Image Production Capstone Unit 3: March 30th
Technique Charts
Standardized guidelines are used to select kVp and mAs for specific exams to ensure consistent image quality and reduce repeats.
Includes:
kVp
mAs
SID
Grid use
Patient thickness
Types:
Fixed kVp
Variable kVp
Anatomically Programmed Radiography (APR)
Anatomically Programmed Radiography (APR)
A computerized system where a technologist selects body part + projection and the machine automatically sets exposure factors.
Advantages:
Faster setup
Fewer technique errors
Consistent exposures
Technologists must still adjust for:
patient size
pathology
casts
Age
Fixed vs Variable kVp
Fixed kVp
kVp stays constant
mAs changes with patient thicknessProduces consistent contrast
Variable kVp
kVp changes with thickness
Rule: 2 kVp change per 1 cm thickness
mAs stays constant
ARRT Tip: Fixed kVp = most consistent contrast
Special Considerations
Casts
Increase exposure because casts absorb X-rays.
Fiberglass cast: +3–4 kVp
Plaster cast:
Dry: +5–7 kVp
Wet: +8–10 kVp
Pathologic Factors
Additive Disease
Increases tissue density → increase technique
Examples:
pneumonia
ascites
edema
tumors
Destructive Disease
Decreases tissue density → decrease technique
Examples:
osteoporosis
emphysema
Memory tip: Additive = Add exposure
Age Considerations
Pediatrics
Smaller body size
Higher radiation sensitivity
Technique:
Reduce exposure
Collimation
Use shielding
Short exposure times
Geriatrics
Often decreased bone density → may require slightly reduced technique, but pathology may increase exposure.
Body Mass Index (BMI)
Higher BMI = more tissue attenuation.
Adjustments:
Increase kVp or mAs
Lower BMI = decreased exposure.
Contrast Media
Positive Contrast
Absorbs x-rays → appears white
Examples:
barium
iodine
Negative Contrast
Allows x-rays to pass → appears black
Examples:
air
CO₂
Grids
Purpose: Reduce scatter → improve image contrast
Used when body part thickness > 10 cm
Effects:
Improves contrast
Requires increased mAs
Higher grid ratio:
better scatter cleanup
Higher patient dose
OID (Object-to-Image Distance)
OID = distance between anatomy and image receptor.
Increased OID allows scatter, which travels widely in divergent paths to exit the patient and miss the IR.
Air gap technique seen on lateral c spine doesn’t require a grid similar to 10:1
Increasing OID causes:
Magnification ↑
Recorded detail ↓
Best image quality:
Small OID
Large SID
Small focal spot
Automatic Exposure Control (AEC)
Definition:
AEC automatically terminates the exposure when the detector receives a preset amount of radiation.
Purpose:
Produces consistent receptor exposure
Reduces technique errors
Requires excellent positioning skills
Helps standardize images between technologists
Key components:
Ionization chambers (detectors)
Backup timer
Density control
Shortest time with an AEC is 1 ms
Important:
AEC controls exposure time, on a fixed kVp.
The technologist still selects:
kVp
mA
detector cell
density setting
Detector Selection
Most systems have 3 AEC chambers:
Left
Center
Right
The technologist chooses the chamber under the anatomy of interest.
Examples:
Chest PA
→ Both outer chambers
AP abdomen
→ Center chamber
Spine
→ Center chamber
ARRT tip:
Incorrect chamber selection → overexposure or underexposure.
Anatomic Alignment
For AEC to work correctly:
The anatomy of interest must completely cover the selected chamber.
If anatomy does NOT cover the chamber:
Exposure may terminate too early
The image may be underexposed
Example:
Chest rotated → lung may not cover the chamber → exposure error.
Backup Timer
Safety feature that stops exposure if AEC fails.
Purpose:
Prevents excessive patient dose
Typically set at:
150% of expected exposure time
Density Control (Exposure Adjustment)
Density control adjusts the amount of radiation reaching the detector.
Does NOT change contrast.
Often labeled:
+1
+2
−1
−2
Each step changes exposure by about 25–30%.
Example:
+1 → slightly darker image
−1 → slightly lighter image
Used for small exposure corrections.
Effects of Exposure Factors on Image Quality
kVp Effects
Increasing kVp:
Increases penetration
Decreases image contrast
Increases scatter
Decreasing kVp:
Less penetration
Higher contrast
Less scatter
ARRT rule:
15% rule
Increasing kVp by 15% ≈ doubles receptor exposure.
mAs Effects
mAs controls the quantity of x-rays.
Increasing mAs:
Increases receptor exposure
Increases patient dose
No effect on contrast
Decreasing mAs:
Decreases exposure
May produce quantum noise
Common AEC Errors (ARRT Favorites)
Wrong chamber selected
Improper positioning
Anatomy does not cover the chamber
Metal prosthesis over the chamber
Incorrect centering
These cause overexposure or underexposure.
Spatial Resolution
Definition:
The ability of an imaging system to distinguish small objects that are close together.
Also called:
Recorded detail
Measured in:
Line pairs per millimeter (lp/mm)
Higher spatial resolution
sharper images
better visibility of fine detail
Factors affecting spatial resolution:
Pixel size
Matrix size
Detector element size
Sampling frequency
MTF
Pixel Characteristics
Pixel
A pixel (picture element) is the smallest unit of a digital image.
Each pixel contains:
a numeric value representing signal intensity.
Pixel Size
Determined by:
Field of view ÷ matrix size
Smaller pixels:
increase spatial resolution
reduce signal per pixel
Larger pixels:
decrease spatial resolution
increase signal
Pixel Pitch
Distance from the center of one pixel to the center of the next pixel.
Smaller pitch:
higher resolution
Larger pitch:
lower resolution
Detector Elements (DEL)
A detector element (DEL) is the physical component of the detector that captures x-ray energy.
