What refers to the output from the tube, exposure to the patients, intensity of radiation, or amount of radiation?
- milligray in air (mGya)
- gray in air (Gya)
- milliroengten (mR)
- roengten (r)
What units is quantity measured in?
Directly proportional
What is the relationship between mAs and intensity?
mA
What is the controlling factor for the number of electrons available to travel across the X-ray tube?
Time
What is the controlling factor for the amount of time the electrons are allowed to travel across the tube?
mAs
mA x time = ?
Double the number of X-ray photons exiting the tube
Double the mAs = ?
Intensity stays the same
What does doubling the mA and halving the time do to the intensity?
They increase
As kVp increases, what happens to penetration and X-ray quantity?
15% rule
What rule states that a 15% increase in kVp will change the IR exposure/density by a factor of two?
- 2x ir exposure unless 1/2 mAs
- 1/2 ir exposure unless 2x mAs
- 15% kVp increase = ?
- 15% kvp decrease = ?
Decrease contrast (low contrast, long gray scale, many shades of gray)
What does a 15% kvp increase do to contrast?
Increase contrast (high contrast, short gray scale, few shades of gray)
What does a 15% kVp decrease do to contrast?
No
Does kVp have a physical effect in digital imaging?
Inverse Square Law
What states that intensity varies inversely with the square of the distance?
intensity 1 / intensity 2 = distance 2^2 / distance 1^2
What is the inverse square formula?
Square Law (Exposure Maintenance Law)
What refers to the amount needed to change technique when the distance changes?
It diverges and only the center part of the beam is used
As SID increases, what happens to the X-ray beam?
- 1/4 IR exposure unless 4x mAs
- 4x IR exposure unless 1/4 mAs
SID Rule Of Thumb
- 2x SID = ?
- 1/2 SID = ?
Intensity decreases
As filtration is added, the beam is "hardened" by absorbing low-energy photons, so what happens to intensity?
Decreasing the patient's skin dose (ESE)
What is the purpose of filtration?
Penetration (Quality / Energy)
What refers to the range or distance that the X-ray beam travels in matter?
Decrease wavelength and increase penetration
What does increased kVp do to wavelength and penetration?
Half Value Layer
What refers to the thickness of absorbing material necessary to reduce the x-ray intensity to 1/2 its original value?
Penetration increases
As filtration increases, what happens to penetration (quality/x-ray energy?
Inherent Filtration
What type of filtration is any material above the tube window that the radiation passes through?
.5 mm Al equivalent
What is the normal amount of inherent filtration in the X-ray tube?
Bronzing of the tube
Why does inherent filtration in the X-ray tube increase with age?
Added Filtration
What type of filtration refers to thin sheets of aluminum placed between the tube housing and the collimator?
1 to 2 mm Al
What is the normal amount of added filtration?
1 mm Al
What is the amount of added filtration in a collimator?
Copper and Aluminum
What are the two types of metal that a compound filter uses?
tube / patient
In a compound filter, copper is closer to the _________ and aluminum is closer to the _________
- .5 / 50
- 1.5 / 50 and 70
- 2.5 / 71
NRCP Total Filtration Limits
- ____ mm Al below ____ kVp
- ____ mm Al between ____ and ____ kVp
- ____ mm Al above _____ kVp
- Increase intensity
- Increase intensity / Increase energy
- Decrease intensity
- Decrease intensity / Increase energy
- Increase mAs = ?
- Increase kVp = ?
- Increase distance = ?
- Increase filtration = ?