1/64
Looks like no tags are added yet.
Name | Mastery | Learn | Test | Matching | Spaced | Call with Kai |
|---|
No analytics yet
Send a link to your students to track their progress
Continuous Quality Improvement (CQI)
ongoing process focused on evaluating and improving performance overtime
proactive problem solving
CQI attempts to _____ the problems
anticipate
standards of quality are set by the___ and measured by _____
FDA, JC and PA DOH
The standards of quality does what
specifies requirements for a quality management system
demonstrates the ability to consistently provide products and services
ASRT involvement in Quality Control
participates in ongoing quality assurance programs
QC tests and monitoring, Preventive Maintenance and Corrective Maintenance
Quality control test of equipment is
the calibration and monitoring of equipment and repeat analysis
what are the 4 components of a QC program
acceptance
baseline
correctable deviations
documentation
Preventative Maintenance Procedure
testing is done on a semi-annual basis
goal of preventing breakdowns due to equipment failing without warning signs detectable by monitoring
Record Keeping
data must be recorded after each performance test and kept on file
“if it wasn’t documented it didn’t happen”
What are the 2 levels of evaluation?
1st= results from the monitoring procedures- used to evaluate performance of equipment
2nd= evaluating the effectiveness of their program
What is a quality assurance program?
organized effort to ensure that images are produced are of high quality, consistently provide adequate diagnostic info at the lowest possible cost and with the least possible exposure (WHO)
What does quality assurance do?
prevent problems; plans and policies
fixes the system
what does quality control do?
detect problems; checks and tests
finds the issues
what are the 2 aspects of image quality control
exposure indicator accuracy (EI or DI)
image integrity (verification that images are saved)
Who is the first line of defense in preventing, recognizing and reporting QC issues?
Techs
Image receptor maintenance
inspection would be done: weekly
cleaning should be done: monthly
RH erasure should be done: weekly
CR acceptable erasure is an EI value of _____ in the re-exposed image
0-80
DR fails erasure test when:
faint remnants of previous anatomy or objects
non-uniform areas related to prior exposure
repeatable artifacts indicating incomplete erasure
CR facilities shall erase all CR cassettes, at a minimum, on a _____ basis
weekly
What is the only way to fix DRIFT
calibration (tube age, temp, natural stresses)
what does equipment calibration do?
checks the equipment’s accuracy and ensures consistent image quality
DR calibration:
pixel sensitivity variations, detector non-uniformity
machine typically automatically performs calibration
BIOMED will calibrate after suervice
ensures accurate signal response across the detector
CR Plate Reader Calibration:
plate wear, change in system sensitivity
ensures accurate plate reading and signal conversion
Uniformity test
all digital detectors are INHERENTLY NONUNIFORM
Uniformity corrections must be repeated on a regular basis
should be tested ANNUALLY according to the AAPM
Spatial resolution test should be done:
semi-annually
spatial resolution tests what?
the resolution test tool indicated line pairs per millimeter
due to changes in image quality
test phantom is used
Repeat rates for radiographic procedures should not exceed:
4-6%
Reject analysis
critical that repeat exposures are identified
repeat reason
# of repeats
tech responsible
allows solutions to be found to minimize repeats
Reject analysis is performed:
quarterly
min of 240-250 to be able to get an accurate quarterly analysis
The department monitors what with reject analysis:
Causal: % of repeats from a specific cause
Total: all repeats no matter what the cause
Patient exposure monitoring part of QA program:
recording of modality specific dose indices, associated equipment settings and patient habitus, obtained
identification and analysis of dose-index values and conditions that consistently deviate from corresponding norms
vendor-supplies software…
radiation does index monitoring (RDIM) systems
Dose Watch- automatically collect and analyze patient radiation and iodine exposure
Filtration and HVL
minimum HVL levels must be achieved for pt safety reasons
minimum of 2.5 mm of Al
if it does not reach the minimum then pt dose will increase
uses graded thicknesses of filters (usually aluminum)
protective filtration:
used to remove low-energy x-rays from the beam that cannot penetrate the body and expose the IR efficiently
the only true measure of actual x-ray beam penetration is____
HVL
How often does HVL need to be tested?