Each DEL corresponds to one pixel in the image matrix.
DEL Size
Smaller DEL size:
higher spatial resolution
Larger DEL size:
lower resolution
greater signal capture
Fill Factor
The percentage of the detector element that is sensitive to radiation.
Higher fill factor:
better detector efficiency
improved image quality
Lower fill factor:
less efficient detection
CCD and CMOS Detectors
Used in some digital imaging systems.
CCD (Charge-Coupled Device)
A digital detector that converts X-ray light into electrical signals and transfers the charge across the chip to produce a digital image.
Characteristics:
High image quality
Good sensitivity
Signal transferred across the chip
Limitation:
slower readout
CMOS (Complementary Metal-Oxide Semiconductor)
A digital detector that converts x-ray light into electrical signals and processes the signal directly at each pixel for faster image readout and lower power use.
Characteristics:
Faster image processing
Lower power consumption
Less electronic noise
Often used in modern digital detectors.
Sampling Frequency (CR)
Important in Computed Radiography (CR).
Definition:
Number of times the CR laser scans the imaging plate per millimeter.
Measured in:
pixels/mm
Higher sampling frequency:
higher spatial resolution
smaller pixel size
Lower sampling frequency:
reduced resolution
Matrix Size
Matrix = number of pixels in the image.
Example:
1024 × 1024
2048 × 2048
Total pixels = rows × columns.
Larger matrix:
smaller pixels
higher spatial resolution
Smaller matrix:
larger pixels
lower resolution
Modulation Transfer Function (MTF)
Definition:
MTF describes the ability of an imaging system to reproduce detail accurately.
It measures how well the system transfers object contrast to the image at different spatial frequencies.
Higher MTF:
better spatial resolution
more accurate image detail
Lower MTF:
loss of fine detail
MTF is considered the best overall measure of spatial resolution.
Contrast Resolution
Definition:
The ability of a digital imaging system to distinguish small differences in tissue density.
In other words:
Ability to see subtle differences in shades of gray.
Digital imaging has excellent contrast resolution compared to film.
Factors affecting contrast resolution:
Bit depth
Detector efficiency
Scatter radiation
Grids
Higher contrast resolution →
better visualization of soft tissues.
Example:
Detecting small differences between organs, fluid, and soft tissue.
Bit Depth
Definition:
The number of gray shades that can be displayed in a digital image.
Bit depth determines how much contrast information is stored.
Formula:
Number of gray shades = 2ⁿ
(n = number of bits)
Examples:
Higher bit depth:
better contrast resolution
more gray shades
ARRT Tip:
Digital systems often have 12–16-bit depth.
Detective Quantum Efficiency (DQE)
Definition:
A measure of how efficiently a detector converts incoming x-ray photons into useful image signals.
Higher DQE means:
Better image quality
Less radiation needed
Lower patient dose
In simple terms:
Higher DQE = better detector efficiency.
Typical comparison:
ARRT concept:
Higher DQE allows dose reduction while maintaining image quality.
Grids
Purpose:
Grids remove scatter radiation before it reaches the detector.
This improves image contrast.
Scatter causes:
fog
reduced contrast
When Grids Are Used
ARRT rule:
Use grids when body part thickness is greater than 10 cm.
Common exams using grids:
Abdomen
Pelvis
Lumbar spine
Adult chest
Grid Ratio
Grid ratio = height of lead strips ÷ distance between strips
Examples:
8:1
12:1
16:1
Higher grid ratios:
Advantages:
remove more scatter
improve contrast
Disadvantages:
increase patient dose
require higher mAs
more positioning accuracy
Grid Errors
Grid Cutoff
Occurs when primary radiation is absorbed by the grid.
Causes include:
Off-level grid
Off-center beam
Wrong SID
Upside-down grid
Grid misalignment
Result:
Reduced receptor exposure (lighter image)
Linear grids, focused grids, stationary gridsmoving gridsand and software that removes scatter from the image without the use of a grid.
Image Signal
Definition:
The useful information produced by x-ray photons detected by the imaging receptor.
In digital imaging:
X-rays interact with the detector
The detector converts them into an electronic signal
The signal becomes the digital image
Higher signal means:
better image visibility
improved image quality
Lower signal may lead to:
noisy images
Dynamic Range
Definition:
The range of exposure values a digital detector can accurately capture.
In simple terms:
The detector’s ability to record very low to very high exposures in one image.
Digital systems have a wide dynamic range.
Advantages:
fewer repeat exposures
ability to visualize structures with different densities
Example:
Both soft tissue and bone can be visible in the same image.
The film screen had a narrow dynamic range, meaning exposure had to be very precise.
Quantum Noise (Quantum Mottle)
Definition:
Random grainy appearance caused by too few X-ray photons reaching the detector.
Occurs when:
mAs is too low
Photon numbers are insufficient
Appearance:
speckled or grainy image
Important ARRT concept:
Quantum noise increases when:
exposure decreases
Solution:
increase mAs
Signal-to-Noise Ratio (SNR)
Definition:
The ratio of useful image signal to background noise.
Indicates overall image quality.
Higher SNR:
clearer image
less noise
better visibility of structures
Lower SNR:
noisy image
reduced diagnostic quality
Factors affecting SNR:
Increase SNR:
increase mAs
improve detector efficiency
reduce noise
Decrease SNR:
low photon exposure
quantum mottle
electronic noise
Image Identification
Purpose of Image Identification
Image identification ensures that radiographic images are correctly matched to the patient and examination.
It is required for:
legal documentation
accurate diagnosis
proper medical records
Improper identification can result in medical errors and legal liability.
Methods of Image Identification
Radiographic Identification
Information is placed directly on the image at the time of exposure.
Examples include:
Patient name
Date of exam
Time of exam
Facility name
Technologist IDAnatomical side marker (R or L)
Radiopaque markers are placed on the image receptor before exposure.
