Quality Control

0.0(0)
studied byStudied by 0 people
0.0(0)
full-widthCall with Kai
learnLearn
examPractice Test
spaced repetitionSpaced Repetition
heart puzzleMatch
flashcardsFlashcards
GameKnowt Play
Card Sorting

1/116

encourage image

There's no tags or description

Looks like no tags are added yet.

Study Analytics
Name
Mastery
Learn
Test
Matching
Spaced
Call with Kai

No study sessions yet.

117 Terms

1
New cards

regulatory agencies

department of health and human services

national center for devices and radiological health

food and drug administration

occupational safety and health administration

environmental protection agency

nuclear regulatory commission

2
New cards

2 impactful government actions

safe medical devices act (SMDA)

mammography quality standards act (MQSA)

3
New cards

ACR

american college of radiologists

4
New cards

AAPM

american association of physicists in medicine

5
New cards

DICOM sets standards for

images

6
New cards

HL-7 sets standards for

transmission of written data

7
New cards

goal of quality control

to calibrate and monitor technical equipment to maintain image quality and ensure radiation exposure as low as reasonably achievable

follow standards set by ACR and AAPM

8
New cards

ACR recommendations for digital QC

must document policies and procedures for monitoring and evaluating equipment

QC program should be designed to maximize quality and accessibility of diagnostic information

9
New cards

components of QC

acceptance testing

diagnosis of deviation from acceptable equipment tolerances

documentation of corrections made

10
New cards

acceptance testing

completed upon installation

confirms equipment is functioning according to specifications set by the vendor

establishes a baseline for performance

11
New cards

people involved in acceptance testing

service engineer & bio med

IS and PACS administrator

medical physicist

manager

12
New cards

performance monitoring

evaluates performance of generator and tube

simple to complex equipment can be used

ACR requires medical physicist to perform tests

most testing is done annually

13
New cards

what is used for monitoring equipment performance

ion chamber dosimeters

  • gas filled chambers

newer solid-state dosimeters

  • small and accurate

  • measures the intensity of xr beam accurately

14
New cards

FDA mandates on the operating console

control booth walls

  • min of 7.5 sq ft with walls 7 inches high

exposure switch

  • must be fixed and a minimum of 30 inches from booth wall

control booth window

  • 1.5 mm of lead equivalent

no exposure with the door open

xr must scatter 2 times before entering the control booth

15
New cards

tube support mechanism

strong and counterbalanced with immobilization locks

16
New cards

metal housing

leakage less than 100 mR/hour at 1 meter

17
New cards

max tabletop thickness

1 mm of Al equivalent

18
New cards

high voltage generators

convers low V to kV

step up transformers for kV

step down for tube filament

rectifiers convert incoming alternating V to pulsating direct current

requires less space and less power

19
New cards

ripple for high frequency generators

less than 1%

20
New cards

timer accuracy

if timer is off, mAs is off

tested annually

tested with spinning top or motorized synchronous top

21
New cards

back up timer

should terminate exposure after 6 seconds or 600 mA

22
New cards

what equipment is tested with a spinning top

single phase equipment

23
New cards

what equipment is tested with a motorized synchronous top

3 phase equipment

24
New cards

spinning top test

circular steel/lead disc with a small hole on the periphery

can indicate a problem with a rectifier or the timer

25
New cards

rectifier issue with spinning top test

you see ½ of the expected dots or arc angle

26
New cards

timer issue with spinning top test

results are off by a few dots

27
New cards

single phase rectified equipment with spinning top test

produces dots

1 second exposure should produce 120 dots

28
New cards

3 phase rectified equipment with spinning top test

produces an arc

1 second exposure should produce an image with 360 degree arc

29
New cards

how to calculate expectation of dots/arc for spinning top test at >1 second

multiply the seconds by the expected dots/arc for 1 sec

EX: 0.5 sec x 120 dots = 6 dots should be seen on properly working equipment

30
New cards

line focus principle

when target angle is less than 45 degrees, the effective focal spot is smaller than the actual focal spot

