Chapter 11

0.0(0)
studied byStudied by 0 people
learnLearn
examPractice Test
spaced repetitionSpaced Repetition
heart puzzleMatch
flashcardsFlashcards
Card Sorting

1/111

flashcard set

Earn XP

Description and Tags

RAD 136

Study Analytics
Name
Mastery
Learn
Test
Matching
Spaced

No study sessions yet.

112 Terms

1
New cards

What should all diagnotic imaging systems include?

-Protective tube housing

-correctly functioning control panel

-console

-exam table

2
New cards

What is required to safeguard the patient and imaging personnel from off-focus, or leakage, radiation by restricting the emission of x-rays to the area of the useful, or primary, beam?

Diagnostic-type protective tube housing

3
New cards

The housing enclosing the x-ray tube must be constructed with _ to reduce radiation leakage through any portion of the housing away from the useful beam.

Lead lining

4
New cards

Leakage radiation is measured at a distance of ____ meter from the x-ray source, and must not exceed an air kerma rate of ____ mGy/hr when the tube is operartd at it;s highest voltage at the highest current that allows continuous operation.

1; 0.88 mGy/hr

5
New cards

The place where technical exposure factors, such as milliamperes (mA), peak kilovoltage (kVp), and exposure duration (ms), are selected and visually displayed.

Control panel/console

6
New cards

For the radiation safety of the radiographer, where must the console be located that also has radiation-absorbent window?

Behind a shield barrier

7
New cards

For some state-of-the-art operating consoles, digital controls for kVp and mAs are availble on a what?

Touch screen

8
New cards

When happens when an x-ray exposure begins?

A tone is emitted and a radiation sign iliminates

9
New cards

What must be strong enough to adequately support patients whose weight can be up to 400 pounds?

Radiographic table

10
New cards

The _ of the tabletop must be uniform and for under-table x-ray tubes as used in fluoroscopy

Thickness

11
New cards

What is commonly used in a tabletop to reduce excessive x-ray tube emission?

Carbon fiber material

12
New cards

Is it OK for the collimator to be bigger than the IR?

NOOOOOOO

13
New cards

An x-ray beam limitation device that permits the equipment operator to adjust the size and shape of the x-ray beam either automatically or manually

Light-localizing, variable-aperture rectangular collimator

14
New cards

Any device that limits the dimensions of the useful beam to a designated size and shape before it enters the area of clinical interest

X-ray beam limitation device

15
New cards

All the non-useful image-forming radiation that arises from the interaction of an x-ray beam with the atoms of a patient or any other object in the path of the beam

Scatter radiation

16
New cards

X-rays emitted from parts of the x-ray tube other than the focal spot. Also called stem radiation

Off-focus radiation

17
New cards

Luminance is a technical term referring to the _ of a surface

Brightness

18
New cards

A feature of radiographic collimators that automatically adjusts them so that the radiation field size matches the size of the IR.

Positive Beal Limitation (PBL)

19
New cards

Elements that are part of or added to the x-ray tube to reduce exposure to the patient's skin and superficial tissue by absorbing most of the lower energy photons from the produced beam.

Filtration

20
New cards

The thickness of a designated absorber required to decrease the intensity of the primary beam by 50% of its initial value

Half-value layer (HVL)

21
New cards

The consistency in radiation output intensity for identical generator settings from one single exposure to subsequent exposures

Exposure Reproducibility

22
New cards

Refers to a consistency in output radiation intensity at any selected kVp when generator settings are changed from 1 milliamperage and time combination to another

Exposure linearity

23
New cards

Also known as "phototiming", that is an x-ray termination device that ends the radiation when an arrangement of sensors receives a predetermined amount of radiation.

Automated Exposure Control (AEC)

24
New cards

A device made of parallel radiopaque strips alternately separated with low-attenuation strips of a material such as aluminum or plastic

Radiographic grid

25
New cards

The use of a flat panel detector to record a radiographic image and render it in digital form without a film developing process or the scanning of the image receptor

Digital Radiography (DR)

26
New cards

The distance from the anode focal spot of an x-ray tube to the skin of the patient

Source-to-skin distance (SSD)

27
New cards

The array of pixels that comprises a digital image

Image Matrix

28
New cards

The term that x-ray systems that generate images using the process of photostimulable luminescence

Computer Radiography (CR)

29
New cards

Converts the pattern of x-rays transmitted through the patient into a corresponding and amplified visible light pattern

Image intensifier

30
New cards

A technique in which the fluoroscopic image is produced in digital form in the image acquisition process

Digital fluoroscopy

31
New cards

A device that measures the collective x-ray beam-on time and sounds an audible alarm, or, in some cases, temporarily interrupts the radiation until it is reset after the fluoroscope has been activated for 5 minutes

Cumulative timer

32
New cards

Limited by federal standards of general-purpose intensified fluoroscopic units to a maximum of 88 mGy, per minute.

