1/87
Looks like no tags are added yet.
Name | Mastery | Learn | Test | Matching | Spaced |
---|
No study sessions yet.
1895
When were x-rays discovered?
Wilhem Conrad Roentgen
-discovered x-rays in 1895
-successfully took radiograph of wife's hand using x-ray tube
-received the first Nobel Prize in physics for his discovery in 1901
x-ray
-energy
-considered form of electromagnetic radiation
-travel in straight lines but may change direction
-composed of energy (quanta) bundles called photons carried along by wave
electromagnetic radiation
-method of transporting energy through space
-distinguished by wavelength, frequency, and energy
-how energy passes through space and matter
wavelength
-distance between two consecutive points
-measured in nanometers
Frequency
-number of wave cycles that pass a single stationary point per second
-measured in Hz
high frequency
shorter wavelength=
shorter wavelength
greater energy=
greater energy
more penetrating power=
xrays result
fast moving particles bump into matter
cathode
-negative electrode
-source for x-ray
-contains wire filament made of tungsten thats heated by a low-energy circuit
-heat excited electrons within atoms of the tungsten filament
anode
-postive electrode
-electrons pulled towards bc of their negative charge
-large amount of energy produced when electrons interact with it
-target within the anode that the electrons hit
-draws heat away from target
wire filament is heated
electrons released when
1%
how much energy produced forms electrons?
99%
how much energy is released as heat?
number of electrons produced
-depends on amount of heat applied to the filament
-under control of mAs and time which controls amount of heat applied to filament
Acceleration
controlled by kVp
kilovoltage
the amount of electrical energy being applied to the anode and cathode to accelerate the electrons from the cathode to the anode
kVp
-determines the penetrating power of the x-ray beams
focal spot
-area on the anode that the electrons hit
-size is related to sharpness of image
-the smaller the sharper the image
production of controlled x-ray beam
What is the purpose of the x-ray tube?
SID (source to image distance)
-distance from the x-ray source (anode target) to the film
-generally 36 to 40 inches
Object-Film Distance
-the distance between object being imaged and the film
radiographic density
-the degree of blackness on a radiograph
-refers to the number of x-rays that the film had been exposed to in a given areas
-the blacker the film the more x-rays that touched it
radiographic contrast
-the difference in radiographic density between tow adjacent areas on the radiograph
-bone vs. soft tissue
tube stand
-supports x-ray tube
-models can vary from tabletop stands to large stands mounted on the celing
collimotor
-most versatile type contains lead shutters
-light source helps the radiograph tech view the x ray field size type
-light also helps to accurately position the patient
-use smallest field size necessary to obtain diagnostic image
grid
-purpose is to absorb scatter radiation to improve the quality of the image (prevents fuzziness)
-lives under table top
-essentially a series of lead strips that absorb radiation and spaces to allow x rays to pass through
grid ratio
-height of the lead strips compared to the width of the space between
-height is usually larger and listed first
"bucky"
-place holder for image receptor
-located under the x ray table or vertically in the wall mounted unit
-not recommended for portable units
-provides mobility for different size patients
-contains a built in grid to absorb scatter
-no grid lines on film
As Low As Reasonably Achievable
What does ALARA stand for?
6 ways to achieve ALARA
-dosimetry badges
-PPE and protective equipment
-patient positioning aids
-technique parts
-quality control measures
-emergency procedures
Primary x-ray beam
a body part is placed in the primary beam due to technical error
scatter
-x-rays which strike the object of interest and are bounced in another direction
-cannot be prevented so personnel must use leaded PPE
upper body and head
Which parts of the body are at greatest risk for exposure?
leaking radiation
-produced by older, poorly maintained machines when quality control and assurance programs are not routinely practiced
-damaged portable
3 rules of radiology safety
-shortest time
-greatest distance
-maximum shielding
secondary and scatter
Which radiation is leaded PPE designed to protect against?
25%
How much of the primary beam is absorbed by leaded PPE?
visual inspection
check for visible tears and cracks
every 3 months
How often should manual inspection of PPE be performed?
manual inspection
-external and internal surfaces
-if a defect occurs in main body of leaded PPE the item must be replaced or repaired before further use
-if defect occurs at the edge, the item should be marked and monitored for changes
-take radiographs to evaluate internal surfaces
Cleaning PPE
-never use products containing bleach
-never machine wash or dry clean leaded PPE
apron
-never fold for storage as it results in cracks over time
-hang vertically over a round surface not less than 3 cm in diameter or lay flat for storage
gloves
-store in a way that air can circulate inside them; this eliminates moisture which can build up inside them
-store vertically on holder
dosimeter purpose
measures the amount of radiation exposure (dose) an individual receives over time
dosimeter donts
-NEVER wear your badge outside of designated area
-NEVER expose badges to the path of the direct beam
-NEVER share badges
dosimeters dos
-store badges outside of the radiology room in areas where they are not likely to be exposed to sunlight, heat, or additional radiation sources
-use a separate badge for each site where radiation occurs
pinpoint spikes/defects
Why should you use separate badges for each site where radiation occurs?
