A received frequency that is less than the transmitted frequency is which of the following. (multiple answers)

Negative Doppler Shift, Flow Away from the Transducer, Retrograde Flow

2
New cards

The major reflector for Doppler ultrasound is which of the following?

Red Blood Cells

3
New cards

The Doppler shift is dependent on all of the following EXCEPT?

thickness of the vessel wall

4
New cards

The cosine of 180 is which of the following?

-1

5
New cards

In regard to Doppler ultrasound, the extracting of lower Doppler frequencies from higher transducer frequencies is called?

demodulation

6
New cards

Which of the following information is recorded by spectral Doppler?

velocity of the reflector

7
New cards

Inaccurate angle correction may result in spurious velocity determinations which may lead to a misdiagnosis.

true

8
New cards

Velocity refers to:

both speed and direction of red blood cells

9
New cards

This form of Doppler is used in a clinical setting simply to identify the presence of flow when direction of flow is of no consequence.

nondirectional Doppler

10
New cards

Simultaneous imaging and Doppler is termed?

Duplex imaging

11
New cards

When discussing "laminar flow" within a vessel, blood velocities are:

highest in the center and decrease to zero at the walls

12
New cards

Which of the following may be present in pulsed Doppler but NOT in continuous wave Doppler?

aliasing

13
New cards

The purpose of the FFT, fast Fourier transform program is to?

display the Doppler frequency spectrum

14
New cards

Aliasing in pulsed doppler occurs when:(May be more than one answer)

the Doppler frequency shift exceeds 1/2 the PRF, the sampling rate is less than 2 times the frequency of the signal.

15
New cards

A patient who has anemia should not have a Doppler exam performed because of the low concentration of red blood cells.

False

16
New cards

The strength of the Doppler signal from flowing blood is most closely related to:

the number of red blood cells in the beam

17
New cards

Absence of color on a color Doppler image always indicates no flow is present.

False

18
New cards

On color Doppler image, red always indicates flow toward the transducer and blue always indicates flow away from the transducer.

False

19
New cards

Color Doppler reports peak velocities.

False

20
New cards

Power Doppler can identify direction of flow but not velocities of flow.

False

21
New cards

The Doppler shift produces information about which of the following?

velocity

22
New cards

What is the current method of processing spectral Doppler signals?

fast Fourier transform

23
New cards

What is the current method for analyzing color Doppler information?

autocorrelation

24
New cards

When high velocities appear negative on the spectral tracing it is termed?

aliasing

25
New cards

All of the following are goals of quality assurance EXCEPT:


increase repeat scans

26
New cards

The AIUM 100 mm test object can evaluate grayscale.

False

27
New cards

Which of the following evaluates mock cysts and solid masses?

tissue equivalent phantom

28
New cards

Which of the following describes the system's ability to display similar reflectors in the phantom with echoes of equal brightness.

uniformity

29
New cards

Which of the following types of resolution displays the smallest distance at which two pins parallel to the sound beam and are displayed as two distinct echoes?

Axial

30
New cards

A region between the transducer to a depth of meaningful reflections is called which of the following


dead zone

31
New cards

Which one of the following is a requirement for any lab?

record keeping

32
New cards

All of the following can be used to evaluate Doppler systems EXCEPT:

tissue equivalent phantom

33
New cards

Which of the following might a Doppler phantom may assess?

All the above

34
New cards

Quality assurance can only be performed by engineers or physical scientists.

False

35
New cards

Quality assurance should be performed:

routinely and periodically

36
New cards

The science of identifying and measuring the characteristics of an ultrasound beam that are relevant to its potential for producing bioeffects is which of the following?

dosimetry

37
New cards

Which of the following is related to the likelihood of harmful bioeffects from cavitation?

mechanical index

38
New cards

Research performed within the living body is known as:

in Vivo

39
New cards

Which of the following is the thermal index of soft tissue?

TIS

40
New cards

A high mechanical index will cause:

more cavitation

41
New cards

It is appropriate to use diagnostic ultrasound in a non medical setting for entertainment.

False

42
New cards

Which of the following is generally true of diagnostic ultrasound?

harmful bioeffects do not occur but it would be irresponsible to ignore the possibility they may in the future

43
New cards

Which of the following is correct for an unfocused beam?

100 mW/cm sq

44
New cards

Which of the following is the correct intensity for a focused beam?

1 W/cm sq

45
New cards

No confirmed bioeffects on patients or sonographers have been found with the use of diagnostic ultrasound.

True

46
New cards

Which of the following is the unit for impedance?

rayls

47
New cards

Which of the following is the unit for pressure?

Pa

48
New cards

Which of the following is the processor used to analyze spectral Doppler?

fast Fourier transform

49
New cards

Which one of the following will give the most accurate spectral Doppler information?

180 degrees

50
New cards

Adding more focal zones will do which of the following?

decrease temporal resolution

51
New cards

You are using spectral Doppler to evaluate the heart pressure and need to know exactly where the velocities are coming from. In this case you would use which of the following?

PW Doppler

52
New cards

Adding too much filter or decreasing the filter too much will lead to a loss of information on the spectral Doppler waveform.

True

53
New cards

Power Doppler can determine the direction of flow but not the speed of blood flow.

False

54
New cards

Which of the following is the half value layer thickness of a 7 MHz transducer?

0.85 cm

55
New cards

What is the attenuation in dB using a 2.5 MHz transducer and imageing to a depth of 16 cm?

20 dB

56
New cards

What is the period of a 10MHz transducer?

0.1 microseconds