Technique Charts

Standardized guidelines for selecting kVp and mAs for specific exams to ensure image quality and reduce repeats.

2
New cards

Fixed kVp

Technique where kVp remains constant and mAs varies according to patient thickness.

3
New cards

Variable kVp

Technique where kVp changes with patient thickness, following the rule of 2 kVp change per 1 cm thickness.

4
New cards

Anatomically Programmed Radiography (APR)

Computerized system that automatically sets exposure factors based on selected body part and projection.

5
New cards

Automatic Exposure Control (AEC)

System that automatically terminates exposure when the detector receives a preset amount of radiation.

6
New cards

kVp

Kilovolt peak; determines the energy and penetration of the x-ray beam, affecting contrast.

7
New cards

mAs

Milliamperes per second; controls the quantity of x-rays produced.

8
New cards

Pixel

Smallest unit of a digital image that contains a numeric value representing signal intensity.

9
New cards

Spatial Resolution

Ability of an imaging system to distinguish small objects that are close together.

10
New cards

Fill Factor

Percentage of the detector element that is sensitive to radiation; higher fill factor means better efficiency.

11
New cards

Bit Depth

Number of gray shades that can be displayed in a digital image, influencing contrast resolution.

12
New cards

Detective Quantum Efficiency (DQE)

Measure of how efficiently a detector converts incoming x-ray photons into image signals.

13
New cards

Quantum Noise (Quantum Mottle)

Random grainy appearance in an image caused by too few x-ray photons reaching the detector.

14
New cards

Image Identification

Ensures radiographic images are matched to the correct patient and examination for legal and diagnostic accuracy.

15
New cards

Contrast Resolution

Ability of a digital imaging system to distinguish small differences in tissue density.

16
New cards

Digital Imaging and Communications in Medicine (DICOM)

Standard format for storing and transmitting medical images to ensure compatibility.”},{

17
New cards

Minimum Exposure Controls

Prevent excessively short exposure times, ensuring AEC operates effectively.

18
New cards

Quantum Noise

Grainy appearance in an image caused by inadequate mAs leading to low signal intensity.

19
New cards

Grid

Device used between the patient and image receptor to reduce scatter radiation and improve image contrast.

20
New cards

Beam Restriction

Limiting the X-ray beam to the area of interest using collimation.

21
New cards

Dynamic Range

Range of exposure values a digital detector can accurately capture, allowing visualization of structures with different densities.

22
New cards

Image Artifacts

Unwanted objects or distortions in an image that do not represent patient anatomy.

23
New cards

Inherent Filtration

Filtration that is built into the X-ray tube, including the glass envelope and insulating oil.

24
New cards

Compensating Filters

Devices used to even out differences in patient thickness and improve image quality.

25
New cards

Collimator

Device that shapes the X-ray beam to reduce exposure to surrounding tissues.

26
New cards

Minification Gain

Increase in brightness of the output image compared to the input due to size difference in phosphors.

27
New cards

Flux Gain

Increased brightness of an image due to acceleration of electrons in the image intensifier.

28
New cards

Automatic Brightness Control (ABC)

System that automatically adjusts exposure parameters to maintain consistent image brightness.