1/34
Looks like no tags are added yet.
Name | Mastery | Learn | Test | Matching | Spaced |
---|
No study sessions yet.
What is the damper?
alkaline buffer reagent for CO2 measurement that prevents bubbles from entering the flow cell
What tests are run on the Modular Chemistry side of the analyzer?
Urea, creatinine, phosphorus, glucose, total protein, an albumin
Ion-selective electrode
the difference in potential between the reference and measuring electrode can be used to calculate the concentration of the ion in solution
What analytes are measured with an ISE?
Na, K, CL, CO2, and CaWh
What coats sodiums electrode?
Aluminum silicate glass
What coats potassiums electrode?
Valinomycin
What coats chlorides electrode?
Ag/AgCl
What coats Calciums electrode?
ionophore membrane
What coats CO2s electrode?
pH rate change
What is a flow cell?
an acrylic block that holds electrodes for ISE measurements
ISE reference
reference reading is taken after each sample and is subtracted from the buffer/sample dilution reading
ISE buffer
used to dilute patient sample
prevents large pH fluctuations due to acidity/alkalinity of sample
suffer contains known concentrations of analytes
What are the advantages of indirect measurement?
uses less sample
analysis is quicker
more consistent measuring environment
reduces interference
less protein-coating
Urea MC principles
urease reagent
gold conductivity electrode - measures rate of increase in conductivity
Creatinine MC principles
Colorimetric: based on jaffe reaction with alkaline picrate
Glucose MC principles
polarographic electrode
measures rate of oxygen consumption in a glucose oxidase reaction
Phosphorus MC principles
Colormetric: ammonium molybdate reagent
Total Protein MC principles
Colormetric: Biuret reaction
Albumin MC principles
Colorimetric: dye binding with BCP
Creatinine MC reagent
combine picric acid with buffer, mix → allow to de-gas overnight
How often do you calibrate electrolytes?
every day
How often do you calibrate Glucose, Urea, Creatinine, and phosphate?
every 48hrs
How often do you calibrate TPm and ALBm?
monthly
Range checks for what?
Accuracy
Back to back checks for what?
Precision (reproducibility)
Span checks for what?
Sensitivity (spread between calibrator levels)
What does failing calibration due to range likely mean?
switched calibrator cups
insufficient, expired, or incorrectly prepared reagents
What does failing calibration due to Back to back likely mean?
bubbles in reagent lines or sample
What does failing calibration due to Span likely mean?
Calibrators: incorrect, reverse positions, contaminated
Reagents: expired, precipitated, temperature
Electrode connections or worn membrane
Reflectance photometry
light reflected form a surface is used to measure the amount of product from a colorimetric reaction → amount of light reflected decreased in proportion to the intensity of colour produced
You need one reflectometer for each test zone plus 2 more, what are their roles?
one to correct for interference due to the colour of the urine
one that monitors and check that the strip is aligned with the instrument correctly
Reflection Density (RD or Dr)
need to determine ratio of light reflected by the surface containing the chromogenic reaction to the intensity of light reflected from a reference surface
Calibrating reflectance tests
white reference standard test used
white surface gives minimum absorbance
all light is reflected or set to 1 = 0abs
Black standard gives maximum absorbance or ) reflectance = 2.0 OD
Reflectometer test reactions
colormetric/enzymatic
endpoint
immunoassay
Potentiometric test reactions
an electrochemical technique used to measure the potential (voltage) between two electrodes in a solution to determine the concentration of an ion or the activity of a species.