31
New cards

focal spot size determines

resolution

smaller FSS = better resolution

32
New cards

blooming

FS size increases as the tube ages and with high mA usage

causes electron cloud to swell

unwanted change in actual focal spot size during a change in exposure

33
New cards

how often is FSS tested

annually

34
New cards

tools for FSS testing

pinhole camera

star or slit camera

  • line pair resolution tool

35
New cards

pinhole camera

CR is aligned through metal sheet with 0.03 and 0.075 mm holes

image of focal spot is displayed using a long exposure

the size of the effective focal spot is measure using a micrometer

36
New cards

blooming change with tube current

higher tube current = more blooming

37
New cards

which is preferred between slit and star cameras

slit cameras

38
New cards

slit camera

metal template with series of slits in groups of 3

  • crosswise and lengthwise

39
New cards

what do slit and star cameras measure

resolution and estimated focal spot size

40
New cards

line pair resolution tool is used in conjunction with

slit and star cameras

41
New cards

line pair resolution tool measures

spatial resolution is measured in lp/mm and corresponds to a certain focal spot size

42
New cards

NEMA specifications for FS variations

FS smaller than 0.8mm may be 50% larger variations

FS middle 0.8mm - 1.5mm may be 40% larger variations

FS larger than 1.5mm may be 30% larger variations

43
New cards

what is kV and mA accuracy

stability of the relationship between kV or mA and the resulting exposure intensity

44
New cards

kV accuracy must be within

± 4 kVp OR ± 5% of the kV output

EX: if you expect 80 kVp and get exposure with 86 kVp is it allowed?

  • 6 kVp is > the 4 allowed, so no

  • 6/80 = 0.075 7.5% which is higher than 5%, so no

45
New cards

how is kV accuracy measured

old

  • WI test cassette

new

  • digital kVp meter

  • more accurate and easier

46
New cards

linearity

sequential increases in mA should produce the same increase in the exposure rate

47
New cards

linearity requirement

exposures must not vary more than ± 10%

48
New cards

reproducibility

output intensity must be consistent from one exposure to the next

same technique should produce the same exposure

49
New cards

reproducibility requirement

must be within ± 5%

50
New cards

gassy tube

tube with fluctuations in mA readings

51
New cards

half-value layer is an expression of

beam quality

52
New cards

filtration of the primary beam

filter absorbs low energy xrs

  • improves image quality

  • reduces pt dose

  • increases average beam energy

expressed in Al/Eq

53
New cards

how is filtration measured

computerized dosimeter provides a HVL readout

54
New cards

NCRP requirement for filtration

for equipment operating above 70 kVp, minimum of 2.5 mm Al/Eq total filtration is required

55
New cards

collimation

beam restrictors limit the size of the XR field

improves image quality

  • less scatter

  • improved contrast

reduces pt dose

should never be larger than the IR

56
New cards

collimation congruency

measures how well a collimator regulates

57
New cards

NCRP requirement for collimation congruency

accurate within ± 2% of the SID

tested annually

58
New cards

how is collimation congruency tested

8 or 9 penny test tool

if variance is smaller than the penny it is acceptable

59
New cards

central ray accuracy is measured with

beam alignment tool

  • plastic cylinder with metal washers or BBs

60
New cards

central ray accuracy requirement

must be accurate within ± 1% of the SID

  • 40 inch SID = 0.4 inch variance

  • 72 inch SID = 0.72 inch variance

tested annually

61
New cards

distance accuracy is measured by

measuring devices and detents are checked with a tape measure

tube angles are checked with a protractor

62
New cards

distortion is checked with

wire mesh test

63
New cards

wire mesh test

mesh is placed on cassette

dark areas indicate poor contrast

can show pincushioning in fluoro

64
New cards

pincushion distortion

warped image appearance from curved surface of input screen to flat surface of output screen

