appears in mulitples
reverberation
appears equally spaced
reverberation
located parallel to sound beam’s main axis
reverberation, comet tail
located at ever-increasing depths
reverberation
appears as a single long hyperechoic echo
comet tail
hypo- or anechoic (background color)
shadowing; edge shadow
result of too much attenuation
shadowing
located beneath the structure with abnormally high attenuation
shadowing
prevents visualization of true anatomy on the scan
shadowing; edge shadow
results when the beam spreads after striking a curved reflector
edge shadow
extends downward from the curved reflector’s edge, parallel to beam
edge shadow
hyperechoic (appears the same as the foreground color)
enhancement
result of too little attenuation
enhancement
located beneath a structure with abnormally low attenuation
enhancement
hyperechoic side to side region (appears the same as the foreground color)
focal enhancement
results from increased intensity at the focus
focal enhancement
second copy of true reflector
mirror image; lobes
artifact appears deeper than the true reflector
mirror image
bright reflector lies on a straight line between the artifact and the transducer
mirror image
true reflector and artifact are equal distances from each other
mirror image
sound travels slower than the US system expects
speed error (medium’s speed is too slow)
pulses return from their trip in the body slow
speed error (medium’s speed is too slow)
go return time is too long
speed error (medium’s speed is too slow)
system assumes reflectors are far from transducer
speed error (medium’s speed is too slow)
reflectors places too deep on image
speed error (medium’s speed is too slow)
distances are overestimated (reporter number is too large)
speed error (medium’s speed is too slow)
sound travels faster than the US system expects
speed error (medium’s speed is too fast)
pulses return from their trip in the body fast
speed error (medium’s speed is too fast)
go return time is too short
speed error (medium’s speed is too fast)
system assumes reflectors are close to transducer
speed error (medium’s speed is too fast)
reflectors are located too shallow on image
speed error (medium’s speed is too fast)
distances are underestimated (reported number is too small)
speed error (medium’s speed is too fast)
the artifact and the true reflector are located side by side at the same depth
lobes
a second copy of the reflector
refraction
the copy is side by side or at the same depth as the true reflector
refraction