1/41
Looks like no tags are added yet.
Name | Mastery | Learn | Test | Matching | Spaced | Call with Kai |
|---|
No analytics yet
Send a link to your students to track their progress
Exposure errors
... can adversely affect the quality of the digital image
>+3
DI:
- excessive radiation exposure
- repeat only if relevant anatomy is clipped or "burned out"
- Require immediate management follow-up
+1 to +3.0
DI:
- overexposure
- repeat only if relevant anatomy is clipped or "burned out"
-0.5 to +0.5
DI:
Target range
<-1.0
DI:
- Underexposure
- Consult radiologist for repeat
<-3.0
DI:
Repeat
F
T or F: mAs affects image brightness
Kilovoltage-peak (kVp)
It affects the variation in radiation intensities exiting the patient and image contrast
lower
Use of kVp for computed radiography should be lower compared to film-screen radiography
50-60
Infant extremities kVp
65-75
Adult extremities kVp
75-90
Bucky extremities kVp
85-95
AP spine kVp
85-100
Cervical kVp
85-100
Thoracic kVp
85-100
Lumbar kVp
110-130
Chest kVp
80-90
Skull kVp
Higher kVp and lower mAs
... kVp and ... mAs are now recommended with digital detectors because of reduction in entrance skin exposure
Fixed kVp systems
The optimal kVp is the maximum kVp level that will produce images with appropriate contrast that are consistently within acceptance limits
Fixed kVp systems
The rule is to double or halve mAs for every 5 cm of thickness
Variable kVp systems
The rule is 2 kVp is adjusted per cm of subject thickness, requiring a base of 30 kVp (40-50 kVp can be used) to be used
kVpₙₑᵥᵥ = (2kVp x part) + 30kVp
Formula for new kVp using variable kVp systems
Fixed kVp systems
... systems are better suited for use with digital image receptors
Collimation
Changes in beam restriction alter the amount of tissue irradiated, affecting the amount of exposure to the image receptor
Source-to-Image Receptor Distance (SID)
It is the distance between the source of radiation and the image receptor, and it affects the amount of radiation reaching the patient
object-to-image receptor distance (OID)
It is the distance between the object radiographed and the image receptor
Decreased
The presence of OID may result in ... beam intensity
Zero
Optimal recorded detail is achieved when the OID is ...
Grid
It is a device placed between the Anatomic area and the image receptor to absorb scatter radiation exiting the patient
Thicker
In computed radiography, ... phosphors have faster speed and absorbs more radiation
Thinner
... phosphors have slower speed and absorbs less radiation, producing sharper image quality
Scatter radiation
... can add unwanted exposure to the radiographic image as a result of Compton interactions
Fog
It is produced as a result of scatter radiation reaching the image receptor, and can be visualized on both a digital and a film image
Unwanted exposure or fog
... or ... on the image does not provide information about the Anatomic area of interest
Increased mAs
Maintaining the exposure to the image receptor when imaging a thicker part requires an ...
Increases, decreases
As thickness increases, scatter radiation ..., and radiographic contrast ...
Filtration
It is the removal of low energy x-rays from the x-ray beam spectrum which does not contribute to image quality
Automatic Exposure Control (AEC)
It provides x-rays with the consistent receptor exposure, reducing changes of overexposure
Automatic Exposure Control (AEC)
It is used to control the amount of radiation reaching the receptors by terminating the length of exposure
Receptor exposure
Ionization chambers measure ...
Lead shielding
... should never be placed over any portion of an active ionization chamber