Subgroup IV: Instrument Operations and Quality Control
Name | Mastery | Learn | Test | Matching | Spaced | Call with Kai | Chat |
|---|
No analytics yet
Send a link to your students to track their progress
CT performance characteristics
slice thickness
CT number accuracy
uniformity
noise
contrast resolution
spatial resolution
CT daily QC procedures
tube warm-up
tomographic uniformity
accuracy of the CT number of water
image noise
CT tube warm-up definition
monitoring of tube output and detector response
tomographic uniformity definition
variation in Hounsfield Units in ROIs within a uniform phantom
image noise definition
measured as std deviation of the CT numbers in an ROI
CT monthly or quarterly QC procedures
laser light alignment
slice thickness, measures in mm
spatial resolution (high-contrast resolution)
CT number accuracy
contrast (low-contrast resolution)
How is spatial resolution measured?
line pairs/cm
CT semiannual QC procedures
linearity of Hounsfield Unit values
CT annual QC procedures
radiation dose
CT radiation dose procedure
measure phantoms with pencil ion chambers in different locations
calculate CT dose index (CTDI)
CT ring artifacts are due to
non0functioning detector (should also be seen on topogram)
most CT artifacts are related to ___
the patient, acquisition parameters, and/or reconstruction technique