1/129
Looks like no tags are added yet.
Name | Mastery | Learn | Test | Matching | Spaced | Call with Kai |
|---|
No analytics yet
Send a link to your students to track their progress
protective tube housing
lead lined metal that protects personnel and patients from leakage and off focus radiation
cannot exceed 1 mGy per hour at 1m away from housing (Air Kerma of 0.88mGy/hr)
control panel/console
technologist workspace where you can set up the console, technique, and take the exposure
properly shielded barrier
exposure button must be affixed to the console so the exposure cant be taken in an unshielded area
displays when the x-ray tube is energized
kVp (kilovoltage peak)
maximum possible energy of a photon that exits the x-ray tube, unit selected on the operating console
indirectly proportional to patient exposure
mA (milliamperage)
measurement of x-ray tube current or the number of electrons crossing the tube from cathode, unit selected on the operating console
directly proportional to patient exposure
mAs (milliampere seconds)
controls the amount of radiation produced by the x-ray tube
mA X seconds = mAs
directly proportional to patient exposure
AEC (automatic exposure control)
exposure will terminate when a predetermined amount of radiation is reached for the selected body part
can only be used on body parts big enough to cover the cells
back up timer for patient safety
x-ray room design
table must adequately support patient and be uniform density
SID must be accurately measured (digital or tape measure)
beam limiting devices
limits the primary beam to a smaller area
decreases exposure by reducing the amount of tissue that is exposed to radiation, reduces scatter
types of beam limiting devices
aperture diaphragm
cones
collimators
aperture diaphragm
flat lead with a shape and size cut into it that is placed below the window, reduces scatter
rectangular (most common)
square
round
cones
circular metal cylinders connected to the tube housing that limits the size of the beam, can be flared or straight
extensive cylinders - can be telescoped 10-12 inches for smaller exposure area
mostly used in dental radiography but can be used for the heel, skull, and spine imaging
collimators
light-localized variable aperture rectangular collimator
must be bright enough to outline the anatomy to include
most versatile beam restriction - can change sizes
reduces exposure by 20-30%
should not be larger than the size of the image receptor
what are collimators
2 sets of shutters 90 degrees from one another
near (upper) shutters
located close to the window, reduces exposure from off focus radiation
far (lower) shutters
located closer to the light source, confines the beam to the area of interest
skin sparing
minimizes skin exposure by requiring a 15cm distance from the skin to the collimator
can be achieved with spacer bars mounted on the tube
PBL (positive beam limitation) (automatic collimation)
electronic sensors in the bucky that senses the image receptor size that you are suing and opens the light field to that size - can be slits or pegs
reduces user errors by matching the light field to the image receptor size
required to be within 2% accuracy
collimation → scatter/dose
increased collimation → smaller scatter/dose
decreased collimation → larger scatter/dose
filtration
hardens the beam by cleaning up the low energy x-rays
reduces the skin and superficial exposure to the patient - decreases patients absorbed dose because the remaining photons and higher energy
2.5mm Al for units operating above 70kVp
inherent filtration
0.5m Al equivalent
glass envelope, insulating oil, and glass window
added filtration
2.0mm Al equivalent
sheets of Al added outside the glass window above the collimator shutters
accessible by service person, can be changed as the tube ages
mobile & fluoro filtration
2.5mm
NCRP 102
list minimum required filtration for x-ray equipment
diagnostic x-ray beam must always have adequate filtration
intensity decreased
1 HVL - 50%
2 HVL - 25%
3 HVL - 12.5%
4 HVL - 6.25%
HVL
insufficient HVL test could mean improper filtration
must get tested
measures beam quality or effective energy of the beam
measured at least once a year by a physicist or if the tube is replaced or repairs are made
TVL (tenth value layer)
thickness that will decrease the intensity of the beam by 1/10
compensating filters
used when x-raying a part that has varying thickness to reduce dose and provide a uniform density across the image
decreases the entrance skin exposure (ESE)
constructed of aluminum or lead-acrylic that is attached to the bottom of the collimator
types of compensating filters
wedge filter
trough filter or bilateral wedge
ferlic
boomerage
wedge filter
used for foot and spines
trough filter or bilateral wedge
used on chest x-rays, thicker on both sides and thin in the middle
ferlic filter
used on hips
boomerang filter
used on shoulders
radiation protection - testing on equipment
exposure reproducibility - system must duplicate exposure when taken at the same exposure - variance 5%
exposure linearity - system must consistently produce radiation output when mA and time are changed - variance 10%
radiographic grids
material is parallel radiopaque material of very thin lead, aluminum, or plastic fiber
removes scatter and improves visibility of detail
increases patient dose
when should you use a grid
thickness is over 12cm at 60 kVp or higher
grid ratio
height of lead strip divided by distance between strips
mobile radiography
minimal source to skin distance on a mobile fluoroscopy unit is 12 inches (30cm)
only perform portable x-ray on patients that cannot be transported to the department
digital imaging
utilization of technique charts
grids
has lower doses, immediate images, image manipulation, less maintenance
fluoroscopy
continuous radiation that shows dynamic motion, largest exposure to patients in diagnostic radiology
image intensification
increases brightness on screen
source to skin distance
15 inches (38 cm) for fixed units
12 inches (30 cm) for mobile units
cumulative timing device
audible alarm or interrupt of fluoro every 5 minutes of time
