meter
max tabletop thickness
1 mm of Al equivalent
high voltage generators
convers low V to kV
step up transformers for kV
step down for tube filament
rectifiers convert incoming alternating V to pulsating direct current
requires less space and less power
ripple for high frequency generators
less than 1%
timer accuracy
if timer is off, mAs is off
tested annually
tested with spinning top or motorized synchronous top
back up timer
should terminate exposure after 6 seconds or 600 mA
what equipment is tested with a spinning top
single phase equipment
what equipment is tested with a motorized synchronous top
3 phase equipment
spinning top test
circular steel/lead disc with a small hole on the periphery
can indicate a problem with a rectifier or the timer
rectifier issue with spinning top test
you see ½ of the expected dots or arc angle
timer issue with spinning top test
results are off by a few dots
single phase rectified equipment with spinning top test
produces dots
1 second exposure should produce 120 dots
3 phase rectified equipment with spinning top test
produces an arc
1 second exposure should produce an image with 360 degree arc
how to calculate expectation of dots/arc for spinning top test at >1 second
multiply the seconds by the expected dots/arc for 1 sec
EX: 0.5 sec x 120 dots = 6 dots should be seen on properly working equipment
line focus principle
when target angle is less than 45 degrees, the effective focal spot is smaller than the actual focal spot
focal spot size determines
resolution
smaller FSS = better resolution
blooming
FS size increases as the tube ages and with high mA usage
causes electron cloud to swell
unwanted change in actual focal spot size during a change in exposure
how often is FSS tested
annually
tools for FSS testing
pinhole camera
star or slit camera
line pair resolution tool
pinhole camera
CR is aligned through metal sheet with 0.03 and 0.075 mm holes
image of focal spot is displayed using a long exposure
the size of the effective focal spot is measure using a micrometer
blooming change with tube current
higher tube current = more blooming
which is preferred between slit and star cameras
slit cameras
slit camera
metal template with series of slits in groups of 3
crosswise and lengthwise
what do slit and star cameras measure
resolution and estimated focal spot size
line pair resolution tool is used in conjunction with
slit and star cameras
line pair resolution tool measures
spatial resolution is measured in lp/mm and corresponds to a certain focal spot size
NEMA specifications for FS variations
FS smaller than 0.8mm may be 50% larger variations
FS middle 0.8mm - 1.5mm may be 40% larger variations
FS larger than 1.5mm may be 30% larger variations
what is kV and mA accuracy
stability of the relationship between kV or mA and the resulting exposure intensity
kV accuracy must be within
± 4 kVp OR ± 5% of the kV output
EX: if you expect 80 kVp and get exposure with 86 kVp is it allowed?
6 kVp is > the 4 allowed, so no
6/80 = 0.075 7.5% which is higher than 5%, so no
how is kV accuracy measured
old
WI test cassette
new
digital kVp meter
more accurate and easier
linearity
sequential increases in mA should produce the same increase in the exposure rate
linearity requirement
exposures must not vary more than ± 10%
reproducibility
output intensity must be consistent from one exposure to the next
same technique should produce the same exposure
reproducibility requirement
must be within ± 5%
gassy tube
tube with fluctuations in mA readings
half-value layer is an expression of
beam quality
filtration of the primary beam
filter absorbs low energy xrs
improves image quality
reduces pt dose
increases average beam energy
expressed in Al/Eq
how is filtration measured
computerized dosimeter provides a HVL readout
NCRP requirement for filtration
for equipment operating above 70 kVp, minimum of 2.5 mm Al/Eq total filtration is required
collimation
beam restrictors limit the size of the XR field
improves image quality
less scatter
improved contrast
reduces pt dose
should never be larger than the IR
collimation congruency
measures how well a collimator regulates
NCRP requirement for collimation congruency
accurate within ± 2% of the SID
tested annually
how is collimation congruency tested
8 or 9 penny test tool
if variance is smaller than the penny it is acceptable