Important ARRT rule:
Lead side markers must be placed before exposure whenever possible.
Electronic Identification
Used in digital imaging systems.
Patient and exam information are added through:
PACS
Radiology information systems (RIS)
Electronic patient records
Data is embedded into the digital image file.
Legal Considerations
Radiographs are legal medical documents.
They must contain accurate patient and exam information.
Required patient data typically includes:
Patient name
Patient identification number
Date of examination
Facility identification
Anatomical side marker (R or L)
Additional exam data may include:
Projection or position
Technologist identification
Time of exam (sometimes required)
Important ARRT Legal Points
Images must be properly labeled before leaving the room.
Markers must indicate:
Right (R)
Left (L)
Annotations such as:
Upright
Supine
Decubitus
May also be included when necessary.
Failure to properly identify images can lead to:
incorrect diagnosis
repeat exams
legal consequences.
Criteria for Image Evaluation
Exposure Indicator
The exposure indicator (EI) is a numerical value that tells the technologist whether the correct amount of radiation reached the digital detector.
Too high → overexposure
Too low → underexposure / quantum noise
It helps technologists determine if the technique used was appropriate.
Quantum Noise (Quantum Mottle)
Quantum noise is a grainy or speckled appearance in the image caused by too few X-ray photons reaching the detector.
It usually occurs when mAs is too low, which reduces the signal and increases noise in the image.
Gross Exposure Error
A gross exposure error occurs when the image is severely overexposed or underexposed.
Overexposure can cause saturation, where image data is lost.
Underexposure causes excessive noise and poor image quality.
These errors may make the image non-diagnostic.
Contrast
Contrast is the difference in brightness between adjacent structures in the image.
It allows radiologists to distinguish between different tissues such as bone, soft tissue, and air.
Contrast is primarily influenced by kVp and scatter radiation.
Controlled by bit depth, the number of bits per pixel, the larger the bit depth, the greater the number of levels of gray
High contrast: few gray tones or short scale at lower kVp
Low contrast: many gray tones, long scale, and produced at higher kvp.
Subject contrast: controlled by the kVp, differential absorption by anatomy and pathology, and variation of atomic number.
Spatial Resolution
Spatial resolution is the ability of the imaging system to show small objects and fine details clearly.
High spatial resolution means the image appears sharp with clear edges, and closely spaced structures can be distinguished.
Controlled by pixel size and pixel pitch
Pixel limited
Influenced by SID, OID, focal spot size, and motion
Distortion (Size and Shape)
Distortion occurs when the size or shape of the anatomy is inaccurately represented on the image.
It is usually caused by improper alignment of the x-ray tube, patient, or image receptor, or by excessive OID (object-to-image distance).
Image Artifacts
Artifacts are unwanted objects or distortions that appear on the image but do not represent patient anatomy.
They can be caused by:
dirty detectors
patient motion
jewelry or clothing
processing or equipment errors.
Artifacts may interfere with diagnostic interpretation.
Radiation Fog (CR)
Radiation fog occurs when a computed radiography (CR) imaging plate is accidentally exposed to radiation before processing.
This results in a uniform gray appearance and reduced image contrast, which can degrade image quality.
The X-ray Imaging System
The three main components of an X-ray system are:
Operating Console
X-ray tube
High-voltage generator
The system operates at a voltage of 25kVp to 150kVp
The tube current at 100mA to 1200mA
A x-ray table is designated by the tilt of the table – most common for fluoroscopy tables is a 90-degree foot tilt and a 30-degree head tilt
Operating Console
Radiation quantity – the number of x-rays in a beam (milligray/milliampere-second: mGya/mAs)
Radiation quality – the energy of the X-ray beam (kilovolt peak: kVp, or half-layer value; HVL)
Automatic Exposure Control (AEC) terminates the exposure when the required amount of radiation intensity is reached.
All electric circuits on the operating console are low voltage
The voltage to an X-ray unit may vary as much as 5%. Line compensators are used to maintain a constant 220 volts to the equipment.
X-ray Generator, Transformers & Rectification
High-Voltage Generator
Used for increasing the voltage to the kilovoltage required for the X-ray imaging equipment
3 primary parts:
High-voltage transformer
Filament transformer
Rectifiers
Falling Load Generator – the X-ray equipment will choose the highest mA allowing for the shortest exposure time (mA x seconds = mAs)
A step-down transformer from the high-voltage supply, the secondary windings are insulated from the primary windings.
Disadvantages of long breathing exams
Autotransformer
Autotransformers
Supplies a precise voltage to the x-ray equipment's high-voltage circuit.
The primary and secondary voltages are directly related to the number of wire turns around the core.
Autotransformer Law: Vs = Ns
Vp Np
kVp
kVp determines the quality of the X-ray beam – the kVp is shown on the console at all times: known as the pre-reading kVp meter
mA
The tube current, measured in milliamperes (mA), flows from cathode to anode.
It represents the number of electrons resulting from thermionic emission.
A step-down transformer is necessary for this milliamperage current.
Fixed mA stations typically include 100mA, 200mA, and 300mA.