technologist is responsible to record the fluoro time and dose in medical record
federal regulations tabletop exposure - fluoroscopy
88 mGy per minute
ABC/ABC (automatic brightness control/stabilization)
no matter the kVp or mA varying, the brightness of the image remains the same
AERC (automatic exposure rate control)
adjusts exposure factors automatically as the beam moves over varying thicknesses
fluoro exposure switch/dead man switch
foot pedal requires direct pressure to continue fluoro exposure
c-arm fluoroscopy
used in the OR, cardiac cath, and IR
12 inch (30 cm) minimal distance to the patient
should be positioned with the image intensifier on the top - reduces scatter and patient dose
digital fluoroscopy
beam turns off while image is scanned and then turns back on - pulsed
dose area product (DAP)
last image hold
DAP (dose area product)
newer fluoro systems provide the sum of the air kerma (energy) over the exposed area of the patient
last image hold
when the foot comes off the pedal, it holds the last image and displays it on the screen until the foot pedal is activated again
digital subtraction angiography (DSA)
used in interventional and vascular
used to visualize occlusions, stenosis, or aneurysms
uses software to remove the superimposing anatomy so you only see the area of interest
interventional radiology
invasive sterile procedures performed by a physician under fluoro
FDA requires documentation in the chart if skin dose is 1-2 Gy
federal regulations for table-top exposure for high level control fluoro (IR)
176 mGy per minute
equipment safety
on and off switches
interlocks - detents & fluoro locks
visual/audio monitors - control panel, laser light, tae measures
emergency controls
holistic patient care
treat the patient as a whole person rather than a body part
makes patients feel more respected, comfortable, and more willing to cooperate
patient motion
involuntary & voluntary
involuntary motion
caused by muscles, not controllable
heart & digestion
manifestations that can cause involuntary motion
chills
tremors
spasms
pain
active withdrawal
voluntary motion
controlled motion, can be caused by
patients age
breathing patterns
anxiety
physical or mental discomfort
fear of exam/prognosis
mental instability
how is involuntary motion corrected
decreasing exposure time and increasing imaging receptor speed
how is voluntary motion corrected
gaining patient cooperation and use of proper immobilization
immobilization
piggostat
papoose/octostop
sponges and sandbags
mummy wrap/bunny wrap
tape
velcro straps
radiolucent plexiglass
non-radiology employee helping to hold
sponges
radiolucent
sandbags
radiopaque
what did the AAPM state in April 2019
shielding of patient gonadal or fetal shielding during diagnostic imaging should be discontinued
what committee was created after the statement made by AAPM
the CARES committee (communicating advances in radiation education for shielding)
who supports the statement made by the AAPM
NCRP
ACR
ASRT
ARRT
radiosensitive organs
lens of eye
breasts
reproductive organs
2 types of shielding
gonadal & specific area
gonadal shielding
could use unless covering the area of interest
shield cannot be in the collimation if AEC is being used
what is the first step for gonadal protection
proper collimation
females receive ___ more exposure than males
3X
flat shields
most effective in the AP or PA recumbent positions
shadow shields
careful to place properly or a repeat image could be caused
not suitable during fluoro
shaped shields
contoured to enclose the male reproductive organs
can be placed by the patient
cannot be used during PA projections
clear shields
transparent lead-plastic material
lap shields (half)
covers only the front or back half of the patient, attached with a velcro strap or on wheels
specific area shields
eyes
breast
thyroid
gloves
what type of kVp and mAs is best for reducing patient exposure
high kVp, low mAs
high kVp and low mAs
ensures adequate anatomy is penetrated & minimizes dose
technique books
provide an average technique for body part/equipment
AEC needs to be properly functioning
the higher the kVp, the ____ the entrance skin dose
lower
penetrated radiation
radiation that has passed through the body and hit the image receptor
absorbed radiation
radiation that has hit and stopped in the body part
scatter radiation
radiation that has hit the body part, but has scattered outside the body part into the air
exposure index
the number that is found on the image after processing that measures receptor exposure
should be in range for the equipment parameters to be a good diagnostic image
100-300 in the lab
what does under exposure cause
quantum noise (grainy appearance) & should be repeated
what does over exposure cause
will appear as a good image in most cases
in extreme cases, will cause saturation & should be repeated
correct positioning
inadequate processing of image results in repeats
artifacts caused by image receptor, software, or patient related
quality control program
monitors that suboptimal images are not being produced
what are quality control programs looking/watching out for
repeats
unnecessary patient dose
incorrect projections being taken
machines are working properly - would affect all images taken if not
what is an alternative to using a grid to clean up scatter
air gap technique
air gap technique
patient is placed 4-6 inches (10-15 cm) away from the image receptor with a 10-12 feet (300-366cm) SID
what is a con regarding the air gap technique
there is an increase in magnification and is not useful if the kVp is higher than 90
repeat image
any image that must be done more than once due to human or mechanical errors
patient receives a “double dose”
repeat images are unacceptable if done due to…
carelessness or poor judgment
positioning
technique
reasons for unacceptable images
patient mispositioning
incorrect centering of the beam
patient motion during exposure
incorrect collimation of the beam
presence of external foreign bodies
post-processing artifacts
chest x-rays
pre-admission
pre employement
routine health check ups
screening for TB
lumbar x-rays
pre employment
CT whole body scans
check for disease