1/20
Looks like no tags are added yet.
Name | Mastery | Learn | Test | Matching | Spaced | Call with Kai |
|---|
No analytics yet
Send a link to your students to track their progress
X-ray Energy
kVp (highest voltage applied across tube) determines maximum energy of electrons and thus x-ray photons (Emax = kVp)
Average photon energy in clinical x-ray beams is ~1/3 to 1/2 of maximum energy
X-ray Emission Spectrum
total number of x-rays emitted is equivalent to area under the curve of x-ray emission spectrum
always bell-shaped
position along the energy (x) axis can change
larger area under the curve = higher x-ray intensity (quantity)
farther to the right a spectrum is, the higher the effective energy (quality) of x-ray beam
Beam Parameters
X-ray tube output controls beam’s quantity and quality
Rule: When intensity/quality (amplitude) is effected it is always for bremsstrahlung and characteristic x-rays
Factors Effects on Emission Spectra:
Tube Current effects intensity
Tube voltage effects intensity and energy of bremsstrahlung x-rays
Filtration effects intensity and energy of bremsstrahlung x-rays; most effective at low energy
Target material effects intensity and energy of characteristic x-rays peaks
Tube Current
Number of electrons passing from cathode to anode, mA
Doubling mA = twice as many electrons flow = twice as many x-rays at every energy (amplitude of each point on emission spectra curve is doubled)
Increasing tube current = better image quality (because image noise decreased) but increased dose
X-ray Quantity (Intensity) - mGy(a)
Number of x-rays in beam
Measured as area under spectrum curve
Measured in units of air kerma (energy transferred to electrons in air), mGya which is proportional to number of ion pairs produced in air by a quantity of x-rays
mAs
determines total radiation quantity used during an exam
= tube current(mA) x time (sec)
Changes x-ray quantity (mGya) and image receptor exposure linearly - Doubling mAs doubles number of electrons (thus number of photons) and thus number of photons hitting image receptor
Tube Output
mGy(a)/mAs
how much radiation the tube produces per mAs at a specific kVp and distance
Signal-to-Noise Ratio (SNR)
Ratio of mean photon intensity (signal) to the random fluctuation in number of detected photons (noise)
SNR =N / √N
Noise = √counts because discrete photons follow Poisson statistics
High SNR = high resolution because signal (N) increases faster than noise (√N)
Image Noise
Mostly caused by a lack of photons (low SNR) reaching receptor
Noise is reduced by increaisng mAs, which increases x-ray quantity to provide more signal to receptor
However caution is needed as increasing mAs will increase dose
Although increasing kVp (thus photon energy) can reduce noise it also increases scatter and reduces overall subject contrast
X-ray Quality (Energy)
Increasing energy increases penetrability
High penetrability = high quality/harder beam
Half Value Layer (HVL)
measures beam quality
the thickness of an absorber needed to reduce beam intensity by half
Image Receptor Exposure
Number of x-rays that reach image receptor
X-rays diverge, speading over larger areas, as they move away from focal spot (source)
Intensity (x-ray quantity) and image receptor exposure = 1/d2 or changes by (d1/d2)2
where d = distance from focal spot
Increasing Tube voltage (kVp) Effects On Quality and Quantity
Increases Quality (position) - faster acceleration, electrons hit target with higher kinetic energy, higher photon mean and max energy
Increases Quantity (amplitude) - higher energy electrons have more interactions with target before losing energy
*No effect on position of characteristic x-rays
Other effects of Increasing Tube voltage (kVp)
Increases HVL
Decreases image noise
Increased forward direction of Bremsstrahlung photons relative to incident electron beam
Increases relative dose because increase in photon number outweighs increase in pentrability (which woud usually decrease dose as more photons pass unabsorbed) due to secondary ionisations depositing more energy in body
Tube voltage (kVp) Equations
X-ray quantity proportional to (kVp)2/changes by (kVp2/kVp1)2
Relative dose proportional to (kVp)3/changes by (kVp2/kVp1)3
Image receptor exposure proportional to (kVp)5/changes by (kVp2/kVp1)5
15% kVp Rule
If you increase kVp by 15% (x1.15), you must halve mAs (and vice versa) to keep image receptor exposure the same
If you decrease kVp by 15% (x0.85), you must double mAs (and vice versa) to keep image receptor exposure the same
Because (1.15)5 ≈ 2, (0.85)5 ≈ 0.5
15% kVp Rule - Application
used to change image contrast or reduce patient dose while maintaining consistent image density (degree of blackening)
15% kVp Rule - Image Contrast
Increasing kVp by 15% produces longer scale of contrast (more shades of grey, lower contrast, more penetration power to visualise deeper structures)
Decreasing kVp by 15% produces shorter scale of contrast (higher contrast between air, tissue, and bone)
15% kVp Rule - Dose
Increasing kVp by 15% and halving mAs reduces patient dose, as higher-energy photons are more likely to pass through patient without being absorbed
What does 15% Rule Not Account For?
Beam hardening
Detector response
Scattering
15% kVp Rule - Digital Radiography Limitations
rule was developed for film–screen radiography
not needed to maintain image brightness because of wide detector dynamic range (ability to produce images over large exposure range with less obvious exposure errors) and automatic brightness adjustment by processing algorithms
radiographers increase exposure factors to reduce noise which increases patient dose, but over-exposure is now automatically adjusted/hidden to look acceptable
standard image protocols and automatic exposure control (AEC) terminates exposure when sufficient detector signal reached