central ray accuracy is measured with
beam alignment tool
plastic cylinder with metal washers or BBs
central ray accuracy requirement
must be accurate within ± 1% of the SID
40 inch SID = 0.4 inch variance
72 inch SID = 0.72 inch variance
tested annually
distance accuracy is measured by
measuring devices and detents are checked with a tape measure
tube angles are checked with a protractor
distortion is checked with
wire mesh test
wire mesh test
mesh is placed on cassette
dark areas indicate poor contrast
can show pincushioning in fluoro
pincushion distortion
warped image appearance from curved surface of input screen to flat surface of output screen
exposure requirements in fluoro
must not exceed 10 mR/min at tabletop
entrance dose must not exceed 5R/min
how are exposure requirements tested in fluoro
physicists measures output with a calibrated ion chamber
artifacts in fluoro
veiling glare
vignetting
distortion
pincushion
s-distortion
required thickness for primary barrier of the housing in fluoro
2 mm Pb/Eq
required thickness for bucky slot cover in fluoro
0.25 mm Pb/Eq
required thickness for fluoro curtain
0.25 mm Pb/Eq
timer requirement for fluoro
5 minutes
sourced to table distance requirement for stationary fluoro units
15 inches
sourced to table distance requirement for mobile fluoro units
12 inches
field uniformity checks for
checks the image visually for defects
ghosting
inadequate CR erasure
loss of lamp intensity or erasure time too short
residual charge is trapped in DEL of digital IR
off-focus/STEM radiation
examples of CR quality control tests
dark noise
uniformity
spatial accuracy
erasure throroughness
dark noise is AKA
intrinsic noise
dark noise test
scans an unexposed plate
should come out as a uniform lighter gray color
uniformity test
expose a plate and run it through the scanner
should come up as a uniform dark image
spatial accuracy test
wire mesh is placed on the CR and exposed and ran through the scanner
should have no distortion
erasure thoroughness test
expose a CR with a step wedge, run through scanner and erase, run through scanner again
after the second time through the CR shouls be blank
AAPM proposed standard for exposure index
to use deviation index numbers
what is the weakest link in the imaging chain
the monitor
ACR and FDA classify monitors as
primary monitors
secondary monitors
primary display monitor
diagnostic monitor
used by the radiologist for interpretations minimum of 2K (2048×2560) required
minimum of 4K for mammography
secondary monitors
technologist review stations
minimum of 1K (1024×1280) required
not considered diagnostic quality
types of viewing monitors
cathode ray tube
flat panel
LCD or Plasma
how often must viewing monitors be cleaned
monthly
cathode ray tube monitors
resolution deteriorates over time (blooming)
higher reflectance
better luminance
lower resolution
pincushion, barrel, and skew distortions
veiling glare cause by scatter of the phosphor layer
flat pannel liquid crystal display (LCD)
layer of liquid crystals sandwiched between two transparent electrodes
used for laptops and flat panel monitors
advantages of flat panel LCD
uses less power
better resolution and contrast
faster refresh rate
higher level of luminance longer
no distortion from:
pincushion
barrel
skew
veiling glare
drawbacks of flat panel LCD
limited viewing angle and more easily damaged
flat panel plasma monitors
acctive array of tiny fluorescent bulbs emit light from each pixel location
advantages of flat panel plasma monitors
thinner than an LCD
higher luminance output
wider viewing angle than LCD
drawbacks of flat panel plasma monitors
more expensive
shorter life
ACR remmonendations for flat panel plasma monitors
monthly performane evaluation
AAPM recommendations for flat panel plasma monitors
acceptance testing evaluation
annual physicist evaluation
daily/monthly/quarterly evaluation by QC tech
types of quality testing on display monitors
luminance
illuminance
resolution
noise
dead and stuck pixels
luminance for display monitors
ability to display different shades of brightness
rate of light emitted from a source
ACR requires min brightness of 250 lumens
illuminance for display monitors
rate of light striking the surface
max ambient lighing in a reading room must be less than ¼ normal lighting