Crooked Ribbon Causes

-edges of block not parallel to blade

-block is unevenly chilled

-imperfections in blade

2
New cards

Crooked Ribbon Resolutions

-make block parallel to blade

-ensure even chilling

-change blade

3
New cards

Uneven Sections Causes

-block holder not aligned to blade
-loose block angle lever
-improper embedding

4
New cards

Uneven Sections Resolutions

-adjust block holder
-use microtome calibration tools
-tighten levers
-re-embed

5
New cards

Holes in Section Causes

-aggressive sectioning/facing
-tissue is under-hydrated
-tissue is under decalcified
-unexpected calcifications, staples, or sutures

6
New cards

Holes in Section Resolution

-slower speed
-face blocks when warm
-properly hydrate
-recognize natural holes in section

7
New cards

Ribbon will not Form/Lifting Sections Causes

-dull blade/paraffin buildup behind blade
-paraffin too hard/sticky
-room /block too warm
-thick sections
-incorrect clearance angle
-loose parts
-static

8
New cards

Ribbon will not Form/Lifting Sections Resolutions

-replace blade/remove buildup
-use different wax
-adjust room temp
-chill block
-decrease microns
-change clearance angle
-tighten parts
-score block
-scroll method

9
New cards

Water Trapped Under Section Cause

scooping section when picking up

10
New cards

Water Trapped Under Section Resolution

use pulling method (angle slide at 45 degrees to pick up tissue than rotate it to 90 degrees to remove from bath)

11
New cards

Washboarding/Undulation/Venetian Blind Causes

-worn/loose parts
-blade/block not clamped correctly
-aggressive sectioning
-block holder overextended
-clearance angle too high
-underdecalcified tissue
-dense/fibrous tissue

12
New cards

Washboarding/Undulation/Venetian Blind Resolutions

-tighten parts/clamp sufficiently
-retract holder
-decrease clearance angle
-surface decal
-soften tissue (fabric softener/dish soap)

13
New cards

Chatter Causes

-underhydrated tissue
-clearance angle too high
-aggressive sectioning
-dull blade

14
New cards

Chatter Resolutions

-longer hydration
-decrease dehydration time on processor
-if tissue is overprocessed: float on bath, and add alcohol to block
-decrease clearance angle
-slower speed

15
New cards

Thick and Thin Causes

-block not secure
-blade not tight
-clearance angle too low
-loose/worn parts
-aggressive sectioning
-overextended chuck

16
New cards

Thick and Thin Resolutions

-increase clearance angle
-tighten parts
-slow speed
-retract chuck

17
New cards

Compression Causes

-dirty/dull blade
-clearance angle too low
-sectioning too rapidly
-room/block too warm
-underhydrated tissue

18
New cards

Compression Resolutions

-change blade
-increase angle
-slower speed
-chill block/hydrate

19
New cards

Knife Lines Causes

-blade defect
-underdecalcified tissue
-staples, sutures, or foreign particles present

20
New cards

Knife Lines Resolutions

-replace blade
-reposition block in chuck holder
-surface decal
-melt block to remove debris

21
New cards

Fragmented, Separated, or Torn Sections Causes

-incomplete fixation, processing
-over manipulation on bath
-bath too warm
-extreme overhydration

22
New cards

Fragmented, Separated, or Torn Sections Resolutions

-reprocess
-super chill block
-add freeze spray, alcohol, or detergent to bath
-decrease bath temp

23
New cards

Blade Gouges out of Block Causes

-improper embedding (air bubbles)
-knife/parts not clamped
-block not secure

24
New cards

Blade Gouges out of Block Resolutions

-re-embed
-tighten parts
-re-secure block

25
New cards

Cracking/Parched Earth Causes

-under-hydrated tissue (blood)
-bath too warm
-overuse of freeze spray
-over manipulation on bath

26
New cards

Cracking/Parched Earth Resolutions

-soak tissue
-ensure proper hydration
-decrease bath temp

27
New cards

Folds/Wrinkles Causes

-bath too cold
-not allowing section to flatten on bath
-section not adhered to slide
-slide not properly dried

28
New cards

Wavy Section Causes

-poor adhesion to slide

-incomplete drying before staining

29
New cards

Wavy Sections Resolutions

-use charges slides
-completely dry before staining

30
New cards

Floaters/Contamination/Debris Causes

-dirty bath
-squames
-debris in sections
-ppt from staining solutions

31
New cards

Floaters/ Contamination/ Debris Resolutions

-use a kimwipe to clean bath
-use lotion
-filter staining solutions

32
New cards

Nuclear Bubbling/Distortion/Cell Shrinkage Causes

-excessive heat

-under fixed tissue

33
New cards

Nuclear Bubbling/Distortion/Cell Shrinkage Resolutions

-monitor oven temp

-ensure proper fixation

34
New cards

Cracks in Paraffin Block Causes

-overuse of freeze spray
-block chilled too fast
-block left on embedding center cold plate too long
-dirty molds

35
New cards

Cracks in Paraffin Block Resolutions

-avoid freeze spray

-re-embed if necessary

36
New cards

Tissue has a White Appearance Causes

-excessive hydration

-underprocessed tissue

37
New cards

Tissue has a White Appearance Resolutions

-allow tissue to dry
-limit soaking time
-re-process

38
New cards

Soft Mushy Tissue Causes

-tissue grossed too thick
-inadequate processing
-saturated/contaminated processing reagents
-inadequate fixation

39
New cards

Soft Mushy Tissue Resolution

-re-process/re-embed

40
New cards

Soft Mushy Tissue Prevention

-gross thinner sections

-change processor reagents

41
New cards

Paraffin Cracking Causes

-cold plate too cold (wax hardens too quickly)
-molds are dirty (may cause wax to stick)
-block removed from mold too soon

42
New cards

Paraffin Cracking Prevention

-decrease cold plate temp or place paper towel on surface if temp cannot be changed
-clean molds
-do not remove blocks from mold until they are sufficiently chilled