What’s the penetration number for the linear array probe?
8 cm
Near field
Area nearest to the transducer
Which TGC will fix the near, mid, and far field?
Near field = top or higher TGC sliders
Mid field = middle sliders
Far field = bottom or lower sliders
Name the types of transducers & for which exams
Linear transducer - for vascular
Phased array or Sector transducer - for echo, abdominal (when penetration is needed)
Curvilinear transducer - for OBGYN/abdominal
Vector transducer - for abdominal (sometimes cardiac or intercostal scan)
Name the faces of the transducers
Sector, Vector, Phased Array : square face
Linear : rectangle face
Curvilinear : curved, convex face
What image shape/scan format does each transducer produce?
Linear → produces rectangular image
Curvilinear → produces sector (pie shaped) image with a CURVED top
Phased/sector → produces sector (pie shaped) image with a POINTED top
Vector → produces sector (pie shaped) image with a FLAT top
Frequency vs Resolution vs Penetration
Low frequency = deeper penetration, but lower resolution
Higher frequency = less penetration, but higher resolution
What to label/annotate on images?
Label
the side of the patient being scanned (left/right)
the organ
location details (proximal, mid, distal)
transducer orientation (transverse or sagittal)
patient position (supine or prone)
Role of a sonographer
Provide diagnostic images for a physician
A-mode
Amplitude mode
Not used often today, except ophthalmology
Spikes represent strength of returning echoes
B-mode
Brightness mode
Range of whiteness from black to white
Echoes create spots that make image
Gray scale (256 shades of gray) that create image
M-mode
Motion of structures along a signal graph overtime
Graphs depth vs. time of structure
Used in cardiac scans
Pulser/transmitter
Provided electrical excitation to the crystals in transducer
What’s the Piezoelectric effect?
Ability for the energy (crystals in transducer) to go from electrical energy to mechanical and back
What’s another name for Overall Gain?
2D Gain or Amplification Gain
What does adjusting overall gain do?
Adjusts overall image brightness
What does adjusting the TCG do?
Adjusts the brightness in a specific area of the image
What does adjusting the focus do?
Allows more focus/clarity on a specific area (better quality in that area)
What does adjusting depth do?
Adjusts the depth of ultrasound wave
Decreasing depth allows you to focus / hone in on specific structure
Where is depth indicated on an image?
Indicated by dashes on the screen on the side of an image
What’s the focal zone?
Ability to focus transducer at various depths for better resolution
How to adjust the focus?
Use focus knob: place the triangle symbol (CARET) at or just below the area of interest to focus
Function of Trackball
• Allows you to navigate screen (similar to computer mouse)
• Annotations
• Focus
• Gain
• Cineloop
• Calipers
• Change size of scan sector/color box
Frequency of linear transducers
5 to 18 MHz
Frequency of curvilinear array transducers
2 to 5 MHz
Frequency of sector/phased array transducers
3 to 5 MHz
Frequency of vector transducers
2 to 5 MHz
Why do you annotate?
Identifies anatomy & pathology
Provides orientation & views for referencing
What’s another term for Dynamic Range?
Log compression
What is Dynamic Range?
Number of shades of gray displayed (keep it around 60 dB)
Narrow Dynamic Range vs. High Dynamic Range
Narrow dynamic range = high contrast image, very black & white, grainy
High dynamic range = low contrast image
Why would you use TGC versus overall gain?
To adjust the brightness in a specific area
Near Zone vs. Far Zone?
Near zone = area closest to transducer; higher quality
Far zone = area furthest from transducer; lower quality due to depth
Near Gain vs. Far Gain?
Near Gain = controls brightness at top of image
Far Gain = controls brightness at bottom of image
Tips for holding probe properly
• Hold it like a pencil, relaxed
• Drop & relax elbow
• Relax shoulders
Keep hand underneath probe
Hold hand as close to body as possible
Rest hand on patient (get permission)
Hang cable on machine or wrap around wrist
What are matrix transducers?
Special transducers that allow 3D/4D imaging
Various MHz
Has columns and rows of crystals
What are multihertz transducers?
Operates over a range of frequencies without needing to switch transducers