Transformers (X-ray Circuit)
Step-Up Transformer (High Voltage Transformer)
Function: Increases voltage (kVp)
Converts volts → kilovolts
Used to accelerate electrons from the cathode → anode
Primary vs Secondary:
The difference is the amplitude (height) of the waveform — higher on the secondary side = higher voltage
Turns Ratio:More turns in the secondary coil than the primary coil
Example: 1:500
Meaning: Voltage increases 500×, current decreases
Step-Down Transformer (Filament Circuit)
Function: Increases current (mA)
Converts amperage → milliamperage (filament heating)
Heats filament → thermionic emission
Turns Ratio:
Fewer turns in the secondary coil than in the primary
Example: 10:1
Meaning: Voltage decreases, current increases
Key Concept: Turns Ratio
Turns Ratio Formula
Ns = Number of turns in the Secondary coil
Np = Number of turns in the Primary coil
What It Means
If Ns > Np → Step-UP transformer
Voltage increases
Current decreases
If Ns < Np → Step-DOWN transformer
Voltage decreases
Current increases
Turn ratio may be 500- 1000
Formula idea (no math needed for ARRT):
Voltage ∝ number of turns
Current is inverse of voltage
Simple memory trick:
Step-up = more turns = ↑ voltage, ↓ current
Step-down = fewer turns = ↓ voltage, ↑ current
Induction Concepts
Mutual Induction
When current flows in the primary coil, it creates a magnetic field
This magnetic field induces a voltage in the secondary coil
This is how transformers work
Self-Induction
A changing current in a coil creates its own opposing magnetic field
This resists changes in current flow
Super simple:
The coil “pushes back” against changes in its own current
Rectification
Converts AC → DC (needed for x-ray production)
Are solid state semiconductor diodes that contain silicon-based n-type and p-type semiconductors.
The current changes direction every 1/60th of a second for alternating
Half-wave: uses only half the power, not conducted during the negative direction of the wave
Full-wave: uses all power (more efficiently), with four diodes, negative direction of the wave is reversed.
Phase, Pulse, Frequency
Single-phase
Voltage Ripple & Efficiency
Single-Phase
100% voltage ripple
Voltage drops to 0 between pulses
Only about 50–60% of the peak voltage is effectively used
Result:
Least efficient
Lower average kVp → less penetrating beam
Three-Phase (6-pulse)
Ripple ~ 13%
Voltage never drops to zero
Uses about ~87% of peak voltage
Key Fact:
Produces ~35% more X-rays than single-phase at the same kVp
Three-Phase (12-pulse)
Ripple ~ 4%
Voltage is almost constant
Uses about ~96% of peak voltage
Key Fact:
Produces ~41% more X-rays than a single-phase
High-Frequency Generators
<1% ripple (almost perfectly constant)
Uses ~99% of peak voltage
Result:
Most efficient, less patient dose.
Highest, most consistent beam quality
Why More X-rays Are Produced
Less ripple = voltage stays closer to peak kVp longer
Electrons hit the anode with higher, more consistent energy
→ More efficient X-ray production
Key Comparison (Test Gold)
Three-phase & high-frequency systems produce an increase of ~12–16% more radiation than single-phase at the same kVp and mAs settings
Because their average voltage is higher
Single-phase generators produce about a 12–16% decrease in radiation output compared to three-phase or high-frequency systems due to greater voltage ripple.
Tube Loading
Refers to the heat capacity of an X-ray tube.
Excess heat = tube damage
99% of energy = heat
Avoid max exposure on a cold anode
Components of the Radiographic Unit
Includes:
Tube
Table
Bucky (grid system)
Support systems (ceiling, C-arm, etc.)
Protective Housing
The metal (lead) exterior housing minimizes isotropic X-ray radiation, prevents electric shock, and dissipates heat. High-capacity tubes use a metal enclosure.
The Pyrex glass enclosure withstands the high heat generated.
The tube maintains a vacuum for efficient X-ray production, known as the Coolidge tube.
X-ray Tube Construction
Cathode (−)
Filament + focusing cup
Produces electrons via thermionic emission
Key concepts:
Small focal spot → better detail
Large focal spot → handles more heat
Space charge effect: electron cloud repulsion at the cathode limits the number of electrons reaching the anode, especially at low kVp.
Anode (+)
Receives electrons → produces x-rays
Types:
Stationary used low power units
Rotating (most common) for high intensity
Functions:
Conduct electricity
Dissipate heat with the material
Produce x-rays
Electron source
Defined as the filament (cathode) made of tungsten
Current (mA) flows through the filament
Filament heats up
Electrons are “boiled off” → thermionic emission
Electrons form a space charge cloud near the filament
mA controls how many electrons → affects X-ray quantity
Forms a space charge cloud that can limit electron flow. The tungsten filament (cathode) is heated by electric current (mA).
Target Material:
Tungsten Atomic number of Tungsten (W) = 74
High atomic number
High melting point
Good heat dissipation
Secondary Materials
Rhenium (Re)
Added to tungsten (≈ 10%)
Prevents cracking and pitting
Increases durability
Molybdenum (Mo)
Used in:
Mammography tubes
Produces lower energy x-rays
Improves soft tissue contrast
Rhodium (Rh)
Also used in mammography
Slightly higher energy than molybdenum
Used for denser breast tissue
Induction Motor
Definition
A device that spins the rotating anode in the X-ray tube
Uses electromagnetic induction (no direct contact)
Components
Stator (outside)
Located outside the glass tube
Contains electromagnets
Creates a rotating magnetic field
Rotor (inside)
Located inside the tube (vacuum)
Made of copper and iron
Connected to the anode disk
Spins when a magnetic field is applied
How It Works (Step-by-Step)
Stator is energized
Produces a rotating magnetic field
The magnetic field interacts with the rotor
The rotor begins to spin
Anode rotates → spreads heat
Purpose
Allows high tube current (mA)
Prevents overheating
Distributes heat over a larger target area
Filtration
Purpose of Filtration
Removes low-energy X-rays
Improves beam quality (hardens beam)
Reduces patient dose
Types of Filtration
1. Inherent Filtration
Definition:
Filtration is built into the X-ray tube.
Includes:
Glass envelope
Insulating oil
Tube window
Amount:
Approximately 0.5 – 1.0 mm Al equivalent
2. Added Filtration
Definition:
Filtration was intentionally added to the beam.
