How does the presence of contrast media in a region of interest affect AEC operation during a radiographic procedure?
Increased exposure time
What is a consequence of improperly calibrated AEC sensors during a pelvis radiograph?
Over or underestimation of required exposure
Which AEC detector should be activated for a right shoulder xray?
Center AEC to align with right shoulder
Which of the following factors can lead to AEC mis operation resulting in an underexposed image?
Incorrect positioning of the body part over AEC detector
Incorrect positioning of the body part relative to the detector can cause the AEC to terminate the exposure prematurely, leading to a _ image.
Underexposed
When would a recalibration of the AEC system be required?
After replacing the xray tube
A quality control test of the consistency of exposure across different patient sizes will assess the function of what system?
AEC
What is the best practice when using AEC for a PA chest radiograph on a patient with severe kyphosis?
Adjust selection of AEC detectors to better align with the anatomical changes
What consideration should be utilized when using AEC for a morbidly obese patient?
Increase AEC density/sensitivity
What impact does activating the AEC for a lateral lumbar spine radiograph have on image quality compared to manual settings?
AEC can achieve more consistent images
Which two of the following are ways a radiologic technologist can adjust the AEC settings for pediatric patients to avoid overexposure?
Decrease the sensitivity of AEC detectors
Increase the sensitivity of AEC detectors
Extend the backup time
Utilize a lower energy xray beam
Increase the sensitivity of AEC detectors
Utilize a lower energy xray beam
How does increasing the kVp settings affect the exposure time when using AEC?
Decreases exposure time
What would be the best approach to configure the AEC for a patient undergoing a chest xray who has a pneumothorax?
Decrease sensitivity of the affected side’s AEC detector
How does the AEC system adjust when imaging thicker body parts?
Increases exposure time
In situations where a prosthetic device is present, which two options would a technologist have when using AEC?
Disable AEC and manually set exposure parameters
Position the device over the AEC detectors
Avoid positioning the device directly over the AEC detectors
Increase the sensitivity of the AEC detectors
Disable AEC and manually set exposure parameters
Avoid positioning the device directly over the AEC detectors