1. Connect oxygen hose to oxygen source or turn on local oxygen source
2. Attach circuit to be used
3. Check vaporizer for adequate level of liquid anesthetic
4. Securely occlude patient end of circuit with thumb, palm in hand
5. Completely close pop-off valve
6. Depress fast flush valve until a pressure reading of 40 cm H2O registers on the manometer
7. Observe needle of manometer for declining movement, which would indicate a leak
8. If there is no leak, proceed to step 12
9. If there is a leak, turn to oxygen flow to 200 ml/min
10. If leak is corrected - indicated by a stable needle on the manometer - no further action is needed. Proceed to step 12
11. If leak is not corrected, perform machine maintenance to determine location of leak. Correct leak
12. Without removing the occlusion from the patient end of the circuit, open the pop-off valve. The rebreathing bag should deflate
13. Remove hand, thumb from circuit end
14. Check scavenging system to ensure that connections are intact
15. Completely open pop-off valve