What error tolerance is allowed for Centering?

1%

2
New cards

What error tolerance is allowed for Distance

2%

3
New cards

What error tolerance is allowed for Collimation

2%

4
New cards

What error tolerance is allowed for Exposure Timer

5%

5
New cards

What error tolerance is allowed for Exposure Reproducibility

5%

6
New cards

What error tolerance is allowed for kVp Calibration

5%

7
New cards

What error tolerance is allowed for Exposure Linearity

10%

8
New cards

What quality control is the ability of an x-ray tube to produce the SAME radiation exposure, or intensity, with the SAME exposure settings

Exposure Reproducibility

9
New cards

What quality control is the ability of an x-ray tube to produce the same radiation exposure with the same mAs, but varying mA and time settings

Exposure Linearity

10
New cards

What is called when detectors are calibrated to produce the same response to the same exposure levels?

Detector Calibration

11
New cards

What is it called when CR readers should be calibrated to produce the same response to the same exposure levels 

Reader Calibration

12
New cards

What is it called when a small lead block is exposed, then the receptor is read, erased, exposed again, and read again. This checks for any residual image of the block

Erasure thoroughness

13
New cards

What is called when the image receptor is exposed, processed, and reviewed for an artifacts or changes in radiation response?

Plate Uniformity 

14
New cards

What is it called when the spatial resolution of image receptors is checked by exposing the line-pairs tool. Resolution is checked vertically, horizontally, in the center of the IR and on the periphery of the IR

Spatial Resolution 

15
New cards

Which QC test measures the uGy/mAs at adjacent mA stations? 

Exposure Linearity 

16
New cards

What QC test requires multiple exposures with identical techniques?

Exposure Reproducibiliity 

17
New cards

What QC test checks for voltage drift?

kVp Calibration

18
New cards

Coincidence testing is an evaluation of

Collimation

19
New cards

What QC test must be performed twice per year?

Collimation

20
New cards

What image receptor test checks for ghost image?

Erasure Thoroughness

21
New cards

What image receptor test uses a small lead block?

Erasure Thoroughness

22
New cards

What image receptor test uses a line-pairs tool?

Spatial Resolution

23
New cards

What QC process ensures detectors produce the same responses to the same exposure levels

Detector Callibration

24
New cards

The brightness of PACS reading monitor is checked with a

Luminance Meter

25
New cards

During the AEC test, increasing the patient thickness should not affect

Receptor Exposure

26
New cards

If the indicated SID is 100 cm, the actual SID cannot be more than

102 cm