what focal spot size is used in routine imaging

0.3 mm

2
New cards

T/F: The AEC detector covers the entire size of the imaging detector

true

3
New cards

T/F: The RWS has a resolution of 2 megapixels and the AWS has a resolution of 5 megapixels.

false (RWS 5 megapixels and AWS 2-3)

4
New cards

Who is in charge of the day to day QC tests?

Technologist

5
New cards

The new ACR DM Phantom has how many speck groups?

6

6
New cards

The compression force for the power drive mode should be at least _____ but not exceed ______.

25 lbs, 45 lbs

7
New cards

How long must a facility keep mammogram records if the patient does not return to the facility?

10 years

8
New cards

The BI-RAD category for a benign mammogram is:

BI-RAD 2

9
New cards

All technologists are required to complete_________.

200 mammograms every 24 months

10
New cards

T/F: The large focal spot is used for magnification views while the small focal spot is used for routine views.

false (small for mag views, large for routine views)

11
New cards

Which filter is used in tomosynthesis (3D) imaging?

Aluminum (Al)

12
New cards

T/F: Lossless Compression is required by the ACR when storing Mammography images.

ctrue

13
New cards

what is the best method to control voluntary motion?

communication

14
New cards

The overall repeat rate ideally is ____ but should not exceed ____.

2%, 5%

15
New cards

How often is the compression thickness indicator test performed?

monthly

16
New cards

Who provides a written report with corrective action and/or recommendations.

medical physicist

17
New cards

A minimum of _____ speck groups must be seen on the new ACR DM phantom.

 

≥3

18
New cards

T/F: QC tests MUST be performed on the same day of each week/month.

false

19
New cards

T/F:Accreditation and Certification are different processes.

true

20
New cards

Clinical images submitted for accreditation must include:

1- Two CC and Two MLO projections

2- One CC and One MLO projections

3- Two fatty and two dense images

4- One fatty and one dense images

1 and 4

21
New cards

A concerning result must be sent to the physician in______. 

3-5 days from the date of interpretation

22
New cards

The technologist's initial mammography qualifications include:

1- Completion of a 40 hour mammography course

2- Completion of 25 supervised examinations

3- Completion of 8 hours of training specific to digital and DBT

4- Completion of 200 unsupervised examination

1, 2, 3

23
New cards

The BI-RAD category for an incomplete mammogram is:

BI-RAD 0

24
New cards

The ideal focal spot size for magnification mammography is:

0.1 mm

25
New cards

The standards for mammographic imaging have designated that the correct placement of the markers (R,L, etc. ) is:

Near the aspect of the breast closest to the axilla

 

26
New cards

T/F: The posterior edge of a compression paddle is rounded along the chest wall.

false

27
New cards

In mammography the kVp range is typically from:

20-40

28
New cards

You are designing a mammographic unit. All of the following would be target material EXCEPT?

aluminum

29
New cards

You are designing a mammographic unit. You need to select the range of focal spot sizes for routine imaging. You decide to go with the most common sizes which are:

0.3 mm to 0.4 mm

30
New cards

The recommended SID for mammographic units ranges from:

60-65 cm

31
New cards

The generator that produces the most accurate and reproducible x-ray output is:

high frequency

32
New cards

Altering the targets and filters can be most beneficial for:

Shaping the x-ray beam to penetrate both dense and fatty tissue

33
New cards

The most common magnification factor used in mammography is:

1.5-2.0

34
New cards

Which of the following is TRUE concerning magnification?

The incidence of patient motion increases with magnification

35
New cards

The factor that the technologist cannot control in the area of subject contrast is:

tissue density

36
New cards

The x-ray tube is typically made of:

1. Glass 

2.  Metal 

3.  Rhodium

1 and 2

37
New cards

In mammography collimation should be:

To the size of the detector

38
New cards

High Transmission Cellular grids (HTC) are constructed:

  

With copper strips in a crossed grid

39
New cards

The grid ratio in mammography is:

3:1-5:1

40
New cards

Which of the following is NOT true concerning AEC devices in mammography?

