This process allows analytical instruments to perform many tests with minimal human involvement?
Automation
This type of analysis subjects a specimen to multiple analytical processes to obtain a panel of results
Multiple-channel analysis
The most common analyzer configuration where any specimen can be analyzed for any test in any order
Random access analysis
A chemistry analyzer uses reusable cuvettes, fixed pipettes, and inadequate probe washing. A potassium result of 9.8 mmol/L immediately follows a specimen with severe hemolysis
Carry-over
This feature automatically releases laboratory results that meet predefined acceptance criteria
Auto-verification
The most critical unit operation in laboratory automation
Patient Identification
These sensors reduce splatter and help determine the exact specimen level before aspiration
Liquid Level Sensors (LLS)
Human specimen label reading has an error rate of about 1 in 300 characters, while this technology reduces the rate to about 1 in this many
1 million
In continuous-flow systems, this device moves fluids by compressing flexible tubing with rotating rollers
Peristaltic pump
A laboratory is deciding between an open and closed chemistry system. The laboratory wants to develop its own assays and purchase reagents from multiple vendors.
Open analyzer system
The most common immunoglobulin used as an antibody reagent in immunoassays
IgG
In a competitive immunoassay, bound label is related to analyte concentration in this way
Inversely proportional
This assay format captures antigen between two antibodies
Sandwich (noncompetitive) immunoassay
A heterogenous immunoassay requires this additional step that homogenous assays do not
Separation or wash step
A patient’s hCG concentration is extremely elevated, yet a sandwich immunoassay reports a result within the reference interval. Dilution of the specimen results in a dramatically higher value
Hook Effect
These labels absorb light at one wavelength and emit it at a longer wavelength
Fluorescent labels
The difference between excitation and emission wavelengths is called this
Stokes Shift
Luminol and acridinium ester are examples of this type of label
Chemiluminescent labels
This enzyme is commonly used as a label because it amplifies weak signals
Horseradish peroxidase (HRP)
A sandwich immunoassay for troponin is positive despite the absence of analyte because an endogenous antibody bridges the capture and labeled antibodies
What are heterophile antibodies (or HAMA)
Testing performed outside the physical laboratory but within a healthcare instituion
point of care testing (POCT)
This is the most frequently measure analyte using POCT devices
Glucose
The CLIA category assigned to most POCT devices because they are simple and have a low risk of harm if performed incorrectly
Waived tests
One major advantage of POCT that allows clinicians to make treatment decisions more quickly
reduced turn around time
Regulatory guidelines require these four identifiers to accompany POCT results
Patient, sample, operator, and device identifiers
Case Study:
A healthcare system performs bedside glucose testing and troponin testing, clinic pregnancy testing, and provider-performed wet preps. The smallest number of CLIA certificates that could legally cover all testing sites would include this certificate
PPMP certificate