What is MDCT
Multiple detector act 2 or more parallel rows
Advantages of MDCT
more spices
Less artefacts
Pediatric scanning
Thinner slice improved z axis resolution
Good for trauma
Different detector configuration
Uniform - all same
Non uniform - all different
Hybrid - variety of sizes has both uniform/ non uniform
Pitch
Speed table moves through gantry
If beam pitch more than 1:
Table travels more wide the gaps : higher pitch : - low image quality - faster - less dose
If beam pitch is less than 1
Overlapping - lower pitch - high patient does - higher axial z- resolution - better image quality
Come beam acquisition and artefacts
More detectors more slices / wider beam the more detectors the more cones can have more artefacts because tube rotates objects may appear off centre
Come beam Interpolation: tilted reconstruction
Filling the gaps in at angle to make axial- straight line
Cone beam interpolation/ Feldman algorithm :
Filling gaps to a 3rd dimensional shape: cube
Characteristics of a good CT image:
Visable anatomical structures
Good image congrats
Good spatial resolution
Fast scanning
Low dose
Scanning parameters
volatage
Tube current
Scan time
Slice thickness
Pitch
Patient size
Noise ?
Effects contrast resolution
Number of photons produced are directly proportion to tube current, scan time , slice width
To decrease noise you need to
Increase:
Tube current
Scan time
Slice width
Factors effecting mouse
Patient size
Scan time
Noise filters slice thickened
Pitch
Beam filtration
Spatial resolution
How far 2 objects have to be before they are seen as separate
Large detectors identify gaps but seen as one object - low spatial resolution
smaller detectors identifying gaps but display them as separate- high spstial resolution
Transaxial resolution
Across patient
Z-axis resolution
Along patient
CT necessary
X -ray taken angles for cross sectional images different slices
Eliminate superimposition
Converts x-ray to electrical signal
CT detector characteristics
Small in size for spatial resolution
High detection effeciencu
High density
Fast response time
High stability for high speeds rotating gantry
Tube characteristics
cooling systems
Larger and thicker anodes to absorb dissipate large amounts
Modified cathode
Focal spot allowing focus position
Valium used as lubricant
Metal ceramic tube instead of glass envelope
Slip rings
Allows transmission of power and electrical signal from stationary to rotating structure
Image reconstruction
Converts the numbers from ray sums to image
Back projection
Different number of gray scale from the images joins together - data converted into back projection to prouduce image
Iterative reconstruction
Similar to back projection
Speed up and improved reconstruction
Improves image quality
Image matrix
Pixel in images
Pixel reprints thickness - voxel ( volume)
Voxel assigned to grey scale measured in hounded hints 0-1000
CT window
Changing brightness and contrast of image by altering hounsfield units
Window width
Contrast - measure range of act numbers
Window level
Brightness - midpoint of range of ct number displayed
Axial ct
Patient still
Stop and start while x-ray
Helical Ct
Around the body continues having added detector
Advantage of helical scan
Quicker
Longer scans
Pitch speed can be controlled
Overlapping allows for reconstruction
Slide sensitivity
Extent to which slice contributed to volume data
Wider slice is
High pitch - less intensity - less image resolution
narrow slice
Small pitch - intensity increased - spatial resolution
Ct artefacts
Physics based
Patient based
Scanner based
Helical and multi slice artefacts
Example of helical and multislice artefacts
Ring artifact
Windmill effect
Gemortical distortion
Example physical artefacts
Beam hardening causes streaking and upping artefacts happens in dense part of body
Lower energy beams absorbed or scattered
Attenuation coefficient
Fraction of beam x-rays that is absorbed or scattered per unit thickness of the absorber
CTDI
Volume Ct dose index not patient dose related to intensity of x-ray beam
DLP
Dose length product - measure of total iraddation of patient