The KVP is held constant for a given range of subject densities and contrast while the mass
is varied to achieve the appropriate image density describes
fixed KVP systems
advantages of fixed KVP systems include
decrease patient dose, increases tube life, consistent contrast levels
The maximum KVP level that will produce images with appropriate contrast that are consistently within acceptable limits describes
optimal KVP
what is the optimal KVP range for skull work
85 kvp
The systems advantage is that it permits small incremental changes an exposure for the body part being examined
variable KVP systems
from earlier days, it was common to change KVP and increments of ___ kvp for every ___ cm
2,1
most medical facilities use ___ kvp over ___ kvp charts
APR stands for
Anatomically Programmed Radiography
which system eliminates the need to set an exposure time
AEC
what does OD stand for
optical density
a density factor of +1 is equivalent to ___ in relative exposure
25% increase
when using AEC, what results in an increased repeat rate
Poor positioning skills
grayscale bit depth contains the number of shades of a gray allowable and one pixel. The equipment can have bits that range from ___ bits - ___ bits
8-32
if the field of view (FOV) remains the same; then as the matrix size increases, the pixels get ___, and the image becomes ___
smaller, sharper
What's a Instagram has been determined, a histogram analysis is been performed within the computer to determine what two things
values of interest and exposure indicator
A wide histogram will have what type of contract
high
The LUT is applied to the image after the histogram is made in the histogram analysis is performed. What does this do to the image
adjusts contrast
if an image is showing only few densities on the monitor, it is said to have high contrast. What type of window width will have high contrast
narrow
which spacial frequency filter removes or suppresses low spatial frequency from the image, resulting in an increase of contrast and producing a sharper image
High-pass Filtering/Edge Enhancement
an increase in signal to noise ratio, or a higher SNR indicates what
little noise
what is the only way to decrease quantum mottle noise
repeating the image with increased technical factors
DR systems have DQE's of ___ - ___ % efficiency
30-70
A device with 100% efficiency of converting incoming data to the output viewing device has a DQE of ___ meaning there is ___ loss of information
1, no
The comparison between the actual exposure and the proper exposure received by the image detector is known as
deviation index
The exposure sensitivity of an imaging detector ranges from
.1 mR - 100 mR
Fuji, Konica and Phillips manufactures have a sensitivity number or 's' value which demonstrates the exposure to the imaging plate. What type of relationship does the S value have in regards to exposure
as exposure increases, the S value decreases, inverse
which of the detector configurations creates the image without a scintillator
direct conversion
what is the active layer of the PSP plate in computed radiography
phosphor layer
in CR, when does the photo stimulated luminescence or PSL occur
I a peso pitch she gets smaller, the ___ the special resolution
greater
once the Layton image is captured by the PSP play, it is fed through the processor and scanned line by line in what pattern
raster pattern
in CR, pixel pitch is determined by the size of what
sampling or Nyquist frequency
A grossly overexposed plate may not erase completely. What artifact results from this on the next image
ghosting artifact
The latent image in computed radiography is formed where
europium phosphor sites
The late imaged on a CR plate will lose about how much of its energy and how many hours
25%, 8 hours
what is the most common type of CR phosphorus
barium fluorohalide bromide with europium activators & barium fluorohalide iodide with europium activators
in PSP technology, what does a photo detector do
converts light into electronic signal
The process that will determine the brightness levels or grayscale for the pixels is
quantification
segmentation (through carestream manufacturers) is also known as what?
exposure recognition
causes of histogram analysis air include
Poor collimation, increased or decreased attenuation located in the body where it's not normally located, too much collimation and prosthetic devices included or other parts with large density differences
in DR, the pixel pitch is determined by the size of what
detector element
each image will generate its own specific ___ using the image data from within the collimated edges. This allows the system to find the useful signal
histogram
grid cut off and digital radiography will have what type of appearance
Poor spatial resolution, poor density or brightness, poor contrast
and certain detectors of DR, this is the last step of the x-ray conversion. This collects electric charges to produce an electric readout. This information is collected in a type of matrix known as a ___ containing ___
TFT array (or active-matrix array) containing detector elements (DELs
Direct conversion detectors use
Amorphous Selenium
which detector uses this conversion process? X-ray strike the scintillator, then the light is optically coupled to special sensors like the lenses in optic fibers. then the light can be converted into an electric charge and sent to the computer to be displayed
indirect converison
which digital artifact is caused by dust or foreign objects on the IP
white spots
And digital radiography, and underexposed image will show ___ even though densities adequate. An image that is overexpose will tend to suffer from ___ contrast
quantum mottle, low
A scintillator can be made using which types of material