99Mo/99mTc generator
- most commonly used generator type in nuclear medicine which allows for extraction of Tc-99m
99Mo/99mTc generator
- saline passes through alumina column
- Mo-99 stays tightly bound to the column & the pertechnetate ion is removed by the Cl- ion
2 weeks
99Mo/99mTc generator shelf-life
23 hours
amount of time needed for maximum extraction of Tc-99m from 99Mo/99mTc generator
1 month
82Sr/82Rb generator shelf-life
82Sr/82Rb generator
- saline passes through alumina column
- Sr-82 stays tightly bound to the column & Rb-82 is removed
transient equilibrium
equilibrium:
- parent half-life is longer than daughter half-life
daughter activity will rise & surpass the parent activity, then decrease to a rate parallel to that of the parent
secular equilibrium
equilibrium:
- parent half-life is very long to the point where the decay rate is negligible over the time period required for the daughter to decay
- parent activity remains constant while the daughter activity gradually accumulates
- met when daughter activity = parent activity
eluate
extracted from the generator
dose calibrator
used to determine the amount of activity eluted & the concentration
0.15uCi Mo-99 per mCi of Tc-99m
radionuclidic 99Mo breakthrough limit
Mo-99 breakthrough
looks for presence of Mo-99 contamination with an eluate
Al+3 breakthrough
confirms amount of aluminum ion present within a radiopharmaceutical
10 ug Al/mL of eluate
Al+3 breakthrough limit
dose calibrator
- used to measure dose activity before/after patient administration
- operates in ionization region
- operates in current mode
constancy
dose calibrator QC performed daily, at install, after repair
constancy
ability of meter to measure same activity reading each day of use
constancy
performed with a single long-lived standard (Cs-137)
linearity
dose calibrator QC performed quarterly, at install, after repair
linearity
correctness of activity reading across a wide range of dose activities
decay method
- measure source at multiple intervals until it decays down to 30 uCi
- graph measurements on semilogarithmic paper
sleeve method (calicheck)
- commercial lead sleeves that simulate decay through attenuation
- absorbs the Tc-99m photons & the Mo-99 photons pass through
accuracy
dose calibrator QC performed annually, at install, after repair
accuracy
- correctness of reading compared to true decay corrected value for a long-lived source
- measured with at least 2 sealed NIST reference standards
geometry
dose calibrator QC performed after change in configuration, at install, after repair
geometry
correctness of reading based on container size/shape/material/volume