Material:
Aluminum
Location:
Collimator or tube exit
Amount:
Usually 1.5 – 2.0 mm Al
3. Total Filtration
Definition:
Sum of inherent + added filtration
Formula:
Total = Inherent + Added
ARRT Requirement:
Minimum 2.5 mm Al equivalent (for >70 kVp)
4. Compensating Filtration (Extra)
Used to balance uneven anatomy.
Examples:
Wedge
Trough
Boomerang
Automatic Exposure Control (AEC)
Stops exposure when enough radiation reaches the detector.
Controls:
Exposure time only
Technologist sets:
kVp
mA
chamber selection
Requires excellent positioning
Not used with casts or prosthetics
Backup Timer
Stops exposure if AEC fails
Shortest time with an AEC is 1ms
Prevents overexposure
Set at 150% of expected exposure time
mAs Timer
Controls exposure by using mA × time (mAs)
Automatically adjusts time to reach the selected mAs
Ensures accurate total radiation output
Located the secondary coil of a high voltage transformer
Used with falling load generators.
Key idea: You set mAs, and the system figures out the time
Electronic Timer
Uses electronic circuits (no moving parts)
Provides very precise exposure times (milliseconds)
Most common timer in modern X-ray units
Microprocesador controlado.
Allows times of 1 ms
Example
Expected exposure time = 0.2 sec
Backup timer = 0.3 sec (150%)
Exposure Adjustment (Density Control)
Adjusts exposure by ±25–30% per step
+1 → darker
−1 → lighter
Minimum Exposure Controls
Prevents too short an exposure time
Ensures AEC works properly
Manual Exposure Controls
The technologist manually sets:
kVp
mA
time
Used when:
AEC not appropriate
small anatomy
portable exams
Prosthetics/casts
Poor positioning
Manual Exposure Controls – Quick Chart
Radiation Detectors
Used in AEC to measure radiation.
Types:
Radiation Detectors – Condensed Study Guide
1. Gas-Filled Detectors
Contain gas that becomes ionized by radiation
Measure radiation exposure
Types include:
Ionization chambers
Geiger counters (less relevant for imaging)
2. Ionization Chambers (MOST IMPORTANT)
Most common AEC detector
Gas-filled device
X-rays ionize gas → creates an electrical signal
Controls exposure time (AEC)Requires accurate positioning and proper chamber selection
3. Photomultiplier Tubes (PMT)
Used in older AEC systems
Detect light from the fluorescent screenConvert light → electrical signal
Largely replaced by ionization chambers
4. Scintillation Detectors
Convert x-rays → light → electrical signal
Use a scintillator (crystal)
Common in:
CT
Nuclear medicine
5. Semiconductor / Solid-State Detectors
(These are closely related and often grouped)
Definition:
Use solid materials (no gas) to convert radiation → electrical signal
Characteristics:
Fast response
High sensitivity
More efficient than gas detectors
Used in:
Modern AEC systems
Digital radiography (DR)
12. Tube Failure & Heat Management
Causes of Failure:
Excess heat (MOST COMMON)
Bearing damage
Tungsten evaporation
Heat Dissipation:
Radiation
Conduction
Convection
COMPUTED RADIOGRAPHY (CR)
Definition
A cassette-based digital imaging system that uses a reusable imaging plate and a separate plate reader.
Key Components
Imaging Plate (IP)
Contains Photostimulable Phosphor (PSP)
Stores energy from X-ray exposure as a latent image
Photostimulable Phosphor (PSP)
Usually barium fluorohalide (BaFBr: Eu)
Traps electrons when exposed to X-rays
Releases energy as light when stimulated by a laser
Plate Reader
A device that processes the imaging plate
Uses a laser to scan the plate
Converts stored energy into a digital image
System Breakdown
Mechanical
Cassette
Imaging plate
Plate transport system (rollers)
Optical
Laser beam
Mirrors
Light guide
Computer
Photomultiplier tube (light detection)
Analog-to-digital converter (ADC)
Image processing software
PACS storage
Process
X-ray exposure → PSP stores energy
Plate inserted into the reader
Laser stimulates PSP → emits light
Light converted to electrical signal
Signal digitized → image displayed
DIGITAL RADIOGRAPHY (DR)
Definition
A cassette-less digital system that produces an image immediately using a flat panel detector.
DIRECT CONVERSION DR
Key Material
Amorphous Selenium (a-Se)
Function
Converts X-rays directly into an electrical signal
No intermediate light step
System Breakdown
Capture
a-Se absorbs x-rays → electrical charge
Coupling
An electric field moves a charge
Collection
Thin-film transistor (TFT) array collects the signal
Components
Mechanical
Flat panel detector
Optical
None
Computer
TFT electronics
ADC
Image processing system
INDIRECT CONVERSION DR
Key Material
Amorphous Silicon (a-Si)
Function
Converts X-rays into light, then into an electrical signal
System Breakdown
Capture
Scintillator (CsI or Gd₂O₂S) converts x-rays → light
Coupling
a-Si photodiodes convert light → electrical signal
Collection
TFT array stores and reads the signal
Components
Mechanical
Flat panel detector
Optical
Scintillator (light production)
Computer
TFT readout
ADC
Image processing
AMORPHOUS SILICON (a-Si)
Definition
Semiconductor used in indirect DR
Function
Converts light into an electrical signal
Key Point
Always associated with indirect conversion
CHARGE-COUPLED DEVICE (CCD)
Definition
A digital detector that converts light into an electrical signal using a chip
System Breakdown
Capture
Scintillator converts x-rays → light
Coupling
Lens or fiber optics focuses light
Collection
CCD chip converts light → electrical signal
Components
Mechanical
Lens alignment system
Optical
Lens system
Mirrors or fiber optics
Computer
ADC
Image processing electronics
Key Points
Requires lens system
Smaller field of view
Older technology compared to flat panel detectors
COMPLEMENTARY METAL-OXIDE SEMICONDUCTOR (CMOS)
Definition
Modern digital detector similar to CCD but more efficient
System Breakdown
Capture
Scintillator converts x-rays → light
Coupling
Fiber optics or direct coupling transfers light
Collection
CMOS sensor converts light → electrical signal
Each pixel has its own amplifier
Components
Mechanical
Compact detector housing
Optical
Scintillator
Fiber optic coupling (if present)
Computer
On-chip amplification
Fast readout electronics
ADC
Image processing
Key Points
Faster than CCD
Lower power consumption
Less noise
More durable
Used in modern systems
BEAM RESTRICTION
Definition
Beam restriction is the limitation of the X-ray beam to the area of clinical interest.
