1/52
Looks like no tags are added yet.
Name | Mastery | Learn | Test | Matching | Spaced | Call with Kai | Chat |
|---|
No analytics yet
Send a link to your students to track their progress
How is pixel size calculated, and what is 12 cm ÷ 512?
Pixel size = FOV ÷ matrix. 12 cm ÷ 512 = 0.0234 cm = 0.234 mm.
How do matrix and FOV affect pixel size and spatial resolution?
At the same FOV: larger matrix → smaller pixels → better resolution. At the same matrix: larger FOV → larger pixels → worse resolution.
What is the effect of a smaller pixel or detector element (DEL)?
Better spatial resolution and recorded detail.
What does increasing bit depth improve?
More gray shades and better contrast resolution—not spatial resolution. Gray shades = 2^bit depth.
What is Nyquist frequency?
The highest spatial frequency that can be accurately recorded: 1 ÷ (2 × pixel pitch).
What are the grid conversion factors?
No grid = 1; 5:1 = 2; 6:1 = 3; 8:1 = 4; 10:1 or 12:1 = 5; 16:1 = 6.
What formula is used when changing grids?
New mAs = old mAs × new GCF ÷ old GCF.
Change 10 mAs from a 5:1 grid to a 16:1 grid.
10 × 6 ÷ 2 = 30 mAs.
Change 25 mAs from an 8:1 grid to a 5:1 grid.
25 × 2 ÷ 4 = 12.5 mAs; use approximately 13 mAs.
Change 10 mAs from a 16:1 grid to a 6:1 grid.
10 × 3 ÷ 6 = 5 mAs.
Change 12 mAs with no grid to an 8:1 grid.
12 × 4 ÷ 1 = 48 mAs.
What is the CR imaging plate made from?
A photostimulable phosphor of europium-activated barium fluorohalide.
What is the CR image-production pathway?
X-rays create trapped electrons/latent image → laser releases light → photomultiplier or photodetector converts it to an electrical signal → ADC digitizes it.
What is the direct-conversion DR pathway?
X-rays → electrical charge in amorphous selenium → TFT → ADC. No light-producing scintillator is used.
What is the indirect-conversion DR pathway?
X-rays → light in CsI or GOS scintillator → photodiode converts light to charge → TFT/ADC.
H16 — What is the key difference between direct and indirect DR?
Direct DR converts x-rays directly to charge; indirect DR converts x-rays to light and then charge.
H17 — What do high-pass and low-pass processing do?
High-pass enhances edges/detail but also noise. Low-pass smooths the image and reduces noise but sacrifices detail.
H18 — What does the LUT do?
Maps stored pixel values to displayed gray shades; it primarily controls displayed contrast.
H19 — What does rescaling do?
Standardizes displayed brightness using histogram information; it cannot repair poor acquisition or positioning.
H20 — What are the histogram and VOI?
The histogram represents image exposure values; the VOI selects the relevant anatomy. Poor collimation, centering, or field recognition can cause histogram errors.
H21 — What is linearity versus reproducibility?
Linearity = proportional output using different mA/time settings at equal mAs. Reproducibility = repeated exposures at the same settings produce the same output.
H22 — What are the magnification and SOD formulas?
MF = SID ÷ SOD. SOD = SID − OID.
H23 — What geometric changes improve recorded detail?
Increase SID, decrease OID, and use a smaller focal spot.
H24 — How is motion blur reduced?
Use the shortest practical exposure time; increase mA if needed to maintain the required mAs.
M01 — What does AEC do?
Terminates the exposure after sufficient exit radiation reaches the selected detector, maintaining receptor exposure.
M02 — What is the course rule for full silicone breast implants and AEC?
Turn off AEC and use a manual technique; use minimal appropriate compression.
M03 — What technique is preferred when a pediatric patient does not cover an AEC chamber?
Use a manual technique with lower pediatric kVp and mAs.
M04 — What special pediatric adjustment applies to children younger than six?
Decrease kVp by at least 15% from the adult technique and limit grid use when possible.
M05 — What happens if a dense prosthesis covers an active AEC chamber?
It blocks radiation from the chamber, prolonging exposure and causing overexposure. Select another chamber or use manual technique.
M06 — What is the approximate intensity across the anode-heel effect?
Cathode side ≈ 120%; center = 100%; anode side ≈ 75%.
M07 — How is the heel effect used clinically?
Place thicker anatomy beneath the cathode and thinner anatomy beneath the anode to produce more uniform exposure.
M08 — How are photon number, SNR, and quantum noise related?
More photons → higher SNR and less relative quantum noise. Too few photons produce quantum mottle.
M09 — What produce saturation and fog?
Excessive receptor exposure can cause saturation/clipping; background or stray radiation produces fog.
M10 — How do poor collimation and scatter affect the image?
Poor collimation increases scatter and reduces contrast. Tight collimation and grids reduce scatter.
M11 — What exposure adjustment does a fiberglass cast require?
Fiberglass itself generally requires no change; select technique using the actual measured thickness. Plaster casts require increased exposure.
M12 — What technique adjustment is generally needed for bronchitis?
Use the standard chest technique unless another condition or patient thickness requires adjustment.
M13 — How should technique change for a patient with a BMI around 35?
Increase both kVp and mAs as needed for the increased tissue thickness.
M14 — How should lumbar technique change for a thin, frail patient?
Reduce mAs while generally maintaining appropriate kVp for penetration.
M15 — What is the image-intensifier brightness-gain formula?
Brightness gain = minification gain × flux gain. Minification gain = (input-phosphor diameter ÷ output-phosphor diameter)².
M16 — What does a contrast-detail phantom test?
The system's ability to demonstrate low-contrast objects of different sizes—contrast resolution.
M17 — What does an SNR test evaluate?
The relationship between useful signal and noise; higher SNR means a cleaner image.
M18 — What is the purpose of a daily CR phantom image?
To verify consistent receptor/reader performance and identify developing artifacts.
L01 — What is the proper rule for anatomical markers?
Place the correct lead side marker inside the collimated field during exposure; a postprocessing annotation is not a proper substitute.
L02 — What does a compensating filter do?
Equalizes receptor exposure across anatomy with unequal thickness.
L03 — What usually causes horizontal banding on a DR image?
Failed or improperly calibrated detector elements or detector rows.
L04 — How should lead aprons be tested and protected?
Inspect radiographically or fluoroscopically at least annually; never fold them because folding causes cracks.
L05 — What does the focusing cup do?
The negatively charged focusing cup directs electrons from the cathode filament toward the anode focal spot.
L06 — What are the major transformer functions?
An iron-core step-up transformer increases voltage; a step-down transformer lowers voltage and raises filament current; the autotransformer selects voltage/kVp.
L07 — What are voltage and amperage?
Voltage is the electrical force that pushes electrons; amperage is the rate of electron flow.
L08 — What are oscillation and frequency?
Oscillation is cyclic positive-to-negative electrical variation; frequency is the number of cycles per second, measured in hertz.
L09 — What does monitor dynamic range determine?
The range of brightness and contrast levels the monitor can display.
L10 — What are typical image-intensifier and monitor resolutions?
Image intensifier ≈ 4-6 lp/mm; monitor ≈ 1-2 lp/mm. Standard monitors use about 525 lines and high-resolution monitors about 1,024.
L11 — What is the difference between lossless and lossy image compression?
Both reduce file size. Lossless preserves all data; lossy permanently discards data and excessive compression can destroy diagnostic detail.