The AEC position is stationary

41
New cards

Dedicated mammography units have a mAs range from:

20-600 mAs

42
New cards

The mammographic x-ray tube is angled to optimally utilize the strengths of the

line focus principle

43
New cards

To minimize the anode heel effect the tube is tilted approximately:

6 degrees

44
New cards

In non-magnification views, the exposure time in 2D mammography will usually vary from:

0.4-1 second

45
New cards

In the design of line focus principle the target is angled approximately:

20-24 degrees

46
New cards

The  breast phantom approximates a breast thickness of:

4.2 cm

47
New cards

the phantom image evaluation is performed at least:

weekly

48
New cards

The action taken on a regularly scheduled basis to prevent deterioration of image quality or any breakdown in the imaging chain is:

preventive maintenance

49
New cards

The monitoring and maintenance of equipment with data collection and controlled variables is called:

Quality Control

50
New cards

The overall management actions which consistently provides high image quality is:

Quality Assurance

51
New cards

MQSA requires all Mammography Technologists participate in _________ of mammography specific continuing education credits.

15 credit hours in a 3 yr period

52
New cards

The infection control procedure (cleaning and disinfecting the mammography equipment after contact with blood or other body fluids) is required by:

FDA

53
New cards

Which of the following is the most common problem encountered when evaluating clinical images?

Incorrect Positioning

54
New cards

The maximum amount of force on compression on the initial power drive allowed by MQSA is:

45 pounds

55
New cards

The Medical Physicist test of the x-ray field cannot extend beyond any edge on the image receptor by more than what percentage?

2% of SID

56
New cards

Which of the following mammographic quality control tests should be performed semiannually?

compression force

57
New cards

Which of the following is FALSE concerning phantom imaging?

The phantom is equal to a compressed breast that is 3.5 mm in thickness

58
New cards

When performing a repeat analysis the test should be performed with

at least 250 patients

59
New cards

On the new ACR DM Phantom Image what is the minimum number of fibers, specks and masses that must be seen?

2 fibers, 3 specks and 2 masses

60
New cards

The monitor used to interpret the mammograms is the:

review workstation monitor RWS

61
New cards

A computer network system that uses DICOM to view images of different modalities at various monitors or transmitted or stored is:

Picture archiving and communication system (PACS)

62
New cards

The transmission of the digital image from one location to another via electronic transfer is:

teleradiography

63
New cards

The program which allows communication between different medical imaging devices such as scanners, printers, servers, PACS, workstations and networks from multiple manufactures is:

DICOM

64
New cards

The type of data compression for storage that when uncompressed contains the same information without losing information is:

lossless compression

65
New cards

The computer program that acts as a second reader to the radiologist by analyzing suspicious areas is:

Computer Aided Detection (CAD)

66
New cards

The agency that performs a yearly inspection as part of MQSA is the:

FDA

67
New cards

Silver is used as the filtration in:

Digital mammography with tungsten targets

68
New cards

Which QC test is performed on a monthly basis?

visual check list

69
New cards

What is the compression thickness indicator test?

 

A check of the indicated compression thickness

70
New cards

Who is responsible for the general oversight of the MQSA QA program

radiologist

71
New cards

Who conducts the Mammography Equipment Evaluation (MEE)?

medical physicist

72
New cards

Communication of results to a patient:

1. Must be within 30 days of the mammogram

2.  Are only given verbally

3. Are only given written

4.  Must be in lay terms

1, 3, 4

73
New cards

The medical audit tracks mammograms that are given a:

BIRAD 4 and 5

74
New cards

T/F: The purpose of the Mammography Quality Standards Act (MQSA) is to establish minimal nation quality standards for mammography.

True

75
New cards

According to the FDA, a facility must maintain a copy of consumer complaints for:

3 years