Entrance skin irradiation rate

33
New cards

A barrier designed to prevent primary, or direct, radiation from reaching personnel or members of the general public on the other side of the barrier

Primary protective beam

34
New cards

When the x-ray beam is deactivated while the image is being scanned, thereby decreasing patient dose, and then pulsed back on for the next image

Pulsed progressive systems

35
New cards

A fluoroscopic technique used in interventional radiology for visualizing blood vessels

Digital subtraction angiography

36
New cards

A method of digital image subtraction in which the frame that contains the greatest amount of contrast material in vessels is indentified and is then subtracted from all subsequent images

Road mapping

37
New cards

An operating mode for state-of-the-art fluoroscopic equipment in which entrance radiation levels are substantially higher than those normally employed in routine procedures

High-level control fluoroscopy (HLCF)

38
New cards

Added filtration plus inherent filtration

Total filtration

39
New cards

What is a compensating filter used for?

To even out the radiation dose when imaging body parts of different thicknesses

40
New cards

What are some examples of a compensating filter?

Wedge filters (for feet), trough filters (for spine)

41
New cards

What is used to reduce scatter radiation from reaching the IR?

Radiographic grid

42
New cards

What are mobile units (portables) usually used for?

For patient's who cannot move

43
New cards

What is the minimum distance from source to patient on a mobile study?

12 inches

44
New cards

SSD much be at least ___ inches for fixed images and ___ inches for mobile studies

15; 12

45
New cards

The tube housing must not leak more than when tested

100 mR/hr

46
New cards

It is required that the _ be mechanically affixed to the console such that it cannot be activated while the operator is in an unshielded location

Exposure hand control

47
New cards

The thickness of the tabletop must be uniform, and for under-table x-ray tubes as used in fluroscopy, the patient support surface should be as _ as possible

Radiolucent

48
New cards

Distance and centering indicators must be accurate to within what percentages of the SID?

2% and 1%

49
New cards

The light-localizing variable-aperature is used in multipurpose x-ray units. IT is box-shaped and contains the radiographic beam-defining system. This system consists of:

  1. Two sets of adjustable lead shutters mounted within the device at different levels

  1. A light source to illuminate the x-ray field and permit it to be centered over the area of clinical interest

  1. A mirror to deflect the light beam toward the patient to be imaged

50
New cards

To minimize skin exposure to electrons produced by photon interaction with the collimator, the patient's skin surface should be at least _ cm below the collimator for fixed radiographic equipment

15

51
New cards

Portable radiographic units are required to maintain a source-to-skin distance of at least how many cm?

30

52
New cards

Luminance is determined by measuring the concentration of light over a particular field of view. The primary unit is the _, known simply as ____.

Candela per square meter; nix

53
New cards

The SI-derived unit of luminous flux, also known as ____.

Lumen

54
New cards

What is equal to the amount of light emitted per second passing into a solid angle of the steradian from a uniform souce of one candela

Lumen

55
New cards

One lumen corresponds to about the same brilliance as a normal birthday candle has at a distance of _ foot.

1

56
New cards

Luminance must be high enough so that a calibrated light meter reading taken at a distance of ____ cm will, in the English system of units, be equal to at least _____ foot candle

100; 15

57
New cards

Besides a tape measure attached to the collimator, what else can be manually measured?

Laser

58
New cards

What is the predominant x-ray beam limitation device?

The collimator

59
New cards

What is the purpose of the collimator?

To reduce radiation scatter

60
New cards
<p>What is this?</p>

What is this?

X-ray beam limitation device

61
New cards
<p>What is “A”?</p>

What is “A”?

Anode (+) focal spot in an X-ray tube that serves as the positive electrode and is responsible for the emission of X-rays when electrons collide with it.