personnel safety rules
-minimize number of people in the room when imaging
-rotate personnel regularly
-use leaded shielding if you must be in the room but are not manually restraining the patient
-pregnant workers should not restrain patients for imaging, especially in the first trimester
-fetal dosimeter required if pregnant worker chooses to restrain
-people 18 and under should not be allowed in radiology room during imaging except for medical purpses
collimation
-controls size of the x-ray field
-reduces scatter
-improves the clarity of the image
when restraining
-never look directly into the primary beam with or without goggles
-avoid sitting or leaning on the table when restraining
-always stand at the ends of the table
positioning aides
-minimize the number of people required to provide manual restraint
-sandbags
-v troughts
-foam wedges
-limb ties
technique charts
-predetermined technique settings for an area of interest based on tissue thickness
-uses the minimum amount of radiation to take a diagnostic radiograph
-reduces amount of scatter
patient restraint
-minimizes patient motion
-proper sedation of anesthesia should be provided
-positioning aides
-manual restraint
breeding and pregnant animals
-gonadal shields
-breast shields
-for pregnant animas: x-rays for emergency only
ultrasound
what should be used in place of an x-ray in pregnant animals?
care of x ray tube
-warm up machine and tube prior to use
-take care when operating portable units to prevent damage
-the tungsten filament can be damaged when dropped if you drop the unit
-monitor for oil leaks
oil leaks
-oil absorbs heat from x-ray tube so leaking leads to overheating and imminent destruction
-cracks can result in leakage of radiation in the form of scatter
40 in
what is the preferred SID?
intensity
When distance is changed _________________ is also changed
less intensity
greater distance=
quality control and assurance
-does not guarantee complete radiation safety for patients and staff
-errors in image quality and clarity can still occur even if the machine is functioning properly
-SID
-bucky-grid centering
-light field/xray field alignment
quality assurance
checks personnel for error
quality control
checks machine for error
nine penny test
the 9th penny is placed in the lower left hand corner to provide orientation of the film to check for collimation
anode heel effect
-variation in number of x-rays: intensity of the xrays of the primary x-ray field as they diverge from point of origin
-cannot be prevented
-decrease of intensity occurs as as you move from cathode to anode which causes a decrease in radiographic density causing a poor image quality
poor images
What happens as a result of a greater anode heel effect?
put the thickest part of the patient towards cathode side
how should you position the patient to decrease the anode heel effect?
more intensity
the cathode has more or less intensity than the anode side?
SID effect (on anode heel)
-decreased SID=increased anode heel
-shorter distances cause a greater variation in x-ray intensity
-the higher the SID the less anode heel effect
x-ray field size
the greater the size, the greater the anode heel effect will be
anode angle
-the smaller it is the greater the anode heel effect will be
-smaller focal spot, greater amount of x-rays to pass through anode
magnification
-occurs when the anatomy is not positioned as close to the image receptor as possible
-the anatomy that is further away from the image receptor appears larger than it really is
-alters appearance and causes misdiagnosis or hides pathology
geometric distortion
-position of anatomy in regards to image receptor
-changes radiographic density and contrast
-magnification
-foreshortening
-elongation
foreshortening
-occurs when the anatomy is not positioned as close to the image receptor as possible, and is also not parallel with the image receptor
-can get magnification and thickness changes
technical errors
-patient motion is most common
-results in retake
-use the shortest exposure time possible with sedation or anesthesia when available and appropriate to prevent patient motion
-human anatomy present
-collars, leashes, tags on the patient
-double exposures
-film flogging
-static electricity on the film
-thumbnail crescents
-fingerprints
intentional artifacts
-external fixators, splints, casts are visible on the film
-buckshot
-can't be prevented
-either want it there or unaware it is there
manubrium
What is used as the cranial border when collimating a thoracic x-ray?
xiphoid
What is used as the caudal border when collomating a thoracic x-ray?
sternum
What is used to center a VD x-ray of the thorax?
spine of scapula
What is used to center a thoracic x-ray?
scapula border
What is used as the cranial border in a lateral x-ray of the thorax?
standard thoracic views
-right lateral
-ventrodorsal
standard cardiac views
-ventrodorsal
-dorsoventral
-lateral
standard lung views
-right lateral
-left lateral
-ventrodorsal
DV
Which view is best for the heart?
VD
Which view is best for the lungs?