annually
If HVL falls below the minimum requirements it indicates:
calibration of kVp for the unit is off
insufficient filtration in the beam
what is kVp accuracy
it is what technique we are setting VS what is actually coming out of the tube
testing tools
computerized radiation dosimeter or kVp meter
can be within + or - 5 kVp of the control/set kVp
when is kVp accuracy tested?
annually
Exposure linearity is
the alignment of a particular mA station relative to other stations in its output of radiation
testing tool:
computerized radiation dosimeter
readout in mR/mAs
when mA is doubled it can be withing + or - 10%
exposure linearity is performed
annually
exposure reproducibility is
ability to repeat the same technique setting and obtain the same results in exposure
testing tool:
computerized dosimeter
should be withing + or - 5%
exposure reproducibility is performed
annually
timer accuracy
computerized radiation dosimeter
acceptance range + or - 5%
unless exposure times less than 10ms
time accuracy tests are performed
annually
OR IF
major repair performed
part replacement to generator or console
beam alignment is
a tests that checks the perpendicularity of the beam
testing tool:
x-ray beam alignment tool
within 1 degree or 5 mm of vertical alignemnt
beam alignment test occurs
semiannually
with age and use the x-ray tube can shift in its housing causing a lack of perpendicularity
collimator accuracy is checked by
beam alignment test tool
9 penny test
collimator accuracy test is
measures field size accuracy, field alignment, and vertical beam alignment
must be within 2% of the SID
40 SID - + or - .8
collimator accuracy is tested
semi-annually
slippage of the collimator shutters
crooked collimator mirrors
SID indicator test is performed
annually
after tube, collimator or suspensions system service
SID indicator is used to
verify the SID accurately reflects the true distance between the x-ray tube focal spot and the IR
ensures accurate exposure technique, image quality, and patient dose consistency
+ or - 2% of the actual SID or + or - 1 inch ( + or - 2 cm), depending on departmental or regulatory guidelines
Automatic exposure control (AEC)
testing is done annually
uses a dosimeter for testing
accuracy range of + or - 10%
Back up timer test
testing is done annually?
place 3 sheets of leaded rubber over the activated cell(s)
should automatically terminate exposure at the set back up time or at 600 mAs
QC of protective apparel
testing is done annually
look for : cracks, tears, and holes
must replace any defected apparel (apron, gloves, gonadal shields)
fluoro exposure rate
testing is done semi-annually
testing tool is a calibrated radiation dosimeter
standard: 88mGy/min
high level control: 176mGy/min
Source to Skin distance (fluoro)
doses established by the FDA
fixed or stationary: 38 cm or 15 in
mobile: 30 cm or 12 in
ABC or AERC (fluoro)
assures that the automatic brightness control is functioning properly
testing tool used is an ACR accredited test object
acceptable range: 10-40 mR/s
exposure reproducibility (fluoro)
testing is done semi-annually
test tool used:
homogenous phantom
dosimeter
should be within + or - 5%
exposure linearity (fluoro)
testing tools used:
homogenous phantom
dosimeter
stop watch
should be withing + or - 10%
focal spot size (fluoro)
done to evaluate the degree of spot BLOOMING
testing tools used:
pinhole cameras
focal spot test tool
resolution chart
homogenous phantom
calculate according to test tool instructions
filtration and HVL (fluoro)
test tools used:
aluminum plates
dosimeter
value found on the x-axis should be greater than the min required HVL (2.5 mm Al)
collimator accuracy (fluoro)
test tool used:
fluoroscopic bean alignment test tool
withing 2% of the SID
beam alignment (fluoro)
field size x spot film
test tool used
fluoroscopic beam alignment device
should be less than 2cm or (2%)
visual/audible monitors (fluoro)
ensures exposure time is accurate
audible alarm should go off after 5 min, exposure is terminated