ELECTROCHEMISTRY
Study of redox reactions
ELECTROCHEMISTRY
Involves the measurement of current
or voltage generated by the activity
of specific ion.
Galvanic Cell
Electrolytic Cell
Two basic electrochemical cells
Galvanic Cell
generates current
Electrolytic Cell
requires current
Current
The flow of electric charge across a certain element
Voltage
Electrical potential difference between two points
I
Symbol of Current
V
Symbol of voltage
A / amps / amperage
unit of current
V / Volts / Voltage
unit of voltage
1 ampere = 1 coulomb/second
SI unit for current
1 volt = 1 joule/coulomb
SI unit for voltage
Ammeter
Measuring intrument of current
Voltmeter
Measuring instrument of voltage
Current
Current or Voltage?
the effect
Voltage
Current or Voltage?
the cause
Current
voltage
______ cannot flow without _______
Voltage
current
________ can exist without _______.
POTENTIOMETRY
Measurement of the differences in voltage at a constant current
Nernst equation
POTENTIOMETRY is expressed in _________
pH
POTENTIOMETRY is used in the measurement of _______
direct measurement of electrical potential
Potentiometric methods of analysis involve the _____________________________ due to the activity of free ions.
Ion-Selective Electrodes (ISE)
composed of electrochemical transducer capable of responding to one given ion.
Ion-Selective Electrodes (ISE)
Very sensitive and specific to the ion it measures
Ion-Selective Electrodes (ISE)
Practical application: measurement of electrolytes namely Na+, K+, and Cl-
Na+, K+, and Cl-
electrolytes measured by potentiometry
Inert-metal electrodes
Metal electrodes
Membrane electrodes
Types of ISE
Inert-metal electrodes
(standard H electrode) for Na+ measurements
Metal electrodes
(Ag/AgCl electrode) for Cl- measurements
Membrane electrodes
(e.g. valinomycin gel electrode) for K+ measurements.
pH Electrode
ISE universally used in the laboratory
pH Electrode
Determines pH values quickly and accurately
Indicator electrode/ Glass electrode
Consists of a silver wire coated with AgCl, immersed into an internal solution of KCl, and placed into a tube containing a special glass membrane tip.
Indicator electrode/ Glass electrode
This membrane is only sensitive to
Hydrogen ions (H).
Composed of an internal wire made of either Silver – silver chloride system or Calomel electrode (mainly mercurious chloride)
Reference electrode
A filling solution of potassium chloride, and a permeable outer casing
Liquid junction
Protects the reference system from the medium to be measured without disconnecting the electrical potential between them.
Potassium chloride (KCl)
is a commonly used filling solution because K+ and Cl- have nearly the same mobilities.
Read-out device/meter (Voltmeter)
Numerically represent the pH measurements
acidic / acidic pH
If the number of H+ ions in the solution exceeds the number of H+ inside the pH meter probe; the solution will register an ________________.
alkaline/basic pH
If the number of H+ ions in the solution is less than the number of H+ inside the pH meter probe; the solution will register an __________________.
neutral pH
If the number of H+ ions in the solution is equal to the number of H+ inside the pH meter probe; the solution will register a _____________.
calibration
Do not measure the pH of unknown solutions without ____________.
7.0
Calibrate pH of unknown sol’n using a solution with a pH of ____.
3M KCI solution
Wash the electrode after use with clean water and store in ______________.
clean water
Apply ___________ to flush the electrode after the cleaning steps
pepsin
HCl
Protein containing samples
Immerse the electrode in _________ or _____ solution for a few hours.
thiourea
HCl
white
Sulfide containing samples
Immerse the electrode in _______ or _____ solution until the diaphragm of the electrode turns _________.
acetone
alcohol
Grease or organic compound samples
Apply ________ or _______ to clean the electrode for a few minutes
0.1 M NaOH
0.1M HCl
General Samples
Immerse the electrodes in _________ or ________ for a few minutes
Gas sensing electrodes
▪ Designed to detect specific gases
Amperometry
Potentiometry
Gas sensing electrodes
Examples are
▪ pO2 (Clark electrode)- ________
▪ pCO2 electrode (Severinghaus electrode)-
Enzyme electrodes
Determined by the reaction product of the immobilized enzyme
Urease
Glucose oxidase
Enzyme electrodes
Examples are:
• ________ – detects urea
• ______________ – detects glucose
COULOMETRY
Measurement of the amount of electricity (in coulombs) at a fixed potential.
