Image Production Capstone Unit 3: March 30th 



Technique Charts

Standardized guidelines are used to select kVp and mAs for specific exams to ensure consistent image quality and reduce repeats.

Includes:

  • kVp

  • mAs

  • SID

  • Grid use

  • Patient thickness

Types:

  • Fixed kVp

  • Variable kVp

  • Anatomically Programmed Radiography (APR)

Anatomically Programmed Radiography (APR)

A computerized system where a technologist selects body part + projection and the machine automatically sets exposure factors.

Advantages:

  • Faster setup

  • Fewer technique errors

  • Consistent exposures

Technologists must still adjust for:

  • patient size

  • pathology

  • casts

  • Age

Fixed vs Variable kVp

Fixed kVp

  • kVp stays constant
    mAs changes with patient thickness

  • Produces consistent contrast

Variable kVp

  • kVp changes with thickness

  • Rule: 2 kVp change per 1 cm thickness

  • mAs stays constant

ARRT Tip: Fixed kVp = most consistent contrast

Special Considerations

Casts

Increase exposure because casts absorb X-rays.

Fiberglass cast: +3–4 kVp

Plaster cast:

  • Dry: +5–7 kVp

  • Wet: +8–10 kVp

Pathologic Factors

Additive Disease

Increases tissue density → increase technique

Examples:

  • pneumonia

  • ascites

  • edema

  • tumors

Destructive Disease

Decreases tissue density → decrease technique

Examples:

  • osteoporosis
    emphysema

Memory tip: Additive = Add exposure

Age Considerations

Pediatrics

  • Smaller body size

  • Higher radiation sensitivity

Technique:

  • Reduce exposure

  • Collimation 

  • Use shielding

  • Short exposure times

Geriatrics

Often decreased bone density → may require slightly reduced technique, but pathology may increase exposure.

Body Mass Index (BMI)

Higher BMI = more tissue attenuation.

Adjustments:

  • Increase kVp or mAs

Lower BMI = decreased exposure.

Contrast Media

Positive Contrast

Absorbs x-rays → appears white

Examples:

  • barium

  • iodine

Negative Contrast

Allows x-rays to pass → appears black

Examples:

  • air

  • CO₂

Grids

Purpose: Reduce scatter → improve image contrast

Used when body part thickness > 10 cm

Effects:

  • Improves contrast

  • Requires increased mAs

Higher grid ratio:

  • better scatter cleanup

  • Higher patient dose

OID (Object-to-Image Distance)

OID = distance between anatomy and image receptor.

Increased OID allows scatter, which travels widely in divergent paths to exit the patient and miss the IR. 

Air gap technique seen on lateral c spine doesn’t require a grid similar to 10:1 

Increasing OID causes:

  • Magnification ↑

  • Recorded detail ↓

Best image quality:

  • Small OID

  • Large SID
    Small focal spot

Automatic Exposure Control (AEC)

Definition:
AEC automatically terminates the exposure when the detector receives a preset amount of radiation.

Purpose:

  • Produces consistent receptor exposure

  • Reduces technique errors

  • Requires excellent positioning skills
    Helps standardize images between technologists

Key components:

  • Ionization chambers (detectors)

  • Backup timer

  • Density control

  • Shortest time with an AEC is 1 ms 

Important:
AEC controls exposure time, on a fixed kVp.

The technologist still selects:

  • kVp

  • mA

  • detector cell

  • density setting

Detector Selection

Most systems have 3 AEC chambers:

  • Left

  • Center

  • Right

The technologist chooses the chamber under the anatomy of interest.

Examples:

Chest PA
→ Both outer chambers

AP abdomen
→ Center chamber

Spine
→ Center chamber

ARRT tip:
Incorrect chamber selection → overexposure or underexposure.

Anatomic Alignment

For AEC to work correctly:

The anatomy of interest must completely cover the selected chamber.

If anatomy does NOT cover the chamber:

  • Exposure may terminate too early

  • The image may be underexposed

Example:

Chest rotated → lung may not cover the chamber → exposure error.

Backup Timer

Safety feature that stops exposure if AEC fails.

Purpose:

  • Prevents excessive patient dose

Typically set at:

  • 150% of expected exposure time

Density Control (Exposure Adjustment)

Density control adjusts the amount of radiation reaching the detector.

Does NOT change contrast.

Often labeled:

  • +1

  • +2

  • −1

  • −2

Each step changes exposure by about 25–30%.

Example:

+1 → slightly darker image
−1 → slightly lighter image

Used for small exposure corrections.

Effects of Exposure Factors on Image Quality

Exposure Factor

Primary Effect

kVp

Penetration and contrast

mAs

Image receptor exposure (density)

SID

Affects exposure intensity

OID

Magnification and recorded detail

Grid

Improves contrast but increases dose

kVp Effects

Increasing kVp:

  • Increases penetration

  • Decreases image contrast

  • Increases scatter

Decreasing kVp:

  • Less penetration

  • Higher contrast

  • Less scatter

ARRT rule:

15% rule

Increasing kVp by 15% ≈ doubles receptor exposure.

mAs Effects

mAs controls the quantity of x-rays.

Increasing mAs:

  • Increases receptor exposure

  • Increases patient dose

  • No effect on contrast

Decreasing mAs:

  • Decreases exposure

  • May produce quantum noise

Common AEC Errors (ARRT Favorites)

  1. Wrong chamber selected

  2. Improper positioning

  3. Anatomy does not cover the chamber

  4. Metal prosthesis over the chamber

  5. Incorrect centering

These cause overexposure or underexposure.

Spatial Resolution

Definition:
The ability of an imaging system to distinguish small objects that are close together.

Also called:

  • Recorded detail

Measured in:

  • Line pairs per millimeter (lp/mm)

Higher spatial resolution 

  • sharper images

  • better visibility of fine detail

Factors affecting spatial resolution:

  • Pixel size

  • Matrix size

  • Detector element size

  • Sampling frequency

  • MTF

Pixel Characteristics

Pixel

A pixel (picture element) is the smallest unit of a digital image.

Each pixel contains:

  • a numeric value representing signal intensity.

Pixel Size

Determined by:

Field of view ÷ matrix size

Smaller pixels:

  • increase spatial resolution

  • reduce signal per pixel

Larger pixels:

  • decrease spatial resolution

  • increase signal

Pixel Pitch

Distance from the center of one pixel to the center of the next pixel.

Smaller pitch:

  • higher resolution

Larger pitch:

  • lower resolution

Detector Elements (DEL)

A detector element (DEL) is the physical component of the detector that captures x-ray energy.

Each DEL corresponds to one pixel in the image matrix.

DEL Size

Smaller DEL size:

  • higher spatial resolution

Larger DEL size:

  • lower resolution

  • greater signal capture

Fill Factor

The percentage of the detector element that is sensitive to radiation.

Higher fill factor:

  • better detector efficiency

  • improved image quality

Lower fill factor:

  • less efficient detection

CCD and CMOS Detectors

Used in some digital imaging systems.

CCD (Charge-Coupled Device)

A digital detector that converts X-ray light into electrical signals and transfers the charge across the chip to produce a digital image.

