1/66
Looks like no tags are added yet.
Name | Mastery | Learn | Test | Matching | Spaced | Call with Kai |
|---|
No analytics yet
Send a link to your students to track their progress
CT cooling system
maintains stable temperature of gantry components.
Cooling mechanisms
include blowers, filters, and oil-to-air heat exchangers.
Cooling importance
prevents damage from temperature fluctuations.
CT x-ray tube
requires high power capacity for imaging.
Rotating anode
used in most CT tubes for heat dissipation.
Anode heat capacity
measured in MHU or Mega Heat Units.
2.1 MHU
most CT scanner anode heating capacity
4-5 MHU
advanced CT scanner
about 0.5 MHU or 500 KHU.
Minimum anode heat capacity
8 MHU
Maximum anode heat capacity
X-ray tube failure
main cause of CT scanner malfunction and principal limitation on sequential scanning frequency
Small focal spot
used to achieve high spatial resolution.
Collimation
controls slice thickness by shaping the x-ray beam.
Source collimator
located before the patient to shape the primary beam.
Source collimator function
controls patient dose and beam width entering the patient.
Predetector collimator
located after the patient before detectors.
Predetector function
reduces scatter and ensures proper beam width.
Detector array
collection of detectors arranged in arc or ring.measures the intensity of transmitted x-ray radiation along a beam projected from the x-ray source to that particular detector element.
Reference detectors
used for calibration and artifact reduction.
SFOV
determines fan beam size and number of detectors.
Detector characteristics
include high efficiency, low afterglow, and stability.
Detector assembly
measures the radiation exiting the patient and feeds back the information (primary data) to the host computer.
CT detector requirements
high efficiency, fast response, wide dynamic range.
Capture efficiency
ability to receive photons from the patient.
Absorption efficiency
ability to absorb incoming x-ray photons.
Conversion efficiency
ability to convert photons into digital signals.
Xenon gas detector
uses pressurized gas to detect x-rays.
Xenon efficiency
absorbs about 60–87% of photons.
Xenon advantage
stable and less expensive.
Xenon disadvantage
requires pressurized aluminum casing and has lower efficiency.
Solid-state detector
uses scintillation crystals to detect x-rays.
Solid-state efficiency
absorbs nearly 100% of photons.
Solid-state advantage
high absorption efficiency.
Solid-state limitation
sensitive to temperature and moisture.
CT computer
processes raw data and reconstructs images.
Reconstruction time
depends on processor and memory speed.
Array processor
instead of a microprocessor because this allows image reconstruction in less than 1 second.
Data acquisition
process of scanning the patient.
Operating console
controls scan parameters and data entry.
CT technique factors
include kVp, mA, scan time, and slice thickness.
Typical scan settings
around 120 kVp, 100 mA, and 1–5 seconds.
Normal Slice thickness
ranges from 1 to 10 mm.
0.5 mm thickness
for high resolution
Scan monitor
used for scanning and entering patient data.
Display monitor
used for viewing and post-processing images.
CT spatial resolution
ability to define small objects distinctly.
CT contrast resolution
ability to differentiate, on the image, objects with similar densities
CT temporal resolution
speed of data acquisition.
Ray sum
The total x-ray transmission measured by each detector is the result of the sum of the attenuation by all the tissues the beam has passed through
Image reconstruction
uses algorithms to form cross-sectional images.
Hounsfield unit
measures tissue attenuation values.
Interpolation
estimation of missing data between measured points.
Extrapolation
estimation of data outside measured range.
CT principle of operation
A mathematical algorithm takes the multiple projection data (raw data) and reconstructs the cross-sectional CT image
Filtered back projection
main reconstruction algorithm in CT.
FBP function
reduces blurring from simple back projection.
Filter function
adjusts balance between resolution and noise.
Soft tissue filters
provide some smoothing, which decreases image noise but also decreases spatial resolution.
Parallel beam geometry
all of the rays in a projection are parallel to each other.
Fan beam geometry
the rays at a given projection angle diverge and have the appearance of a fan.
Cone beam geometry
uses cone-shaped beams for volumetric imaging.
Step-and-shoot scanning
acquires one slice at a time with gantry stopping.
Axial scanning, conventional scanning, or serial scanning
another term for step-and-shoot scanning.
Helical scanning
continuous rotation with simultaneous table movement.
Helical path
spiral trajectory of tube around patient.
Multidetector CT
uses multiple detector rows for faster imaging.
MDCT advantage
acquires multiple slices per rotation.