data collected from the scan, there is no HU unit assigned. Because of this it cannot be reformatted to 2D or 3D. However this information can be stored and used to reconstruct in different formats
image data
after raw data is given a HU value. can be used for reformatting iamges in other planes.
what is retrospective reconstruction
information can be stored and used to reconstruct in different formats. this is in raw data
MDCT: Pitch
uTable movement during each 3600 rotation (1 revolution of x-ray tube) Beam width
beam width formula
(BW= number of slices x slice thickness)
what is pitch
parameter that describes the movement of the table throughout helical scan. essentially the travel distance of the table per 360 roation o fthe xray tube DIVIDED by the collimated xray beam width
what does pitch affect
dose
quality image
scan time
what is the ideal pitch
1:1 or just 1
is pitch usually changed from pre-programmed standards
no but can be
increasing pitch…
uLarger volume of tissue covered
uDecreases patient dose
uIncreases table speed
uDecreases image quality
anatomy covered in 1 rotation
decreasing pitch…
uSmaller volume of tissue covered
uIncreases patient dose
uDecreases table speed
uIncreases image quality
anatomy not covered in 1 rotation
ideally pitch should remain below…why?
1.5. maintain image quality and reduce patient dose
T/F binning occurs when multiple detectors are combined provide the desired overall section thickness
true
x occurs when raw data is retrieved to create new images
Retrospective Reconstruction