Characteristics:

  • High image quality

  • Good sensitivity

  • Signal transferred across the chip

Limitation:

  • slower readout

CMOS (Complementary Metal-Oxide Semiconductor)

A digital detector that converts x-ray light into electrical signals and processes the signal directly at each pixel for faster image readout and lower power use.

Characteristics:

  • Faster image processing

  • Lower power consumption

  • Less electronic noise

Often used in modern digital detectors.

Sampling Frequency (CR)

Important in Computed Radiography (CR).

Definition:
Number of times the CR laser scans the imaging plate per millimeter.

Measured in:

  • pixels/mm

Higher sampling frequency:

  • higher spatial resolution

  • smaller pixel size

Lower sampling frequency:

  • reduced resolution

Matrix Size

Matrix = number of pixels in the image.

Example:

  • 1024 × 1024

  • 2048 × 2048

Total pixels = rows × columns.

Larger matrix:

  • smaller pixels

  • higher spatial resolution

Smaller matrix:

  • larger pixels

  • lower resolution

Modulation Transfer Function (MTF)

Definition:
MTF describes the ability of an imaging system to reproduce detail accurately.

It measures how well the system transfers object contrast to the image at different spatial frequencies.

Higher MTF:

  • better spatial resolution

  • more accurate image detail

Lower MTF:

  • loss of fine detail

MTF is considered the best overall measure of spatial resolution.

Contrast Resolution

Definition:
The ability of a digital imaging system to distinguish small differences in tissue density.

In other words:
Ability to see subtle differences in shades of gray.

Digital imaging has excellent contrast resolution compared to film.

Factors affecting contrast resolution:

  • Bit depth

  • Detector efficiency

  • Scatter radiation

  • Grids

Higher contrast resolution →
better visualization of soft tissues.

Example:
Detecting small differences between organs, fluid, and soft tissue.

Bit Depth

Definition:
The number of gray shades that can be displayed in a digital image.

Bit depth determines how much contrast information is stored.

Formula:

Number of gray shades = 2ⁿ
(n = number of bits)

Examples:

Bit Depth

Gray Levels

8 bit

256 shades

10 bit

1,024 shades

12 bit

4,096 shades

16 bit

65,536 shades

Higher bit depth:

  • better contrast resolution

  • more gray shades

ARRT Tip:
Digital systems often have 12–16-bit depth.

Detective Quantum Efficiency (DQE)

Definition:
A measure of how efficiently a detector converts incoming x-ray photons into useful image signals.

Higher DQE means:

  • Better image quality

  • Less radiation needed

  • Lower patient dose

In simple terms:

Higher DQE = better detector efficiency.

Typical comparison:

System

DQE

Film-screen

Lowest

CR

Moderate

DR

Highest

ARRT concept:
Higher DQE allows dose reduction while maintaining image quality.

Grids

Purpose:
Grids remove scatter radiation before it reaches the detector.

This improves image contrast.

Scatter causes:

  • fog

  • reduced contrast

When Grids Are Used

ARRT rule:

Use grids when body part thickness is greater than 10 cm.

Common exams using grids:

  • Abdomen

  • Pelvis

  • Lumbar spine

  • Adult chest

Grid Ratio

Grid ratio = height of lead strips ÷ distance between strips

Examples:

  • 8:1

  • 12:1

  • 16:1

Higher grid ratios:

Advantages:

  • remove more scatter

  • improve contrast

Disadvantages:

  • increase patient dose

  • require higher mAs

  • more positioning accuracy

Grid Errors

Grid Cutoff

Occurs when primary radiation is absorbed by the grid.

Causes include:

  • Off-level grid

  • Off-center beam

  • Wrong SID

  • Upside-down grid

  • Grid misalignment

Result:
Reduced receptor exposure (lighter image)

Linear grids, focused grids, stationary gridsmoving gridsand and software that removes scatter from the image without the use of a grid. 



Type

Key Idea

Linear grid

Lead strips run in one direction

Focused grid

Lead strips angled to match x-ray beam

Stationary grid

Grid does not move

Moving grid

Grid moves to blur grid lines

Software scatter reduction

Digital processing removes scatter without a grid

Image Signal

Definition:
The useful information produced by x-ray photons detected by the imaging receptor.

In digital imaging:

  • X-rays interact with the detector

  • The detector converts them into an electronic signal

  • The signal becomes the digital image

Higher signal means:

  • better image visibility

  • improved image quality

Lower signal may lead to:

  • noisy images

Dynamic Range

Definition:
The range of exposure values a digital detector can accurately capture.

In simple terms:
The detector’s ability to record very low to very high exposures in one image.

Digital systems have a wide dynamic range.

Advantages:

  • fewer repeat exposures

  • ability to visualize structures with different densities

Example:
Both soft tissue and bone can be visible in the same image.

The film screen had a narrow dynamic range, meaning exposure had to be very precise.

Quantum Noise (Quantum Mottle)

Definition:
Random grainy appearance caused by too few X-ray photons reaching the detector.

Occurs when:

  • mAs is too low

  • Photon numbers are insufficient

Appearance:

  • speckled or grainy image

Important ARRT concept:

Quantum noise increases when:

  • exposure decreases

Solution:

  • increase mAs

Signal-to-Noise Ratio (SNR)

Definition:
The ratio of useful image signal to background noise.

Indicates overall image quality.

Higher SNR:

  • clearer image

  • less noise

  • better visibility of structures

Lower SNR:

  • noisy image

  • reduced diagnostic quality

Factors affecting SNR:

Increase SNR:

  • increase mAs

  • improve detector efficiency

  • reduce noise

Decrease SNR:

  • low photon exposure

  • quantum mottle

  • electronic noise

Image Identification

Purpose of Image Identification

Image identification ensures that radiographic images are correctly matched to the patient and examination.

It is required for:

  • legal documentation

  • accurate diagnosis

  • proper medical records

Improper identification can result in medical errors and legal liability.

Methods of Image Identification

Radiographic Identification

Information is placed directly on the image at the time of exposure.

Examples include:

  • Patient name

  • Date of exam

  • Time of exam

  • Facility name
    Technologist ID

  • Anatomical side marker (R or L)

Radiopaque markers are placed on the image receptor before exposure.

Important ARRT rule:
Lead side markers must be placed before exposure whenever possible.

Electronic Identification

Used in digital imaging systems.

Patient and exam information are added through:

  • PACS

  • Radiology information systems (RIS)

  • Electronic patient records

Data is embedded into the digital image file.

Legal Considerations

Radiographs are legal medical documents.

They must contain accurate patient and exam information.

Required patient data typically includes:

  • Patient name

  • Patient identification number

  • Date of examination

  • Facility identification

  • Anatomical side marker (R or L)

Additional exam data may include:

  • Projection or position

  • Technologist identification

  • Time of exam (sometimes required)

Important ARRT Legal Points

Images must be properly labeled before leaving the room.

Markers must indicate:

  • Right (R)

  • Left (L)

Annotations such as:

  • Upright

  • Supine

  • Decubitus

May also be included when necessary.

Failure to properly identify images can lead to:

  • incorrect diagnosis

  • repeat exams

  • legal consequences.

Criteria for Image Evaluation 

Exposure Indicator

The exposure indicator (EI) is a numerical value that tells the technologist whether the correct amount of radiation reached the digital detector.

