Why does PDD increase with increasing SSD?
The effect of z decreases (inverse-square factor)
The meyneord factor works well for what sized fields and depths?
Small (small scatter contribution)
What does the meyneord factor overestimate?
increase in PDD with increased SSD
Backscatter factor
TAR at dmax
TAR vs. Mayneord factor
TAR considers scatter
When is the SSD setup used for dose calculations?
single field treatments
When is the SAD setup used for dose calculations?
multiple field treatments
If the field size is increased, does beam output increase or decrease?
increase
rc
the side of the equivalent square for the collimator field size
rd
the side of the equivalent square for the field size in the phantom
Does TMR increase or decrease with increasing depth?
decrease
Does TMR increase or decrease with increasing field size?
increase
Does TMR increase or decrease with increasing energy?
increase
How is PDD and TMR related?
inverse square correction
TPR
dose ratios for different depths but same point distance from the source
Which is an SSD setup?
A