Jen's ABG equipment study tips (2/17)

0.0(0)
studied byStudied by 0 people
GameKnowt Play
learnLearn
examPractice Test
spaced repetitionSpaced Repetition
heart puzzleMatch
flashcardsFlashcards
Card Sorting

1/16

encourage image

There's no tags or description

Looks like no tags are added yet.

Study Analytics
Name
Mastery
Learn
Test
Matching
Spaced

No study sessions yet.

17 Terms

1
New cards

3 components to a blood gas analyzer

  1. Electronic circuitry (displays visuals)

  2. Electrolyte solution (promotes reactions + current)

  3. blood gas electrodes (measures pH, PCO2 and PO2, all other values are calculated)

2
New cards

Measured values of ABG? Calculated ones?

Measured: pH PCO2 and PO2

Calculates all other values

3
New cards

Name of each electrode

pH: Sanz (measured @37C)

PCO2: Severinghaus (measured indirectly)

PO2: Clark/polarographic

4
New cards

Mean definition

Average calculated by dividing sum of all numbers in a group by number of entries

5
New cards

Standard deviation

High + low that represents 95% of where normal population falls when subjected to a certain test, being a measure of variance around the mean.

Aka., measure of spread from the average (mean).

6
New cards

Quality control

Control samples have their range of normal based on ±2 SD’s from the mean, being able to determine mean and SD by running 20 control samples through machine. All 3 levels (acidosis, alkalosis and normal) are run every 24h/shift through vials with a known concentration matching each level.

7
New cards

Calibration

When equipment is adjusted/corrected to mach the control standards to determine accuracy and precision

8
New cards

Accuracy vs precision

Accuracy = mean value of several measurements

Precision = how consistently same measurement will produce same results

9
New cards

Levey-Jennings chart

Functions to tell you whether an ABG machine is ready or not.

Control: shows if all QC (quality control) runs are w/i acceptability (±2 SD’s)
Trend: shows if QC’s start trending towards one side of SD (ex. can still use, but its trending out of control)
Random error: all w/i acceptability except one run (still ok!)
Out of control: 2+ out of acceptability.

10
New cards

What causes a “shift/drift” on L-J chart?

Electrode or sensor malfunction if protein is deposited on membrane or there’s a leak in the membrane, electrolyte, depletion or sensor failure.

11
New cards

What causes it to be “out of control” on L-J chart?

Temperature control failure to maintain 37C or thermometer inaccuracy.

12
New cards

What causes it to be “consistently high or low” on L-J chart?

Improper calibration of either inadequate or contaminated buffer or calibration gas.

13
New cards

What might cause clotting in system?

Mechanical problems with leak/improper fxn pump tubing or valves.

14
New cards

ABG sample troubleshooting air bubbles

Room air bubble causes pH, PO2 and SaO2 to rise and pCO2 to lower due to equilibrium.

15
New cards

ABG sample troubleshooting venous admixture

Risk of puncturing vein causing a decreased PO2, and increased PCO2 and SaO2.

16
New cards

ABG sample troubleshooting metabolism. What increases after prolonged sample duration? What decreases?

Cellular metabolism continues even after blood is collected so prolonged storage will cause decreased pH, pO2 and cGlu, and increased pCO2, cCa2 and cLac.

17
New cards

ABG sample troubleshooting heparin and dilution

anticoagulants dilute sample which can decrease pCO2.