  • Too high → overexposure

  • Too low → underexposure / quantum noise
    It helps technologists determine if the technique used was appropriate.

Quantum Noise (Quantum Mottle)

Quantum noise is a grainy or speckled appearance in the image caused by too few X-ray photons reaching the detector.
It usually occurs when mAs is too low, which reduces the signal and increases noise in the image.

Gross Exposure Error

A gross exposure error occurs when the image is severely overexposed or underexposed.

  • Overexposure can cause saturation, where image data is lost.

  • Underexposure causes excessive noise and poor image quality.
    These errors may make the image non-diagnostic.

Contrast

Contrast is the difference in brightness between adjacent structures in the image.
It allows radiologists to distinguish between different tissues such as bone, soft tissue, and air.
Contrast is primarily influenced by kVp and scatter radiation.

Controlled by bit depth, the number of bits per pixel, the larger the bit depth, the greater the number of levels of gray 

High contrast: few gray tones or short scale at lower kVp 

Low contrast: many gray tones, long scale, and produced at higher kvp. 

Subject contrast: controlled by the kVp, differential absorption by anatomy and pathology, and variation of atomic number. 

Spatial Resolution

Spatial resolution is the ability of the imaging system to show small objects and fine details clearly.
High spatial resolution means the image appears sharp with clear edges, and closely spaced structures can be distinguished.

  • Controlled by pixel size and pixel pitch 

  • Pixel limited 

  • Influenced by SID, OID, focal spot size, and motion 

Distortion (Size and Shape)

Distortion occurs when the size or shape of the anatomy is inaccurately represented on the image.
It is usually caused by improper alignment of the x-ray tube, patient, or image receptor, or by excessive OID (object-to-image distance).



Distortion Type

What Happens

Main Cause



Size Distortion

Object appears larger

Increased OID



Elongation

Object appears stretched

Excessive CR angle



Foreshortening

Object appears shortened

Part not parallel to IR

Image Artifacts

Artifacts are unwanted objects or distortions that appear on the image but do not represent patient anatomy.
They can be caused by:

  • dirty detectors

  • patient motion

  • jewelry or clothing

  • processing or equipment errors.

Artifacts may interfere with diagnostic interpretation.

Radiation Fog (CR)

Radiation fog occurs when a computed radiography (CR) imaging plate is accidentally exposed to radiation before processing.
This results in a uniform gray appearance and reduced image contrast, which can degrade image quality.

The X-ray Imaging System

The three main components of an X-ray system are:

  • Operating Console

  • X-ray tube

  • High-voltage generator

  • The system operates at a voltage of 25kVp to 150kVp 

  • The tube current at 100mA to 1200mA

  • A x-ray table is designated by the tilt of the  table – most common for fluoroscopy tables is a 90-degree foot tilt and a 30-degree head tilt



Operating Console

  • Radiation quantity – the number of x-rays in a beam (milligray/milliampere-second: mGya/mAs)

  • Radiation quality – the energy of the X-ray beam (kilovolt peak: kVp, or half-layer value; HVL)

  • Automatic Exposure Control (AEC) terminates the exposure when the required amount of radiation intensity is reached.

  • All electric circuits on the operating console  are low voltage

  • The voltage to an X-ray unit may vary as much as 5%. Line compensators are used to maintain a constant 220 volts to the equipment.

X-ray Generator, Transformers & Rectification

High-Voltage Generator

  • Used for increasing the voltage to the kilovoltage required for the X-ray imaging equipment

  • 3 primary parts:

    • High-voltage transformer

    • Filament transformer

    • Rectifiers

Falling Load Generator – the X-ray equipment will choose the highest mA allowing for the shortest exposure time (mA x seconds = mAs)

  • A step-down transformer from the high-voltage supply, the secondary windings are insulated from the primary windings.

  • Disadvantages of long breathing exams  

Autotransformer

Autotransformers

  • Supplies a precise voltage to the x-ray equipment's high-voltage circuit.

  •  The primary and secondary voltages are directly related to the number of wire turns around the core.

  • Autotransformer Law:     Vs = Ns

                                      Vp     Np

kVp

  • kVp determines the quality of the X-ray beam – the kVp is shown on the console at all times: known as the pre-reading kVp meter



mA

  • The tube current, measured in milliamperes (mA), flows from cathode to anode.

  • It represents the number of electrons resulting from thermionic emission.

  • A step-down transformer is necessary for this milliamperage current.

  • Fixed mA stations typically include 100mA, 200mA, and 300mA.

Transformers (X-ray Circuit)

Step-Up Transformer (High Voltage Transformer)

  • Function: Increases voltage (kVp)

  • Converts volts → kilovolts

  • Used to accelerate electrons from the cathode → anode

  • Primary vs Secondary:
    The difference is the amplitude (height) of the waveform — higher on the secondary side = higher voltage
    Turns Ratio:

    • More turns in the secondary coil than the primary coil

    • Example: 1:500

    • Meaning: Voltage increases 500×, current decreases

Step-Down Transformer (Filament Circuit)

  • Function: Increases current (mA)

  • Converts amperage → milliamperage (filament heating)

  • Heats filament → thermionic emission

  • Turns Ratio:

    • Fewer turns in the secondary coil than in the primary

    • Example: 10:1

    • Meaning: Voltage decreases, current increases

Key Concept: Turns Ratio



 

Turns Ratio Formula

  • Ns = Number of turns in the Secondary coil

  • Np = Number of turns in the Primary coil

What It Means

  • If Ns > Np → Step-UP transformer

    • Voltage increases

    • Current decreases

  • If Ns < Np → Step-DOWN transformer

    • Voltage decreases

    • Current increases

  • Turn ratio may be 500- 1000

Formula idea (no math needed for ARRT):

  • Voltage ∝ number of turns

  • Current is inverse of voltage

 Simple memory trick:

  • Step-up = more turns = ↑ voltage, ↓ current

  • Step-down = fewer turns = ↓ voltage, ↑ current

Induction Concepts

Mutual Induction

  • When current flows in the primary coil, it creates a magnetic field

  • This magnetic field induces a voltage in the secondary coil
      This is how transformers work

Self-Induction

  • A changing current in a coil creates its own opposing magnetic field

  • This resists changes in current flow

Super simple:
The coil “pushes back” against changes in its own current

Rectification

Converts AC → DC (needed for x-ray production)

  • Are solid state semiconductor diodes that contain silicon-based n-type and p-type semiconductors. 

The current changes direction every 1/60th of a second for alternating 

  • Half-wave: uses only half the power, not conducted during the negative direction of the wave

  • Full-wave: uses all power (more efficiently), with four diodes, negative direction of the wave is reversed. 

Phase, Pulse, Frequency

Single-phase

Voltage Ripple & Efficiency

Single-Phase

  • 100% voltage ripple

  • Voltage drops to 0 between pulses

  • Only about 50–60% of the peak voltage is effectively used

Result:

  • Least efficient

  • Lower average kVp → less penetrating beam

Three-Phase (6-pulse)

  • Ripple ~ 13%

  • Voltage never drops to zero

  • Uses about ~87% of peak voltage

Key Fact:

  • Produces ~35% more X-rays than single-phase at the same kVp

Three-Phase (12-pulse)

  • Ripple ~ 4%

  • Voltage is almost constant

  • Uses about ~96% of peak voltage

Key Fact:

  • Produces ~41% more X-rays than a single-phase

High-Frequency Generators

  • <1% ripple (almost perfectly constant)

  • Uses ~99% of peak voltage
    Result:

  • Most efficient, less patient dose. 