65
New cards

exposure requirements in fluoro

must not exceed 10 mR/min at tabletop

entrance dose must not exceed 5R/min

66
New cards

how are exposure requirements tested in fluoro

physicists measures output with a calibrated ion chamber

67
New cards

artifacts in fluoro

veiling glare

vignetting

distortion

  • pincushion

  • s-distortion

68
New cards

required thickness for primary barrier of the housing in fluoro

2 mm Pb/Eq

69
New cards

required thickness for bucky slot cover in fluoro

0.25 mm Pb/Eq

70
New cards

required thickness for fluoro curtain

0.25 mm Pb/Eq

71
New cards

timer requirement for fluoro

5 minutes

72
New cards

sourced to table distance requirement for stationary fluoro units

15 inches

73
New cards

sourced to table distance requirement for mobile fluoro units

12 inches

74
New cards

field uniformity checks for

checks the image visually for defects

75
New cards

ghosting

inadequate CR erasure

  • loss of lamp intensity or erasure time too short

residual charge is trapped in DEL of digital IR

off-focus/STEM radiation

76
New cards

examples of CR quality control tests

dark noise

uniformity

spatial accuracy

erasure throroughness

77
New cards

dark noise is AKA

intrinsic noise

78
New cards

dark noise test

scans an unexposed plate

should come out as a uniform lighter gray color

79
New cards

uniformity test

expose a plate and run it through the scanner

should come up as a uniform dark image

80
New cards

spatial accuracy test

wire mesh is placed on the CR and exposed and ran through the scanner

should have no distortion

81
New cards

erasure thoroughness test

expose a CR with a step wedge, run through scanner and erase, run through scanner again

after the second time through the CR shouls be blank

82
New cards

AAPM proposed standard for exposure index

to use deviation index numbers

83
New cards

what is the weakest link in the imaging chain

the monitor

84
New cards

ACR and FDA classify monitors as

primary monitors

secondary monitors

85
New cards

primary display monitor

diagnostic monitor

used by the radiologist for interpretations minimum of 2K (2048×2560) required

  • minimum of 4K for mammography

86
New cards

secondary monitors

technologist review stations

minimum of 1K (1024×1280) required

not considered diagnostic quality

87
New cards

types of viewing monitors

cathode ray tube

flat panel

  • LCD or Plasma

88
New cards

how often must viewing monitors be cleaned

monthly

89
New cards

cathode ray tube monitors

resolution deteriorates over time (blooming)

higher reflectance

better luminance

lower resolution

  • pincushion, barrel, and skew distortions

  • veiling glare cause by scatter of the phosphor layer

90
New cards

flat pannel liquid crystal display (LCD)

layer of liquid crystals sandwiched between two transparent electrodes

used for laptops and flat panel monitors

91
New cards

advantages of flat panel LCD

uses less power

better resolution and contrast

faster refresh rate

higher level of luminance longer

no distortion from:

  • pincushion

  • barrel

  • skew

  • veiling glare

92
New cards

drawbacks of flat panel LCD

limited viewing angle and more easily damaged

93
New cards

flat panel plasma monitors

acctive array of tiny fluorescent bulbs emit light from each pixel location

94
New cards

advantages of flat panel plasma monitors

thinner than an LCD

higher luminance output

wider viewing angle than LCD

95
New cards

drawbacks of flat panel plasma monitors

more expensive

shorter life

96
New cards

ACR remmonendations for flat panel plasma monitors

monthly performane evaluation

97
New cards

AAPM recommendations for flat panel plasma monitors

acceptance testing evaluation

annual physicist evaluation

daily/monthly/quarterly evaluation by QC tech

98
New cards

types of quality testing on display monitors

luminance

illuminance

resolution

noise

dead and stuck pixels

99
New cards

luminance for display monitors

ability to display different shades of brightness

rate of light emitted from a source

ACR requires min brightness of 250 lumens

100
New cards

illuminance for display monitors

rate of light striking the surface

max ambient lighing in a reading room must be less than ¼ normal lighting