Faraday’s Law
COULOMETRY follows __________ law
COULOMETRY
Involves electrochemical titration and is used for the measurement of chloride in both sweat and serum.
Chloride ISE
COULOMETRY is now obsolete and is being replaced today by
AMPEROMETRY
Used in the determination of :
▪ Partial pressure of Oxygen (pO2)
▪ Glucose
▪ Chloride
▪ Peroxidase
Polarography
▪ Utilizes the principle of amperometry
Polarography
Measurement of differences in current at a constant voltage
trace metals
O2
Vitamin C
amino acid concentrations
Polarography is used in the measurement of ____________ , ____, ___________ and ______________________.
Ilkovic equation
Polarography follows what equation?
Anodic Stripping Voltametry
the analyte is first concentrated onto the
surface of an electrode at a constant
potential and then goes back into solution as
the voltage is changed.
iron
lead
Anodic Stripping Voltametry is used in the measurement of ______ and ______.
Electrophoresis
is a separation technique based on the principle that a charged particle in solution will migrate towards one of the electrodes when placed in an electrical field.
Proteomics
large scale study of proteins particularly the structures and functions.
Genomics
study of genome (complete set of DNA or complete set of hereditary information)
basic type of electrophoresis that….
▪ Uses agarose gel
▪ Used in Genomics
Vertical
Electrophoresis
basic type of electrophoresis that….
▪ Uses Sodium Dodecyl Sulfate Polyacrylamide Gel (SDS- PAGE)
▪ Used in Proteomics
Ions
charged particles
Cations
(+) Charged ions
Anions
(-) Charged ions
Zwitterions
neutrally–charged ions
Cathode
(-) electrode; where reduction occurs
Anode
(+) electrode; where oxidation occurs.
Iontophoresis
migration of small charged ions
Zone Electrophoresis
migration of charged macromolecules.
Electrophoretogram
result of zone electrophoresis and consists of a macromolecule.
Driving Force (electrical power)
Supplies current or voltage
Support Medium (gel)
Provide a matrix that allows molecules to separate
Paper
Obsolete type of support medium
Cellulose acetate
(replacement for paper) – separates particles by molecular size.
Agarose Gel
separates particles by electrical charge (usually used for Genomics)
Starch Gel
separates particles by charge and size.
Polyacrylamide Gel
separates particles by charge and size (usually used for proteomics).
Buffer (Fluid)
Provides an ionic solution that allows current to pass through the water.
Barbital (Veronal) Buffer : pH 8.6
Tris-Borate-EDTA: pH 8.7
Most commonly used buffers
Sample
Can be pre-treated serum or DNA
Sample
Placed in wells created by comb
Sample
Usually combined with a dye (ethidium bromide) for easy visualization of migration.
Detecting System
Converts bands into algorithm to know the exact concentration of substances
▪ Densitometer
▪ Documentation System
Detecting System can be:
negative
net electric charge of cathode to anode
positive
net electric charge of anode to cathode
Large molecules
molecules that travels slowly
Small molecules
molecules that travels fast
Electric Field Strength
Increasing the strength of the field (increase Current) also increases migration
▪ Amino Black
▪ Ponceau S
▪ Oil red O
▪ Sudan Black
▪ Fat Red 7B
▪ Coomassie Brilliant Blue
▪ Silver Nitrate
▪ Nitrotetrazolium Blue
Stains