  • Highest, most consistent beam quality

Why More X-rays Are Produced

  • Less ripple = voltage stays closer to peak kVp longer

  • Electrons hit the anode with higher, more consistent energy

  • → More efficient X-ray production

Key Comparison (Test Gold)

  • Three-phase & high-frequency systems produce an increase of ~12–16% more radiation than single-phase at the same kVp and mAs settings

  • Because their average voltage is higher

  • Single-phase generators produce about a 12–16% decrease in radiation output compared to three-phase or high-frequency systems due to greater voltage ripple.

Tube Loading

Refers to the heat capacity of an X-ray tube.

  • Excess heat = tube damage

  • 99% of energy = heat

  • Avoid max exposure on a cold anode

Components of the Radiographic Unit

Includes:

  • Tube

  • Table

  • Bucky (grid system)

  • Support systems (ceiling, C-arm, etc.)

Protective Housing

  • The metal (lead) exterior housing minimizes isotropic X-ray radiation, prevents electric shock, and dissipates heat. High-capacity tubes use a metal enclosure.

  • The Pyrex glass enclosure withstands the high heat generated.

  • The tube maintains a vacuum for efficient X-ray production, known as the Coolidge tube.

X-ray Tube Construction


Cathode (−)

  • Filament + focusing cup

  • Produces electrons via thermionic emission

Key concepts:

  • Small focal spot → better detail

  • Large focal spot → handles more heat

  • Space charge effect: electron cloud repulsion at the cathode limits the number of electrons reaching the anode, especially at low kVp.

Anode (+)

  • Receives electrons → produces x-rays

  • Types:

    • Stationary used low power units 

    • Rotating (most common) for high intensity 

Functions:

  • Conduct electricity

  • Dissipate heat with the material 

  • Produce x-rays

Electron source

  • Defined as the filament (cathode) made of tungsten

  • Current (mA) flows through the filament

  • Filament heats up

  • Electrons are “boiled off” → thermionic emission

  • Electrons form a space charge cloud near the filament

mA controls how many electrons → affects X-ray quantity

Forms a space charge cloud that can limit electron flow. The tungsten filament (cathode) is heated by electric current (mA).

Target Material:

  • Tungsten Atomic number of Tungsten (W) = 74

    • High atomic number

    • High melting point
      Good heat dissipation

Secondary Materials

Rhenium (Re)

  • Added to tungsten (≈ 10%)

  • Prevents cracking and pitting

  • Increases durability

Molybdenum (Mo)

  • Used in:

    • Mammography tubes

  • Produces lower energy x-rays

  • Improves soft tissue contrast

Rhodium (Rh)

  • Also used in mammography

  • Slightly higher energy than molybdenum

  • Used for denser breast tissue

Induction Motor

Definition

  • A device that spins the rotating anode in the X-ray tube

  • Uses electromagnetic induction (no direct contact)

Components

Stator (outside)

  • Located outside the glass tube

  • Contains electromagnets

  • Creates a rotating magnetic field

Rotor (inside)

  • Located inside the tube (vacuum)

  • Made of copper and iron

  • Connected to the anode disk

  • Spins when a magnetic field is applied

How It Works (Step-by-Step)

  1. Stator is energized

  2. Produces a rotating magnetic field

  3. The magnetic field interacts with the rotor

  4. The rotor begins to spin

  5. Anode rotates → spreads heat

Purpose

  • Allows high tube current (mA)

  • Prevents overheating

  • Distributes heat over a larger target area

Filtration

Purpose of Filtration

  • Removes low-energy X-rays

  • Improves beam quality (hardens beam)

  • Reduces patient dose

Types of Filtration

1. Inherent Filtration

Definition:
Filtration is built into the X-ray tube.

Includes:

  • Glass envelope

  • Insulating oil

  • Tube window

Amount:

  • Approximately 0.5 – 1.0 mm Al equivalent

2. Added Filtration

Definition:
Filtration was intentionally added to the beam.

Material:

  • Aluminum

Location:

  • Collimator or tube exit

Amount:

  • Usually 1.5 – 2.0 mm Al

3. Total Filtration

Definition:
Sum of inherent + added filtration

 Formula:
Total = Inherent + Added

ARRT Requirement:

  • Minimum 2.5 mm Al equivalent (for >70 kVp) 

4. Compensating Filtration (Extra)

Used to balance uneven anatomy.

Examples:

  • Wedge

  • Trough

  • Boomerang


Automatic Exposure Control (AEC)

Stops exposure when enough radiation reaches the detector.

Controls:

  • Exposure time only

Technologist sets:

  • kVp

  • mA

  • chamber selection

  • Requires excellent positioning 

  • Not used with casts or prosthetics 

Backup Timer

  • Stops exposure if AEC fails

  • Shortest time with an AEC is 1ms 

  • Prevents overexposure

  • Set at 150% of expected exposure time

mAs Timer

  • Controls exposure by using mA × time (mAs)

  • Automatically adjusts time to reach the selected mAs

  • Ensures accurate total radiation output

  • Located the secondary coil of a high voltage transformer 

  • Used with falling load generators. 

Key idea: You set mAs, and the system figures out the time

Electronic Timer

  • Uses electronic circuits (no moving parts)

  • Provides very precise exposure times (milliseconds)

  • Most common timer in modern X-ray units

  • Microprocesador controlado. 

  • Allows times of 1 ms 

Example

  • Expected exposure time = 0.2 sec

  • Backup timer = 0.3 sec (150%)

Exposure Adjustment (Density Control)

  • Adjusts exposure by ±25–30% per step

+1 → darker
−1 → lighter

Minimum Exposure Controls

  • Prevents too short an exposure time

  • Ensures AEC works properly

Manual Exposure Controls

The technologist manually sets:

  • kVp

  • mA

  • time

Used when:

  • AEC not appropriate

  • small anatomy

  • portable exams

  • Prosthetics/casts

  • Poor positioning

Manual Exposure Controls – Quick Chart

Step

Control

What It Does

How to Set

ARRT Tip

1

kVp

Controls beam quality (penetration & contrast)

Choose based on body part (high for chest, low for extremities)

↑ kVp = ↓ contrast, ↑ penetration

2

mAs

Controls beam quantity (image exposure & noise)

Adjust based on patient size

↑ mAs = ↑ exposure, ↓ noise

3

mA

Controls electron flow rate

Choose station (100, 200, 300+)

Higher mA = shorter time

4

Time (sec)

Length of exposure

Keep it as short as possible

Short time = ↓ motion blur

5

mAs Calculation

Total exposure

mA × time

Must be correct for a proper image

6

Patient Adjustment

Adjust for size/pathology

Increase or decrease technique

Additive = ↑, Destructive = ↓

7

SID

Distance affects intensity

Use standard (e.g., 40” or 72”)

↑ SID = ↓ exposure

8

Grid Use

Reduces scatter, improves contrast

Use if >10 cm thickness

Requires ↑ mAs

9

Focal Spot

Affects detail

Small = detail, Large = heat

Small focal spot = sharper image

Radiation Detectors

Used in AEC to measure radiation.