PURPOSE
1. Reduce Patient Dose
Limits the amount of tissue exposed to radiation
2. Reduce Scatter Radiation
Smaller field size → less scatter produced
3. Improve Image Quality
Less scatter → increased image contrast
METHODS OF BEAM RESTRICTION
1. Collimator (MOST IMPORTANT)
Description
Adjustable lead shutters inside the X-ray tube housing
Produces a rectangular light field
Components
Mechanical
Lead shutters
Adjustment knobs
Optical
Light bulb
Mirror (reflects light to simulate x-ray field)
Computer/Electrical
Light circuit (for field visualization)
Key Functions
Adjust field size
Align light field with x-ray field
Improve positioning accuracy
2. Cones
Description
Cylindrical or conical metal devices
Attach to tube housing
Function
Restrict beam to a circular field
3. Cylinders
Description
Similar to cones but longer and open-ended
Function
Provide more precise restriction
Often used in dental imaging
COLLIMATOR ACCURACY (QUALITY ASSURANCE)
Light Field vs X-ray Field Alignment
Requirement
Light field must match x-ray field within ±2% of SID
Example
At 100 cm SID → allowable error = 2 cm
Why This Matters
Prevents:
Cutting off anatomy (under-collimation)
Unnecessary exposure (over-collimation)
BEAM RESTRICTION AND SCATTER
Relationship
Increased field size → increased scatter
Increased scatter → decreased contrast
Key Concept
Beam restriction:
Decreases Compton scatter reaching the detector
Improves visibility of anatomical detail
PENUMBRA EFFECT
Definition
Blurring at the edges of the image
Beam Restriction Effect
Smaller field size → reduced penumbra → sharper edges
POSITIVE BEAM LIMITATION (PBL)
Definition
Automatic collimation system
Function
Adjusts beam size to match image receptor size
Requirements
Must be accurate within 3% of SID
Limitation
Does not replace proper manual collimation
QUALITY ASSURANCE TESTS
1. Collimator Alignment Test
Checks light field vs x-ray field accuracy
2. PBL Test
Ensures automatic collimation accuracy
3. Beam Centering Test
Ensures central ray aligns with image receptor
FLUOROSCOPIC UNIT
Components of a Fluoroscopic Unit
Fixed Fluoroscopy Unit
Mechanical
X-ray tube (under or over table)
Fluoroscopic table (tilting or stationary)
Image receptor (image intensifier or flat panel)
Tower (supports IR)
Compression device
Electrical
Generator
Control panel
Automatic Brightness Control (ABC/AERC)
Computer
Digital processing system
Image storage (PACS)
Post-processing workstation
Mobile Fluoroscopy Unit (C-Arm)
Mechanical
C-arm structure (holds tube and IR)
X-ray tube
Image receptor (II or flat panel)
Electrical
Generator (smaller, less powerful than fixed)
Computer
Monitor display
Image storage system
Limited post-processing
Key Details: it is better to have the tube below the patient to hit the bottom of aprons like resistant gonads than the tube above to avoid unnecessary exposure to sensitive areas like eyes and thyroid
IMAGE RECEPTORS (IR)
1. Image Intensifier (II)
Purpose
Converts low-intensity x-ray image into a brighter visible image
Components of Image Intensifier
Input Phosphor
Cesium iodide (CsI)
Converts x-rays → light
Photocathode
Converts light → electrons
Electrostatic Focusing Lenses
Focus electron beam toward output phosphor
Accelerating Anode
Speeds up electrons
Output Phosphor
Converts electrons → visible light
Key Concepts
Minification Gain
Smaller output phosphor compared to input phosphor
Increases brightness
Formula:
Minification Gain = (Input diameter² / Output diameter²)
Flux Gain
Acceleration of electrons increases brightness
Total Brightness Gain
Total Brightness Gain = Minification Gain × Flux Gain
Typical range: 5,000 – 20,000
Decreases as the tube ages
Conversion Factor
Efficiency of II
Ratio of output brightness / input exposure
2. Flat Panel Detector (Digital Fluoroscopy)
Advantages
Higher spatial resolution (better small detail)
Larger field of view (some systems)
No geometric distortion
Improved post-processing
More uniform image quality
Replacement
Replaces the image intensifier in modern systems
VIEWING AND RECORDING SYSTEMS
Image Distribution
Image is sent from the image receptor to:
Monitor (TV or LED display)
Recording systems
Viewing Systems
Television monitor (older systems)
LED/LCD monitors (modern systems)
Recording Systems
Digital storage (PACS)
Hard drive storage
Video recording (older systems)
Digital Fluoroscopy
Similar x-ray equipment as conventional fluoroscopy
Includes:
Operating console
Computer workstation
Allows:
Image storage
Image processing
Frame capture
AUTOMATIC BRIGHTNESS CONTROL (ABC / AERC)
Definition
Automatically adjusts exposure to maintain constant image brightness
How It Works
Adjusts:
kVp
mA
Pulse rate (in pulsed fluoro)
Purpose
Maintain consistent image brightness
Compensate for patient thickness/density
MAGNIFICATION MODE
Requirements
Requires dual-focus or trifocus x-ray tube
How It Works
Uses a smaller portion of the input phosphor
Changes voltage in image intensifier
Effects
Advantages
Increased spatial resolution
Disadvantages
Increased patient dose (very testable)
Decreased brightness (ABC compensates by increasing exposure)
FLUOROSCOPIC UNIT TABLE
Types of Fluoroscopy Tables
TABLE COMPONENTS
Mechanical Components
X-ray Tube Position
IMAGE RECEPTOR MOVEMENT
TABLE OPERATION FEATURES
Controls
Safety Features
PATIENT POSITIONING FEATURES
IMAGE QUALITY + DOSE RELATION
3. COUNTERWEIGHTS (STATIONARY FLUORO)
Purpose
Balance x-ray tube and image receptor
Prevent drifting or “floating”
Allow smooth controlled movement
Key Points
Located inside support system
Improve positioning accuracy
Reduce mechanical strain
Lead Curtain
Bucky Slot Cover
FLUOROSCOPY ACCESSORIES
1. STATIONARY GRIDS
Definition
A grid placed between the patient and image receptor to reduce scatter radiation and improve image contrast.
