1/16
Looks like no tags are added yet.
Name | Mastery | Learn | Test | Matching | Spaced |
---|
No study sessions yet.
3 components to a blood gas analyzer
Electronic circuitry (displays visuals)
Electrolyte solution (promotes reactions + current)
blood gas electrodes (measures pH, PCO2 and PO2, all other values are calculated)
Measured values of ABG? Calculated ones?
Measured: pH PCO2 and PO2
Calculates all other values
Name of each electrode
pH: Sanz (measured @37C)
PCO2: Severinghaus (measured indirectly)
PO2: Clark/polarographic
Mean definition
Average calculated by dividing sum of all numbers in a group by number of entries
Standard deviation
High + low that represents 95% of where normal population falls when subjected to a certain test, being a measure of variance around the mean.
Aka., measure of spread from the average (mean).
Quality control
Control samples have their range of normal based on ±2 SD’s from the mean, being able to determine mean and SD by running 20 control samples through machine. All 3 levels (acidosis, alkalosis and normal) are run every 24h/shift through vials with a known concentration matching each level.
Calibration
When equipment is adjusted/corrected to mach the control standards to determine accuracy and precision
Accuracy vs precision
Accuracy = mean value of several measurements
Precision = how consistently same measurement will produce same results
Levey-Jennings chart
Functions to tell you whether an ABG machine is ready or not.
Control: shows if all QC (quality control) runs are w/i acceptability (±2 SD’s)
Trend: shows if QC’s start trending towards one side of SD (ex. can still use, but its trending out of control)
Random error: all w/i acceptability except one run (still ok!)
Out of control: 2+ out of acceptability.
What causes a “shift/drift” on L-J chart?
Electrode or sensor malfunction if protein is deposited on membrane or there’s a leak in the membrane, electrolyte, depletion or sensor failure.
What causes it to be “out of control” on L-J chart?
Temperature control failure to maintain 37C or thermometer inaccuracy.
What causes it to be “consistently high or low” on L-J chart?
Improper calibration of either inadequate or contaminated buffer or calibration gas.
What might cause clotting in system?
Mechanical problems with leak/improper fxn pump tubing or valves.
ABG sample troubleshooting air bubbles
Room air bubble causes pH, PO2 and SaO2 to rise and pCO2 to lower due to equilibrium.
ABG sample troubleshooting venous admixture
Risk of puncturing vein causing a decreased PO2, and increased PCO2 and SaO2.
ABG sample troubleshooting metabolism. What increases after prolonged sample duration? What decreases?
Cellular metabolism continues even after blood is collected so prolonged storage will cause decreased pH, pO2 and cGlu, and increased pCO2, cCa2 and cLac.
ABG sample troubleshooting heparin and dilution
anticoagulants dilute sample which can decrease pCO2.