Types:

Radiation Detectors – Condensed Study Guide

1. Gas-Filled Detectors

  • Contain gas that becomes ionized by radiation

  • Measure radiation exposure

  • Types include:

    • Ionization chambers

    • Geiger counters (less relevant for imaging)

2. Ionization Chambers (MOST IMPORTANT)

  • Most common AEC detector

  • Gas-filled device
    X-rays ionize gas → creates an electrical signal
    Controls exposure time (AEC)

  • Requires accurate positioning and proper chamber selection

3. Photomultiplier Tubes (PMT)

  • Used in older AEC systems
    Detect light from the fluorescent screen

  • Convert light → electrical signal

  • Largely replaced by ionization chambers

4. Scintillation Detectors

  • Convert x-rays → light → electrical signal

  • Use a scintillator (crystal)

  • Common in:

    • CT

    • Nuclear medicine

5. Semiconductor / Solid-State Detectors

(These are closely related and often grouped)

Definition:

  • Use solid materials (no gas) to convert radiation → electrical signal

Characteristics:

  • Fast response

  • High sensitivity

  • More efficient than gas detectors

Used in:

  • Modern AEC systems

  • Digital radiography (DR)

12. Tube Failure & Heat Management

Causes of Failure:

  • Excess heat (MOST COMMON)

  • Bearing damage

  • Tungsten evaporation

Heat Dissipation:

  1. Radiation

  2. Conduction

  3. Convection

COMPUTED RADIOGRAPHY (CR)

Definition

A cassette-based digital imaging system that uses a reusable imaging plate and a separate plate reader.

Key Components

Imaging Plate (IP)

  • Contains Photostimulable Phosphor (PSP)

  • Stores energy from X-ray exposure as a latent image

Photostimulable Phosphor (PSP)

  • Usually barium fluorohalide (BaFBr: Eu)

  • Traps electrons when exposed to X-rays

  • Releases energy as light when stimulated by a laser

Plate Reader

  • A device that processes the imaging plate

  • Uses a laser to scan the plate

  • Converts stored energy into a digital image

System Breakdown

Mechanical

  • Cassette

  • Imaging plate

  • Plate transport system (rollers)

Optical

  • Laser beam

  • Mirrors

  • Light guide

Computer

  • Photomultiplier tube (light detection)

  • Analog-to-digital converter (ADC)

  • Image processing software

  • PACS storage

Process

  1. X-ray exposure → PSP stores energy

  2. Plate inserted into the reader

  3. Laser stimulates PSP → emits light

  4. Light converted to electrical signal

  5. Signal digitized → image displayed

DIGITAL RADIOGRAPHY (DR)

Definition

A cassette-less digital system that produces an image immediately using a flat panel detector.

DIRECT CONVERSION DR

Key Material

  • Amorphous Selenium (a-Se)

Function

  • Converts X-rays directly into an electrical signal

  • No intermediate light step

System Breakdown

Capture

  • a-Se absorbs x-rays → electrical charge

Coupling

  • An electric field moves a charge

Collection

  • Thin-film transistor (TFT) array collects the signal

Components

Mechanical

  • Flat panel detector

Optical

  • None

Computer

  • TFT electronics

  • ADC

  • Image processing system

INDIRECT CONVERSION DR

Key Material

  • Amorphous Silicon (a-Si)

Function

  • Converts X-rays into light, then into an electrical signal

System Breakdown

Capture

  • Scintillator (CsI or Gd₂O₂S) converts x-rays → light

Coupling

  • a-Si photodiodes convert light → electrical signal

Collection

  • TFT array stores and reads the signal

Components

Mechanical

  • Flat panel detector

Optical

  • Scintillator (light production)

Computer

  • TFT readout

  • ADC

  • Image processing

AMORPHOUS SILICON (a-Si)

Definition

  • Semiconductor used in indirect DR

Function

  • Converts light into an electrical signal

Key Point

  • Always associated with indirect conversion

CHARGE-COUPLED DEVICE (CCD)

Definition

  • A digital detector that converts light into an electrical signal using a chip

System Breakdown

Capture

  • Scintillator converts x-rays → light

Coupling

  • Lens or fiber optics focuses light

Collection

  • CCD chip converts light → electrical signal

Components

Mechanical

  • Lens alignment system

Optical

  • Lens system

  • Mirrors or fiber optics

Computer

  • ADC

  • Image processing electronics

Key Points

  • Requires lens system

  • Smaller field of view

  • Older technology compared to flat panel detectors

COMPLEMENTARY METAL-OXIDE SEMICONDUCTOR (CMOS)

Definition

  • Modern digital detector similar to CCD but more efficient

System Breakdown

Capture

  • Scintillator converts x-rays → light

Coupling

  • Fiber optics or direct coupling transfers light

Collection

  • CMOS sensor converts light → electrical signal

  • Each pixel has its own amplifier

Components

Mechanical

  • Compact detector housing

Optical

  • Scintillator

  • Fiber optic coupling (if present)

Computer

  • On-chip amplification

  • Fast readout electronics

  • ADC

  • Image processing

Key Points

  • Faster than CCD

  • Lower power consumption

  • Less noise

  • More durable

  • Used in modern systems




System

Capture

Coupling

Collection

CR

PSP

Lens/Fiber Optics

Photodetector 

Indirect DR

CsI, Scintillator

Fiber Optics

CCD, CMOS

Direct DR

A-Se

Contact Layer

TFT

Indirect DR

CsI, GdOs

Contact Layer

TFT

BEAM RESTRICTION

Definition

Beam restriction is the limitation of the X-ray beam to the area of clinical interest.

PURPOSE 

1. Reduce Patient Dose

  • Limits the amount of tissue exposed to radiation

2. Reduce Scatter Radiation

  • Smaller field size → less scatter produced

3. Improve Image Quality

  • Less scatter → increased image contrast

METHODS OF BEAM RESTRICTION

1. Collimator (MOST IMPORTANT)

Description

  • Adjustable lead shutters inside the X-ray tube housing

  • Produces a rectangular light field

Components

Mechanical

  • Lead shutters

  • Adjustment knobs

Optical

  • Light bulb

  • Mirror (reflects light to simulate x-ray field)

Computer/Electrical

  • Light circuit (for field visualization)

Key Functions

  • Adjust field size

  • Align light field with x-ray field

  • Improve positioning accuracy

2. Cones

Description

  • Cylindrical or conical metal devices

  • Attach to tube housing

Function

  • Restrict beam to a circular field

3. Cylinders

Description

  • Similar to cones but longer and open-ended

Function

  • Provide more precise restriction

  • Often used in dental imaging

COLLIMATOR ACCURACY (QUALITY ASSURANCE)

Light Field vs X-ray Field Alignment

Requirement

  • Light field must match x-ray field within ±2% of SID

Example

  • At 100 cm SID → allowable error = 2 cm

Why This Matters

  • Prevents:

    • Cutting off anatomy (under-collimation)

    • Unnecessary exposure (over-collimation)

BEAM RESTRICTION AND SCATTER

Relationship

  • Increased field size → increased scatter

  • Increased scatter → decreased contrast

Key Concept

Beam restriction:

  • Decreases Compton scatter reaching the detector

  • Improves visibility of anatomical detail

PENUMBRA EFFECT

Definition

  • Blurring at the edges of the image

Beam Restriction Effect

  • Smaller field size → reduced penumbra → sharper edges

POSITIVE BEAM LIMITATION (PBL)

Definition

  • Automatic collimation system

Function

  • Adjusts beam size to match image receptor size

Requirements

  • Must be accurate within 3% of SID

Limitation

  • Does not replace proper manual collimation

QUALITY ASSURANCE TESTS

1. Collimator Alignment Test

  • Checks light field vs x-ray field accuracy

2. PBL Test

  • Ensures automatic collimation accuracy

3. Beam Centering Test


Ensures central ray aligns with image receptor

FLUOROSCOPIC UNIT

Components of a Fluoroscopic Unit

Fixed Fluoroscopy Unit

Mechanical

  • X-ray tube (under or over table)

  • Fluoroscopic table (tilting or stationary)

  • Image receptor (image intensifier or flat panel)

  • Tower (supports IR)

  • Compression device

Electrical

  • Generator

  • Control panel

  • Automatic Brightness Control (ABC/AERC)

Computer

  • Digital processing system

  • Image storage (PACS)

  • Post-processing workstation

Mobile Fluoroscopy Unit (C-Arm)

Mechanical

  • C-arm structure (holds tube and IR)

  • X-ray tube

  • Image receptor (II or flat panel)

Electrical

  • Generator (smaller, less powerful than fixed)

Computer

  • Monitor display

  • Image storage system

  • Limited post-processing

Key Details: it is better to have the tube below the patient to hit the bottom of aprons like resistant gonads  than the tube above to avoid unnecessary exposure to sensitive areas like eyes and thyroid 

IMAGE RECEPTORS (IR)

1. Image Intensifier (II)

Purpose

  • Converts low-intensity x-ray image into a brighter visible image

Components of Image Intensifier

Input Phosphor

  • Cesium iodide (CsI)

  • Converts x-rays → light

Photocathode

  • Converts light → electrons

Electrostatic Focusing Lenses

  • Focus electron beam toward output phosphor

Accelerating Anode

  • Speeds up electrons

Output Phosphor

  • Converts electrons → visible light

Key Concepts

Minification Gain

  • Smaller output phosphor compared to input phosphor

  • Increases brightness

Formula:
Minification Gain = (Input diameter² / Output diameter²)

Flux Gain

  • Acceleration of electrons increases brightness

Total Brightness Gain

Total Brightness Gain = Minification Gain × Flux Gain

  • Typical range: 5,000 – 20,000

  • Decreases as the tube ages

Conversion Factor

  • Efficiency of II

  • Ratio of output brightness /  input exposure

2. Flat Panel Detector (Digital Fluoroscopy)

Advantages

  • Higher spatial resolution (better small detail)

  • Larger field of view (some systems)

  • No geometric distortion

  • Improved post-processing

  • More uniform image quality

Replacement

  • Replaces the image intensifier in modern systems

VIEWING AND RECORDING SYSTEMS

Image Distribution

  • Image is sent from the image receptor to:

    • Monitor (TV or LED display)

    • Recording systems

Viewing Systems

  • Television monitor (older systems)

  • LED/LCD monitors (modern systems)

Recording Systems

  • Digital storage (PACS)

  • Hard drive storage

  • Video recording (older systems)

Digital Fluoroscopy

  • Similar x-ray equipment as conventional fluoroscopy

  • Includes:

    • Operating console

    • Computer workstation

  • Allows:

    • Image storage

    • Image processing

    • Frame capture

AUTOMATIC BRIGHTNESS CONTROL (ABC / AERC)

Definition

  • Automatically adjusts exposure to maintain constant image brightness

How It Works

Adjusts:

  • kVp

  • mA

  • Pulse rate (in pulsed fluoro)

Purpose

  • Maintain consistent image brightness

  • Compensate for patient thickness/density

MAGNIFICATION MODE

Requirements

  • Requires dual-focus or trifocus x-ray tube

How It Works

  • Uses a smaller portion of the input phosphor

  • Changes voltage in image intensifier

Effects

Advantages

  • Increased spatial resolution

Disadvantages

  • Increased patient dose (very testable)

  • Decreased brightness (ABC compensates by increasing exposure)

FLUOROSCOPIC UNIT TABLE

Types of Fluoroscopy Tables

Type

Description

Key Use

Tilting Table

Can move from horizontal to vertical positions

GI studies (barium swallow, UGI)

Stationary Table

Fixed horizontal position

Interventional procedures

TABLE COMPONENTS

Mechanical Components

Component

Description

Function

Tabletop

Radiolucent surface

Supports patient while allowing x-rays to pass

Tilting Mechanism

Motorized system

Changes table angle (vertical ↔ horizontal)

Footboard

Support at end of table

Prevents patient from falling when upright

Shoulder Supports

Adjustable supports

Stabilize patient in upright position

Compression Device

Paddle or cone

Compresses abdomen to improve image quality

Tower

Supports IR (II or flat panel)

Moves vertically and locks into position

X-ray Tube Position

Configuration

Description

Advantage

Under-table tube

Tube below table, IR above

Reduces patient skin dose

Over-table tube

Tube above table, IR below

Easier access for procedures but ↑ dose

IMAGE RECEPTOR MOVEMENT

Feature

Description

Vertical Movement

IR moves up/down to maintain proper distance

Centering

Aligns IR with anatomy

Locking Mechanism

Keeps IR stable during exposure

TABLE OPERATION FEATURES

Controls

Control Type

Function

Table tilt control

Adjusts table angle

Tower movement control

Moves IR vertically

Fluoro foot pedal

Activates x-ray beam

ABC/AERC system

Maintains brightness automatically

Safety Features

Feature

Purpose

Dead-man switch

Requires continuous pressure to produce x-rays

Interlocks

Prevent improper movement or exposure

Footboard & supports

Prevent patient falls

PATIENT POSITIONING FEATURES

Feature

Purpose

Trendelenburg position

Head down for certain GI studies

Upright position

Swallowing and esophageal studies

Supine position

Standard imaging

IMAGE QUALITY + DOSE RELATION

Factor

Effect

Under-table tube

↓ patient dose

Over-table tube

↑ patient dose

Compression

↓ scatter, ↑ contrast, ↓ dose

Proper centering

Prevents repeats

3. COUNTERWEIGHTS (STATIONARY FLUORO)

Purpose

  • Balance x-ray tube and image receptor

  • Prevent drifting or “floating”

  • Allow smooth controlled movement

Key Points

  • Located inside support system

  • Improve positioning accuracy

  • Reduce mechanical strain

Lead Curtain

Property

Value

Location

Between patient and operator

Thickness

0.25–0.5 mm Pb

Reduction

~90–95%

Bucky Slot Cover

Property

Value

Location

Table slot opening

Thickness

0.25 mm Pb

Reduction

~90%

FLUOROSCOPY ACCESSORIES

1. STATIONARY GRIDS

Definition

A grid placed between the patient and image receptor to reduce scatter radiation and improve image contrast.

