CT Imaging Instrumentation

0.0(0)
Studied by 0 people
call kaiCall Kai
Locked
learnLearn
examPractice Test
spaced repetitionSpaced Repetition
heart puzzleMatch
flashcardsFlashcards
GameKnowt Play
Card Sorting

1/11

flashcard set

Earn XP

Description and Tags

Subgroup IV: Instrument Operations and Quality Control

Last updated 8:14 PM on 7/18/26
Name
Mastery
Learn
Test
Matching
Spaced
Call with Kai
Chat

No analytics yet

Send a link to your students to track their progress

12 Terms

1
New cards

CT performance characteristics

  • slice thickness

  • CT number accuracy

  • uniformity

  • noise

  • contrast resolution

  • spatial resolution

2
New cards

CT daily QC procedures

  • tube warm-up

  • tomographic uniformity

  • accuracy of the CT number of water

  • image noise

3
New cards

CT tube warm-up definition

monitoring of tube output and detector response

4
New cards

tomographic uniformity definition

variation in Hounsfield Units in ROIs within a uniform phantom

5
New cards

image noise definition

measured as std deviation of the CT numbers in an ROI

6
New cards

CT monthly or quarterly QC procedures

  • laser light alignment

  • slice thickness, measures in mm

  • spatial resolution (high-contrast resolution)

  • CT number accuracy

  • contrast (low-contrast resolution)

7
New cards

How is spatial resolution measured?

line pairs/cm

8
New cards

CT semiannual QC procedures

linearity of Hounsfield Unit values

9
New cards

CT annual QC procedures

radiation dose

10
New cards

CT radiation dose procedure

  • measure phantoms with pencil ion chambers in different locations

  • calculate CT dose index (CTDI)

11
New cards

CT ring artifacts are due to

non0functioning detector (should also be seen on topogram)

12
New cards

most CT artifacts are related to ___

the patient, acquisition parameters, and/or reconstruction technique