62
New cards
<p>What is “B”?</p>

What is “B”?

Cathode (-)

63
New cards
<p>What is “C”?</p>

What is “C”?

X-ray tube window (port)

64
New cards
<p>What is “D”?</p>

What is “D”?

First set or upper shutters

65
New cards
<p>What is “E”?</p>

What is “E”?

Light source

66
New cards
<p>What is “F”?</p>

What is “F”?

Second set or lower shutters

67
New cards
<p>What is “G”?</p>

What is “G”?

Useful (primary) beam

68
New cards
<p>Whatis “H”?</p>

Whatis “H”?

Image receptor

69
New cards
<p>What is “I”?</p>

What is “I”?

Imaging plate

70
New cards
<p>What is “J”?</p>

What is “J”?

Light source projection on image plate area of radiographic beam

71
New cards
<p>What is “K”?</p>

What is “K”?

Off-focus radiation

72
New cards
<p>What is “L”?</p>

What is “L”?

Collimator

73
New cards
<p>What is “M”?</p>

What is “M”?

Mirror

74
New cards
<p>What is “N”?</p>

What is “N”?

Aluminum filter

75
New cards
<p>What is “O”?</p>

What is “O”?

X-ray tube

76
New cards

What are the two types of filters?

Added and Inherent

77
New cards

Inherent filtration always includes:

  1. Glass envelope encasing the tube

  2. Insulating oil surrounding the tube

  3. Glass window in the tube housing

78
New cards

Added filtration + inherent filtration equals what?

Total filtration

79
New cards

A minimum of how much ____ mma aluminum equivalent total filtration is required for fixed radiographic units operating at above 70 kVp

2.5 mm

80
New cards

What metal is the most widely selected as a filter material?

Aluminum

81
New cards

Why is aluminum the most widely used metal as a selected filter material?

It effectively removes low-energy x-rays from a polyenergetic (heterogenous) x-ray beam without severely decreasing the x-ray beam intensity

82
New cards

What is HVL expressed in?

Millimeters of aluminum

83
New cards

What is HVL a measure of?

A measurement of beam quality, or effective energy of the xray beam

84
New cards

What are compensating filters made of?

  1. Aluminum

  2. Lead-acrylic

  3. Or other suitable materials

85
New cards

Why are compensating filters used?

Used to accomplish dose reduction and uniform imaging of body parts

86
New cards

In exposure reproducibility, a variance of how much percent or less is acceptable?

5%

87
New cards

The ratio of the difference in mSv/mAs or mR/mAs values between two successive generator stations to the sum of these mSv/mAs or mR/mAs values. Must be less than 0.1

Linerarity

88
New cards

When settings are changed from one mA to a neighboring mA station, the most the linearity can vary is what by what percent?

10%

89
New cards

What are AECs designed to do?

To produce an acceptable diagnostic image while limiting the total amount of radiation exposure to the patient

90
New cards

Old terminiology for referring to photomultiplier tubes

Phototiming

91
New cards

Today’s AEC systems utilizie what?

Ionization chambers

92
New cards

T or F: Radiographic grids reduce patient dose

FALSE

93
New cards

When the SSD is small, because of the relatively greater rate of spreading out of the x-ray beam at a close distances fro the x-ray source than at larger distances, the patient’s entrance does is significally _____ than the exit exposure

Larger

94
New cards

Distance generally for mobile radiography is _____ cm to ______ cm

100;120

95
New cards

What are the 5 digital processed radiography imaging modes of xrays?

  1. Computed tomography (CT)

  2. Computed radiography (CR)

  3. Digital radiography (DR)

  4. Digital fluoroscopy (DF)

  5. Digital mammography (DM)

96
New cards

For under-table tubes as used in fluoroscopy, the patient support surface also be what?

Radiolucent

97
New cards

What must be used to confine the useful beam before it enters the anatomic area of clinical interest?

X-ray beam limitation device

98
New cards

With digital radiography, the image receptor is divided into small detector elements that make up a 2-dimensional picture elements, or ______ of the image

Pixels

99
New cards

What kind of diagnostic imaging procedure results in the most radiation rates in radiology?

Fluoroscopy imaging

100
New cards

During what kind of fluroscopic procedures does the patient-image intensifier distance should be as short as possible?

C-arm fluoroscopic procedures