Structure
Made of lead strips separated by radiolucent material (aluminum or plastic)
Strips are aligned to allow primary beam through while absorbing scatter
Key Characteristics
Function
Absorbs scatter radiation
Improves image contrast
Disadvantages
Increases patient dose (more exposure needed)
Can produce grid lines (less noticeable in fluoro)
2. BUCKY ASSEMBLY
Definition
A device that holds the grid and image receptor and allows proper positioning during imaging.
Components
Function
Aligns grid with x-ray beam
Holds image receptor securely
Allows movement of IR during procedures
In Fluoroscopy
Usually integrated into the table or tower
May include stationary grid (not moving like radiographic Bucky)
High-Yield Points
Bucky assembly includes grid + IR holder
Bucky slot cover reduces operator exposure (~90%)
3. COMPENSATING FILTERS
Definition
Devices placed in the x-ray beam to even out differences in patient thickness/density.
Purpose
Produce more uniform image exposure
Prevent overexposure of thinner areas
Improve contrast
Types
Function
Absorbs more x-rays in thinner areas
Allows more x-rays through thicker areas
Benefits
Improves image quality
Reduces patient dose in thinner regions
Image Processing and Display
1. Raw Data (Pre-processing)
This is the initial signal captured by the detector.
Still unprocessed → contains noise, inconsistencies, and detector imperfections.
2. Analog-to-Digital Converter (ADC)
Converts the signal from analog (continuous) → digital (discrete numbers).
Each pixel gets a numerical value representing intensity.
3. Quantization
Assigns each pixel a discrete value (bit depth).
Example:
8-bit = 256 shades of gray
12-bit = 4096 shades (more detail!)
More bits = better contrast resolution
4. Corrections
Fixes imperfections in the raw image:
Rescaling → adjusts pixel values to usable range
Flat fielding → corrects uneven detector response
Dead pixel correction → replaces non-functioning pixels
Goal: make the image accurate and uniform
5. Histogram
A graph of pixel intensity distribution
Shows how many pixels fall into each brightness level
Used to:
Evaluate exposure
Guide processing decisions
6. Corrected Data for Processing
Now the image is clean and standardized
Ready for enhancement algorithms
7. Image Processing (Enhancement Techniques)
Grayscale
Assigns pixel values to shades of gray
Determines overall image brightness/contrast
Edge Enhancement
Sharpens borders between structures
Makes anatomy easier to distinguish
Example: bone edges appear crisper
Equalization
Redistributes grayscale values
Improves contrast in areas with poor visibility
Smoothing
Reduces noise (graininess)
Makes image look cleaner—but can slightly blur detail
8. Data for Display
Final processed data ready to be shown on the monitor
9. Values of Interest (VOI)
Selects the range of pixel values you want to emphasize
This is where:
Window level (brightness)
Window width (contrast)
are applied
10. Look-Up Table (LUT)
Maps pixel values → display brightness
Determines:
How the image actually appears on the screen
Can invert or adjust contrast
Post-Processing
Happens AFTER the image is acquired and processed
Does NOT change the raw data, only how it’s displayed
Brightness
Controls overall lightness/darkness of the image
Adjusted using window level (WL)
Higher brightness = lighter image
Contrast
Controls the difference between light and dark areas
Adjusted using window width (WW)
Narrow WW = high contrast (more black & white)
Wide WW = low contrast (more grays)
Region of Interest (ROI)
Selected specific area of the image for analysis or processing
Used for:
Exposure evaluation
Histogram analysis
Measurements
Incorrect ROI selection = wrong brightness/contrast
Electronic Cropping (Masking)
Hides unwanted areas of the image
Important:
Does NOT remove data, only covers it
Such as the white border from cropping, but should not be used to hide uncollimated anatomy.
