1/29
Looks like no tags are added yet.
Name | Mastery | Learn | Test | Matching | Spaced | Call with Kai | Chat |
|---|
No analytics yet
Send a link to your students to track their progress
Bright-field microscopy
The most commonly used type of light microscopy in introductory microbiology and clinical laboratories.
Types of light microscopy
The four major types discussed in the manual are bright-field, dark-field, fluorescence, and phase-contrast microscopy.
Image production in bright-field microscopy
It uses light transmitted through the specimen to produce an image.
Specimen appearance in bright-field microscopy
A specimen appears shadowy against a bright background.
Use of stains in bright-field microscopy
Stains are commonly used because most biological specimens are transparent, and staining improves contrast between the specimen and background.
Disadvantage of staining microorganisms
The staining process usually kills the cells.
Condenser lens
It concentrates light and makes illumination of the specimen more uniform.
Objective lens
It uses refracted light from the specimen to produce a magnified real image.
Ocular lens
It magnifies the real image again to produce a virtual image seen by the eye.
Total magnification formula
Total magnification = objective magnification × ocular magnification.
Binocular microscope magnification
The magnifications of both ocular lenses aren't multiplied together because each eye views the image through one ocular lens.
Practical limit of magnification
Approximately 1300× for a light microscope.
Greater magnification utility
Greater magnification isn't always useful because image clarity becomes increasingly difficult to maintain.
Resolution
The clarity of an image and the ability to distinguish two points as separate objects.
Limit of resolution
The minimum distance two points must be apart for the microscope to distinguish them as separate.
Resolution and limit of resolution relationship
Resolution is better when the limit of resolution is smaller.
Best limit of resolution for light microscope
About 0.2 μm.
Wavelength's effect on resolution
Shorter wavelengths produce better resolution than longer wavelengths when other variables remain constant.
Wavelength as limiting factor for resolution
The wavelength of light limits how closely spaced two points can be while still being distinguished as separate.
Numerical aperture (NA)
A measure of a lens's ability to capture light from the specimen and use it to form an image.
Units of numerical aperture
Numerical aperture has no units.
Effect of increasing numerical aperture on resolution
It improves resolution by making the limit of resolution smaller.
Immersion oil
It increases the numerical aperture of the oil-immersion objective, improving resolution.
Objective lens for immersion oil
The oil-immersion objective is designed to be used with immersion oil.
Purpose of the iris diaphragm
To adjust the amount of light entering the specimen and help optimize illumination, contrast, and image quality.
Action with iris diaphragm as magnification increases
It should generally be opened more because a smaller portion of the light beam enters the lens.
Coarse-focus knob use
It is used for bringing the image into approximate focus.
Fine-focus knob use
It is used for bringing the image into sharpest focus.
Nosepiece ring usage
The nosepiece ring should be used to change objectives to avoid damaging them.
Parfocal definition
The objectives are designed so that once a specimen is focused with one objective, only a slight adjustment of fine focus should be needed after switching objectives.