Structure

  • Made of lead strips separated by radiolucent material (aluminum or plastic)

  • Strips are aligned to allow primary beam through while absorbing scatter

Key Characteristics

Feature

Description

Type

Stationary (does not move in fluoro)

Grid ratio

Usually lower than radiography (e.g., 6:1–8:1)

Line density

Number of lead strips per inch

Function

  • Absorbs scatter radiation

  • Improves image contrast

Disadvantages

  • Increases patient dose (more exposure needed)

  • Can produce grid lines (less noticeable in fluoro)

2. BUCKY ASSEMBLY

Definition

A device that holds the grid and image receptor and allows proper positioning during imaging.

Components

Component

Function

Grid

Absorbs scatter

Tray

Holds image receptor

Bucky slot

Opening for IR movement

Bucky slot cover

Lead shield to reduce scatter

Function

  • Aligns grid with x-ray beam

  • Holds image receptor securely

  • Allows movement of IR during procedures

In Fluoroscopy

  • Usually integrated into the table or tower

  • May include stationary grid (not moving like radiographic Bucky)

High-Yield Points

  • Bucky assembly includes grid + IR holder

  • Bucky slot cover reduces operator exposure (~90%)

3. COMPENSATING FILTERS

Definition

Devices placed in the x-ray beam to even out differences in patient thickness/density.

Purpose

  • Produce more uniform image exposure

  • Prevent overexposure of thinner areas

  • Improve contrast

Types 

Type

Shape

Use

Wedge filter

Thick → thin gradient

Extremities, abdomen

Trough filter

Thick edges, thin center

Chest

Boomerang filter

Curved shape

Shoulder, chest

Function

  • Absorbs more x-rays in thinner areas

  • Allows more x-rays through thicker areas

Benefits

  • Improves image quality

  • Reduces patient dose in thinner regions

Image Processing and Display 

1. Raw Data (Pre-processing)

  • This is the initial signal captured by the detector.

  • Still unprocessed → contains noise, inconsistencies, and detector imperfections.

2. Analog-to-Digital Converter (ADC)

  • Converts the signal from analog (continuous) → digital (discrete numbers).

  • Each pixel gets a numerical value representing intensity.

3. Quantization

  • Assigns each pixel a discrete value (bit depth).

  • Example:

    • 8-bit = 256 shades of gray

    • 12-bit = 4096 shades (more detail!)

More bits = better contrast resolution

4. Corrections

Fixes imperfections in the raw image:

  • Rescaling → adjusts pixel values to usable range

  • Flat fielding → corrects uneven detector response

  • Dead pixel correction → replaces non-functioning pixels

Goal: make the image accurate and uniform

5. Histogram

  • A graph of pixel intensity distribution

  • Shows how many pixels fall into each brightness level

 Used to:

  • Evaluate exposure

  • Guide processing decisions

6. Corrected Data for Processing

  • Now the image is clean and standardized

  • Ready for enhancement algorithms

7. Image Processing (Enhancement Techniques)

Grayscale

  • Assigns pixel values to shades of gray

  • Determines overall image brightness/contrast

Edge Enhancement

  • Sharpens borders between structures

  • Makes anatomy easier to distinguish

Example: bone edges appear crisper

Equalization

  • Redistributes grayscale values

  • Improves contrast in areas with poor visibility

Smoothing

  • Reduces noise (graininess)

  • Makes image look cleaner—but can slightly blur detail

8. Data for Display

  • Final processed data ready to be shown on the monitor

9. Values of Interest (VOI)

  • Selects the range of pixel values you want to emphasize

This is where:

  • Window level (brightness)

  • Window width (contrast)

are applied

10. Look-Up Table (LUT)

  • Maps pixel values → display brightness

Determines:

  • How the image actually appears on the screen

  • Can invert or adjust contrast

Post-Processing

Happens AFTER the image is acquired and processed
Does NOT change the raw data, only how it’s displayed

Brightness

  • Controls overall lightness/darkness of the image

  • Adjusted using window level (WL)

 Higher brightness = lighter image

Contrast

  • Controls the difference between light and dark areas

  • Adjusted using window width (WW)

Narrow WW = high contrast (more black & white)
Wide WW = low contrast (more grays)

Region of Interest (ROI)

  • Selected specific area of the image for analysis or processing

  • Used for:

    • Exposure evaluation

    • Histogram analysis

    • Measurements

Incorrect ROI selection = wrong brightness/contrast

Electronic Cropping (Masking)

  • Hides unwanted areas of the image

Important:

  • Does NOT remove data, only covers it

  • Such as the white border from cropping, but should not be used to hide uncollimated anatomy. 

  • The original image can still be retrieved

Used to:

  • Improve appearance

  • Reduce distractions

Stitching

  • Combines multiple images into one long image

Common uses:

  • Full spine imaging

  • Long bone studies

5. Display Monitors

Final step—how the image is actually seen by the radiologist

Viewing Conditions

Viewing Angle

  • Image quality decreases if viewed off-angle
    Best viewed straight on

Ambient Lighting

  • Room lighting affects visibility

Ideal:

  • Low, controlled lighting

  • Too bright = reduces contrast perception

Spatial Resolution (Monitor Quality)

Pixel Size

  • Size of each pixel on the monitor

Smaller pixels = sharper image

Pixel Pitch

  • Distance between pixels

Smaller pitch = higher resolution

Brightness and Contrast (Monitor Performance)

  • Monitors must display a wide range of grayscale values

High-quality monitors:

  • Better contrast resolution

  • More accurate diagnosis 

6. IMAGING INFORMATICS

1. INFORMATION SYSTEMS

HIS (Hospital Information System)

Definition

  • The main system that manages the entire hospital operations

Includes

  • Patient registration

  • Billing

  • Scheduling

  • Medical records

RIS (Radiology Information System)

Definition

  • Specialized system for radiology workflow

Functions

  • Exam scheduling

  • Tracking imaging orders

  • Reporting (radiologist reports)

  • Patient tracking within radiology

EMR (Electronic Medical Record)

Definition

  • Digital version of a patient’s chart within one facility

Key Point

  • Not easily shared outside that facility

EHR (Electronic Health Record)

Definition

  • Comprehensive patient record that can be shared across multiple facilities

Includes

  • History

  • Imaging

  • Lab results

  • Medications





System

Scope

HIS

Entire hospital

RIS

Radiology only

EMR

One facility

EHR

Multiple facilities

2. NETWORKING

Definition

  • A system that connects computers and devices to share data

Key Components

Component

Function

Server

Stores and manages data

Workstation

User access point

Network

Communication system

Types

Type

Description

LAN

Local (hospital) network

WAN

Wide (multiple locations)

Key Concepts

  • Enables image transfer and access

  • Supports PACS and RIS integration

3. PACS (Picture Archiving and Communication System)

Definition

  • System used to store, retrieve, and distribute images

Components

Component

Function

Image acquisition

Modality (CT, XR, etc.)