The original image can still be retrieved
Used to:
Improve appearance
Reduce distractions
Stitching
Combines multiple images into one long image
Common uses:
Full spine imaging
Long bone studies
5. Display Monitors
Final step—how the image is actually seen by the radiologist
Viewing Conditions
Viewing Angle
Image quality decreases if viewed off-angle
Best viewed straight on
Ambient Lighting
Room lighting affects visibility
Ideal:
Low, controlled lighting
Too bright = reduces contrast perception
Spatial Resolution (Monitor Quality)
Pixel Size
Size of each pixel on the monitor
Smaller pixels = sharper image
Pixel Pitch
Distance between pixels
Smaller pitch = higher resolution
Brightness and Contrast (Monitor Performance)
Monitors must display a wide range of grayscale values
High-quality monitors:
Better contrast resolution
More accurate diagnosis
6. IMAGING INFORMATICS
1. INFORMATION SYSTEMS
HIS (Hospital Information System)
Definition
The main system that manages the entire hospital operations
Includes
Patient registration
Billing
Scheduling
Medical records
RIS (Radiology Information System)
Definition
Specialized system for radiology workflow
Functions
Exam scheduling
Tracking imaging orders
Reporting (radiologist reports)
Patient tracking within radiology
EMR (Electronic Medical Record)
Definition
Digital version of a patient’s chart within one facility
Key Point
Not easily shared outside that facility
EHR (Electronic Health Record)
Definition
Comprehensive patient record that can be shared across multiple facilities
Includes
History
Imaging
Lab results
Medications
2. NETWORKING
Definition
A system that connects computers and devices to share data
Key Components
Types
Key Concepts
Enables image transfer and access
Supports PACS and RIS integration
3. PACS (Picture Archiving and Communication System)
Definition
System used to store, retrieve, and distribute images
Components
Functions
Image storage
Image retrieval
Image display
Image sharing
Advantages
Eliminates film
Faster access
Remote viewing
4. DICOM (Digital Imaging and Communications in Medicine)
Definition
Standard format for storing and transmitting medical images
Purpose
Ensures compatibility between:
Modalities
PACS
Workstations
Key Features
Standard file format
Includes:
Patient data
Image data
Enables communication between different systems
5. DOWNTIME PROCEDURES
Definition
Backup procedures used when computer systems fail
When Used
PACS failure
Network outage
System maintenance
Procedures
Use paper requisitions
Manual patient identification
Temporary image storage (local system)
Document exams manually
Enter data into system once restored
Key Goals
Maintain workflow
Prevent data loss
Ensure patient safety
BEAM RESTRICTION
Definition
Limiting the x-ray beam to the area of interest using collimation.
Purpose
Reduce patient dose
Reduce scatter radiation
Improve image contrast
Quality Control Check
Field size should match area of interest
Collimator shutters must function properly
Light field must accurately represent x-ray field
Key Relationship
Smaller field size → less scatter → higher contrast
LIGHT FIELD TO RADIATION FIELD ALIGNMENT
Definition
Ensures the visible light field matches the actual x-ray beam field.
Why It Matters
Prevents:
Cutting off anatomy (under-collimation)
Excess exposure (over-collimation)
Reduces repeat exams
Alignment Tolerance
Must be within ±2% of SID
Example
SID = 100 cm
2% = 2 cm
→ Light field and x-ray field can differ by no more than 2 cm
How It Is Tested
Place test tool (collimator test plate or markers)
Expose image
Compare light field edges to radiation field edges
3. CENTRAL RAY ALIGNMENT
Definition
Ensures the central ray (CR) is aligned with the center of the image receptor (IR).
Purpose
Prevents image distortion
Ensures proper centering
Avoids anatomy cutoff
Tolerance (VERY TESTABLE)
Must be within ±1% of SID
Example
SID = 100 cm
1% = 1 cm
→ CR must be within 1 cm of IR center
How It Is Tested
Use alignment test tool
Check if CR hits exact center of IR
SUMMARY TABLE
RECOGNITION AND REPORTING OF MALFUNCTIONS
Purpose
Maintain image quality
Ensure patient safety
Prevent repeat exposures
Common Signs of Malfunction
Technologist Responsibilities
Recognize abnormal performance
Stop using unsafe equipment
Report immediately (supervisor/biomedical)
Document issue per protocol
DIGITAL IMAGING RECEPTOR SYSTEM QC
Systems
CR (PSP plates)
DR (flat panel detectors)
Common Problems
QC Procedures
Detector calibration (flat-field correction)
Uniformity checks
Artifact evaluation
Exposure indicator accuracy
MAINTENANCE
Erasure Thoroughness (CR)
Ensures no residual image remains
Poor erasure → ghost artifacts
Plate Uniformity (CR)
Ensures consistent response across plate
Issues → shading or uneven brightness
Spatial Resolution
Ability to see fine detail
Tested with line pair patterns
DISPLAY MONITOR QUALITY ASSURANCE
GSDF (Grayscale Standard Display Function)
Ensures consistent grayscale display
Based on DICOM standard
Equal pixel changes → equal perceived brightness
Luminance
Brightness of monitor
Measured in cd/m²
Typical range: 50–300 cd/m²
Illuminance
Ambient light
Measured in lux
Measurement Tools
Display QC Tests (AAPM TG-18)
Common Monitor Issues
DISPLAY PERFORMANCE STANDARDS
SMPTE
Tests contrast and resolution
5% and 95% patches must be visible
DICOM (GSDF)
Standardizes grayscale display
Ensures consistency across systems
AAPM TG-18
Primary QC standard
Provides test patterns and procedures
Other Standards
SHIELDING ACCESSORIES QC
Includes
Lead aprons
Thyroid shields
Lead gloves
Purpose
Protect from radiation exposure
Testing Methods
Visual Inspection
Check for cracks, tears, damage
Radiographic/Fluoroscopic Testing
Detect internal cracks
Frequency
Typically annually
Common Problems
Action
Remove defective shielding immediately
QUALITY CONTROL PROGRAM STRUCTURE
Required Components
Acceptance testing (new equipment)
Routine QC (technologist)
Periodic review (physicist)
Annual evaluations
Post-repair testing
RESPONSIBILITIES
VII. Organizational Factors Affecting Medical Image Quality
A. IR (Image Receptor) Factors
Pixel size, dynamic range, intensity response, signal-to-noise ratio, and post-processing.
B. Geometric Factors
Distortion, magnification, and blur.
C. Subject Factors
Contrast (influenced by thickness, density, and atomic number) and motion.
SUMMARY
Ghost image = incomplete CR erasure
Dead pixels = DR detector issue
GSDF = consistent grayscale display
Luminance = brightness (cd/m²)
Illuminance = ambient light (lux)
TG-18 = primary QC testing system
SMPTE = 5% and 95% visibility test
Cracked shielding = remove immediately
The technologist reports malfunctions immediately