1/74
Looks like no tags are added yet.
Name | Mastery | Learn | Test | Matching | Spaced | Call with Kai |
|---|
No analytics yet
Send a link to your students to track their progress
AEC is programmed to terminate the _____
exposure time
AEC is also commonly referred to as
phototiming or Automatic Exposure Detection (AED)
AEC can help improve _______ of our image quality.
consistency
what device may be used to ensure consistency of radiographic quality from one exposure to the next?
Automatic Exposure Control (AEC)
what is a radiation detection device that measures the ionization of air in a gas-filled chamber?
ionization chamber
which cell of the ionization chambers has the most influence on the overall exposure when more then one is activated?
the cell receiving the most radiation
When a predetermined level of ionization is reached in the ionization chamber, what does the machine do?
the exposure is terminated
how is the minimum reaction time for AEC determined?
length of time necessary for the AEC to respond to the radiation and for the exposure to be terminated; modern chambers have less than 0.001 second (1/1000 second; 1 millisecond)
Backup timers should be set at _____% of the expected exposure time.
150
Federal law required automatic termination at ______ mAs when working above ______ kvp
600; 50
increasing kVp can _______ the exposure time, but will also _______ contrast
reduce; decrease
a high mA will lead to a ______ in the exposure time
decrease
what program suggests a kVp, mA and active chambers for a given exam?
Anatomically Programmed Radiography (APR)
AEC devices are equipped with _____ _____ that allow the radiographer to adjust the amount of present radiation detected in the cells before termination
density controls (changes the detector sensitivity)
how are the density control buttons indicated on the control panel?
-2, -1, 0, 1, 2
zero indicates neutrality
density control buttons changes exposures by what measure?
some predetermined amount or increment expressed as a percentage- common increment is 25%
*routinely using density controls to produce an acceptable radiograph indicates a possible problem with the AEC device
density controls should not be used to compensate for changes in part thickness
true; AEC does this automatically
1 multiple choice option
too tight collimation around ionization chambers (chamber "clipping") can lead to the chambers operating as it the tissue is _____
dense , leading to overexposure
too wide collimation can lead to an increase in scatter which can lead to an _______ image
underexposed
written tables to help radiographers obtain a consistent, standardized image while using the lowest radiation dose possible
technique chart
a technique chart
all of the above
3 multiple choice options
a thorough technique chart should include the following:
kVp
mA
backup exposure time or mAs
SID
focal spot size
use of grid and grid ratio
AEC detector(s)
acceptable exposure indicator range
a way to measure the path of the central ray; measures the thickness of the patient's part being examined
calipers
how are technique charts typically developed?
based on patient's thickness
designed for average/typical patient
do not consider pathology and condition of patient
adjustments should be made for history and condition
what are the two main types of technique charts?
variable kVp
fixed kVp
when using a variable kVp chart, the _____ remains the same for that particular body part no matter the patient
mAs; kVp is varied based on different patient part thickness as measured with calipers
what rule does the variable kVp- fixed mAs chart used to account for patient thickness?
(2 kVp x part cm) + 30 kVp = new kVp
If the kVp is dropped very low when using a variable kVp chart it is possible to:
under penetrate the body part
1 multiple choice option
when using the fixed kVp chart, _____ is varied according to part thickness as measured with calipers
mAs; assumes optimal kVp for the part being x-rayed (exception is considerably large patients)
optimum kVp for small extremities
50-60
optimum kVp for large extremities
70
optimum kVp for skull
80
optimum kVp for abdomen and ribs
80
optimum kVp for barium-based contrast studies
120
optimum kVp for iodine-based contrast studies
68
optimum kVp for chest with a grid
120
optimum kVp for chest without a grid
80
optimum kVp for spine
80
optimum kVp for pelvis and hip
80
Fixed kVp chart: it makes sense to adjust only mAs since its the controlling factor of _____; a minimum change of _______% is required since anything less is not significant
IRE; 30
which chart is better suited for use with digital imaging receptors?
fixed kVp
1 multiple choice option
using a fixed kVp chart generally uses a higher kVp, which gives:
all of the above
3 multiple choice options
photons in a higher kV beam will have higher energy, they will be more likely to _______ patient's body
pass straight through
1 multiple choice option
at the ______ kVp settings, the change in exposure is more noticeable if the kVp is change (15% rule)
lower
1 multiple choice option
which chart can better control motion?
fixed kVp chart; using a high kVp and lower mAs (reciprocity law)
Variable technique (change to both kVp and mAs) charts provide a way to alter routine settings due to:
pathological conditions
patient age
ability to cooperate
BMI
casts
contrast media
Casts will require an increase in ____
kVp
kVp for wet plaster cast
8-10 kvp increase
kVp for dry plaster cast
5-7 kVp increase
kVp for fiberglass cast
3-4 kVp increase
Additive pathology will lead to a ______ in IR exposure if the technique is not adjusted
decrease
Destructive pathology will lead to an _____ in IR exposure if the technique is not adjusted
increase
Pediatric patients are more radiosensitive, and they are smaller, so the technical factors can be ______
reduced
Geriatric patients typically have bones that are not as dense, thus require a _____ in technical factors
decrease
if the patient was given contrast media, such as barium, the technical factors will need to be _____ due to the higher atomic number of the contrast
increased
if you decide not to use a grid for an exam that typically uses one, than your technique will need to be ___
decreased
the best practice for digital radiography is to use the ______ within the optimal range for the part and position along with the _______ that will still provide adequate exposure to the image receptor
highest kVp; lowest mAs
which type of technique will produce the least radiation dose to the patient?
high kVp technique
2 multiple choice options
what device is attached to the opening in the tube housing and is used to regulate the size and shape of the beam?
beam restrictor
beam restriction reduces the number of _____ ______ being emitted from the tube
primary photons; the helps reduce the amount of scatter, you may sometimes have to increase your technical factors
scattered photons are not part of the useful beam
true
1 multiple choice option
where is scatter produced?
in the patient's body when the primary photons hit the patient's atoms and are deflected with a loss of energy
what is the most used beam restrictor in radiography?
collimator
what are the types of beam restrictors?
1. aperture diaphragms
2. cones and cylinders
3. collimators- aka variable aperture diaphragm
what is the most simplest type of beam restrictor?
aperture diaphragms; made of a sheet with a hole in the center and is placed directly below the tube window (size and shape of hole determine size and shape of beam) ^ no longer use today
which team restrictors are circular aperture diaphragms with metal extensions?
cones and cylinders
which beam restrictor has two sets of lead shutters at right angles to one another that move in opposing pairs?
the collimator
what are the parts of the collimator?
1. light source
2. mirror
3. two sets of shutters
what is the purpose of the light bulb in the collimator?
shows the exact configuration of the x-ray field
target of the tube and the light bulb should be the same distance from the center of the mirror
bottom shutters reduces _______ along the edges of the beam
penumbra (unsharpness)
upper shutters help reduce the amount of _______ ______ reaching the IR by absorbing radiation
off-focus radiation
Features of the shutters
all of these
3 multiple choice options
what is a positive beam limiting device (PBL) ?
automatically collimates the beam to the size of the image receptor
by limiting collimation to the area of interest we are:
improving image quality
decreasing patient dose
what are ancillary devices?
lead blockers and lead masks; help give the beam a specific shape *such as when we place lead blockers behind a patient for a lateral lumbar