Storage server

Stores images

Workstations

Viewing images

Network

Transfers images

Functions

  • Image storage

  • Image retrieval

  • Image display

  • Image sharing

Advantages

  • Eliminates film

  • Faster access

  • Remote viewing

4. DICOM (Digital Imaging and Communications in Medicine)

Definition

  • Standard format for storing and transmitting medical images

Purpose

  • Ensures compatibility between:

    • Modalities

    • PACS

    • Workstations

Key Features

  • Standard file format

  • Includes:

    • Patient data

    • Image data

  • Enables communication between different systems

5. DOWNTIME PROCEDURES 

Definition

  • Backup procedures used when computer systems fail

When Used

  • PACS failure

  • Network outage

  • System maintenance

Procedures

  • Use paper requisitions

  • Manual patient identification

  • Temporary image storage (local system)

  • Document exams manually

  • Enter data into system once restored

Key Goals

  • Maintain workflow

  • Prevent data loss

  • Ensure patient safety

BEAM RESTRICTION

Definition

Limiting the x-ray beam to the area of interest using collimation.

Purpose

  • Reduce patient dose

  • Reduce scatter radiation

  • Improve image contrast

Quality Control Check

  • Field size should match area of interest

  • Collimator shutters must function properly

  • Light field must accurately represent x-ray field

Key Relationship

  • Smaller field size → less scatter → higher contrast

LIGHT FIELD TO RADIATION FIELD ALIGNMENT

Definition

Ensures the visible light field matches the actual x-ray beam field.

Why It Matters

  • Prevents:

    • Cutting off anatomy (under-collimation)

    • Excess exposure (over-collimation)

  • Reduces repeat exams

Alignment Tolerance 

Must be within ±2% of SID

Example

  • SID = 100 cm

  • 2% = 2 cm

→ Light field and x-ray field can differ by no more than 2 cm

How It Is Tested

  • Place test tool (collimator test plate or markers)

  • Expose image

  • Compare light field edges to radiation field edges

3. CENTRAL RAY ALIGNMENT

Definition

Ensures the central ray (CR) is aligned with the center of the image receptor (IR).

Purpose

  • Prevents image distortion

  • Ensures proper centering

  • Avoids anatomy cutoff

Tolerance (VERY TESTABLE)

  • Must be within ±1% of SID

Example

  • SID = 100 cm

  • 1% = 1 cm

→ CR must be within 1 cm of IR center

How It Is Tested

  • Use alignment test tool

  • Check if CR hits exact center of IR

SUMMARY TABLE

Test

What It Checks

Tolerance

Beam restriction

Field size control

Proper collimation

Light vs radiation field

Field alignment

±2% SID

Central ray alignment

CR to IR center

±1% SID

RECOGNITION AND REPORTING OF MALFUNCTIONS

Purpose

  • Maintain image quality

  • Ensure patient safety

  • Prevent repeat exposures

Common Signs of Malfunction

Category

Signs

Image

Noise, artifacts, poor contrast, distortion

Detector

Dead pixels, lines, non-uniform brightness

CR Plates

Ghost images, scratches, debris

Monitor

Dim display, poor grayscale, non-uniform brightness

Equipment

Error messages, delays, unusual sounds

Technologist Responsibilities

  • Recognize abnormal performance

  • Stop using unsafe equipment

  • Report immediately (supervisor/biomedical)

  • Document issue per protocol

DIGITAL IMAGING RECEPTOR SYSTEM QC

Systems

  • CR (PSP plates)

  • DR (flat panel detectors)

Common Problems

Problem

Cause

Artifacts

Dust, dirt, scratches

Ghost image

Incomplete erasure (CR)

Non-uniform image

Calibration issue

Dead pixels

Detector damage

QC Procedures

  • Detector calibration (flat-field correction)

  • Uniformity checks

  • Artifact evaluation

  • Exposure indicator accuracy

MAINTENANCE

Erasure Thoroughness (CR)

  • Ensures no residual image remains

  • Poor erasure → ghost artifacts

Plate Uniformity (CR)

  • Ensures consistent response across plate

  • Issues → shading or uneven brightness

Spatial Resolution

  • Ability to see fine detail

  • Tested with line pair patterns

DISPLAY MONITOR QUALITY ASSURANCE

GSDF (Grayscale Standard Display Function)

  • Ensures consistent grayscale display

  • Based on DICOM standard

  • Equal pixel changes → equal perceived brightness

Luminance

  • Brightness of monitor

  • Measured in cd/m²

  • Typical range: 50–300 cd/m²

Illuminance

  • Ambient light

  • Measured in lux

Measurement Tools

Tool

Function

Photometer

Measures luminance

Near-range

Close measurement

Telescopic

Distance (~1 m)

Lux meter

Measures ambient light

Colorimeter

Measures color accuracy



Display QC Tests (AAPM TG-18)

Test

Purpose

TG-18 QC

General image quality

TG-18 AD

Reflection

TG-18 LPV/LPH

Distortion

TG-18 CX

Resolution

TG-18 AFC

Noise

Common Monitor Issues

Issue

Effect

Low luminance

Poor visibility

Non-uniform brightness

Missed pathology

Reflection

Reduced contrast

Distortion

Shape inaccuracies



DISPLAY PERFORMANCE STANDARDS

SMPTE

  • Tests contrast and resolution

  • 5% and 95% patches must be visible

DICOM (GSDF)

  • Standardizes grayscale display

  • Ensures consistency across systems

AAPM TG-18

  • Primary QC standard

  • Provides test patterns and procedures

Other Standards

Standard

Purpose

DIN 2001

Image QA and acceptance testing

VESA

Flat panel performance standards

SHIELDING ACCESSORIES QC

Includes

  • Lead aprons

  • Thyroid shields

  • Lead gloves

Purpose

  • Protect from radiation exposure

Testing Methods

Visual Inspection

  • Check for cracks, tears, damage

Radiographic/Fluoroscopic Testing

  • Detect internal cracks

Frequency

  • Typically annually

Common Problems

Problem

Risk

Cracks

Radiation leakage

Worn shielding

Reduced protection

Action

  • Remove defective shielding immediately

QUALITY CONTROL PROGRAM STRUCTURE

Required Components

  • Acceptance testing (new equipment)

  • Routine QC (technologist)

  • Periodic review (physicist)

  • Annual evaluations

  • Post-repair testing

RESPONSIBILITIES

Role

Responsibility

Technologist

Routine QC, identify/report issues

Medical Physicist

Acceptance testing, annual QA, program oversight

VII. Organizational Factors Affecting Medical Image Quality

  • A. IR (Image Receptor) Factors

    • Pixel size, dynamic range, intensity response, signal-to-noise ratio, and post-processing.

  • B. Geometric Factors

    • Distortion, magnification, and blur.

  • C. Subject Factors

    • Contrast (influenced by thickness, density, and atomic number) and motion.

SUMMARY

  • Ghost image = incomplete CR erasure

  • Dead pixels = DR detector issue

  • GSDF = consistent grayscale display

  • Luminance = brightness (cd/m²)

  • Illuminance = ambient light (lux)

  • TG-18 = primary QC testing system

  • SMPTE = 5% and 95% visibility test

  • Cracked shielding = remove immediately

  • The technologist reports malfunctions immediately