Energy sources of MRI
Main magnetic field
Gradient field
Radiofrequency pulse {RF}
Main Magnetic field
B0
1/15
Looks like no tags are added yet.
Name | Mastery | Learn | Test | Matching | Spaced |
---|
No study sessions yet.
Energy sources of MRI
Main magnetic field
Gradient field
Radiofrequency pulse {RF}
Main Magnetic field
B0
B0 used for
Clinically used for molecular alignment. This is the cause for projectiles flying into the magnet.
Spatial distribution of static magnetic field
How rapidly the magnetic field changes is what causes pulling of ferromagnetic objects.
The greater the differences, more pull.
Items in the center would fall to the bottom from gravity.
B0 is responsible for
translation-projectile tork-rotation
cause for projectiles flying into the magnet
Main magnetic field
what causes pulling of ferromagnetic objects
rapidly the magnetic field change.
greater differences the greater the pull
Gradient Coils
Db/Dt
Db/Dt
Rate of change of magnetic field over time..
Db/Dt does what
Uses electricity to slope the magnetic field in a linear fashion.
dangers of Db/Dt
This can cause electrical malfunctions to internal devices and or burns.
Can induce current into devices
where is Db/Dt most dangerous
Uses electrical charge
Greatest at ends and riskiest there.
Db/Dt is clinically used for
Spatial resolution, image formation
Radio Frequency Pulse
B1
B1 does what
Resonate protons to create the MR signal.
Dangers of B1
This can cause proximity burns where the pt may be touching the bore. Increase on sides of magnet
Large patients more susceptible
Electricity can also be induced into abandoned leads.
how to protect against burns
Important to use manufacture pads where pts are in contact with the bore.