43 Terms

1
New cards

Absorbance is calculated from transmittance by which formula? (L1S8)

A = 2-log %T

2
New cards

Which statement best describes Beer’s Law? (L1S9)

Absorbance is directly proportional to the concentration

3
New cards

A fluorescing molecule absorbs light then emits light of (L2S6)

A longer wavelength (lower energy)

4
New cards

Which statement about atomic absorption (AA) spectrophotometry is correct? (L2S3)

Measures absorbance of light by atoms, not molecules

5
New cards

Which statement about nephelometry is correct? (L2S19)

Measures scattered light blocked at 180 degrees from incident light

6
New cards

Milky, lipemic serum run on a nephelometric estrogen assay: likely outcome? (L2S23)

Lipemia causes background light scatter/interference

7
New cards

Testing 50%T on either side of peak transmittance assesses a filter’s

Bandpass (spectral bandwitdth)

8
New cards

125 mg/dL glucose std, Abs 0.20. Patient ABs 0.24 Conc = ? (L1S25)

150 mg/dL

9
New cards

Protein calibrator 6 g/dL, Abs 0.5. Patient Abs 0.6. Conc = ? (L1S25)

7.2 g/dL

10
New cards

Which wavelength falls within the near-ultraviolet region? (L1S5)

220 nm

11
New cards

Diluted 1:10, very high FT4 fluorometric result reads “<0.05” Why? (L2S23)

Dilution wasn’t enough, hook effect suppressed signal

12
New cards

Fluorescent technique using rotation of small vs large labeled molecules? (L2S11)

Fluorescence polarization immunoassay (FPIA)

13
New cards

Which component isolates a specific wavelength, excluding others? (L1S17)

Monochromator

14
New cards

AAS is frequently used in clinical labs to measure (L2S14)

Elements and trace metals

15
New cards

The light source in an atomic absorption spectrophotometer is typically a (L2S3)

Hollow-cathode lamp

16
New cards

Which technique is more sensitive at low analyte concentration, and why? (L2S12)

Fluorescence; measured against a dark (zero) background

17
New cards

Which optical technique is NOT considered an emission technique? (L2S3-5)

Atomic absorption spectrophotometry

18
New cards

Difference between max excitation and max emission wavelength is called (L2S8)

Stokes shift

19
New cards

Wavelengths of the electromagnetic spectrum are traditionally measured in (L1S4)

Nanometers (nm)

20
New cards

True or false: The visible spectrum spans about 380-750nm (L1S5)

True

21
New cards

Which electrochemical technique routinely measures serum/plasma chloride?

Ion-selective electrode (potentiometry)

22
New cards

Measuring the voltage difference between two half-cells is called (L3S17)

Potentiometry

23
New cards

A voltage potential from chemical equilibrium of electron transfer is (L3S18)

Redox potential

24
New cards

A common glass reference electrode for pH measurement is the ____ electrode (L3S25)

Calomel

25
New cards

Which statement correctly describes a PCO2 electrode? (L3S29)

CO2 diffuses through a membrane, forming carbonic acid

26
New cards

Which neutral antibiotic in a PVC membrane makes a K+ selective electrode? (L3S28)

Valinomycin

27
New cards

That type of K+ ISE belongs to which category of electrodes? (L3S24)

Polymer membrane electrodes

28
New cards

Potentiometric measurements (galvanic cells) consist of (L3S16)

Two conducting electrodes joined by electrolyte

29
New cards

Voltammetry and amperometry are based on which cell type? (L3S8)

Electrolytic cell

30
New cards

The predominant potentiometric electrode in clinical labs is the (L3S24)

Polymer membrane ISE

31
New cards

Most immunoassays use which immunoglobulin for antigen specificity?

IgG

32
New cards

Two examples of luminescent labels used in chemiluminescence are (L2S17)

Luminol and acridinium esters

33
New cards

If a signal reads below the analytic measurement range, you should

Repeat testing and report <AMR

34
New cards

Which statement best contrasts galvanic and electrolytic cells? (L3S6)

Galvanic is spontaneous; electrolytic needs voltage

35
New cards

A substance that loses/donates electrons is called a (L3S9-11)

Reductant

36
New cards

Max potential difference between electrodes at zero current defines (L3S19)

Electromotive force (EMF)

37
New cards

The formula for calculating absorbance of a solution is (L1S13)

Absorptivity x light path x concentration

38
New cards

Correct light pathway through a single-beam spectrophotometer? (L1S14-15)

Source → monochromator → sample → detector → readout

39
New cards

What is the purpose of a monochromator? (L1S21)

Isolate a single wavelength/band of light

40
New cards

Stray light in a spectrophotometer limits accurate measurement of (L1S27)

Absorbance at the upper end of linearity

41
New cards

175 mg/dL glucose std, Abs 0.655. Patient Abs 0.445. Conc = (l1S28)

119 mg/dL

42
New cards

Which component is found only in a fluorometer, not a spectrophotometer? (L2S15)

An excitation source

43
New cards

Current-based technique; gold standard for chloride in serum/plasma (L3S30)

Coulometry