The main circuit of an x-ray machine is composed of what
controle counsole, x-ray tube, low voltage & high voltage
the purpose of the tube circuit is to
supply the x-ray tube with potential difference
which is not part of the control console circuit?
line monitor
rheostat
ma selector
kvp selector
all
when setting the ma station you adjust what flowing through the filament circuit?
amps
all the controls that a radiographer control are located on the
low voltage side of the main circuit
- most be about 5 mm and radiolucent
- the x-rays ionize the gas in the chamber and liberate the electrons which cause a current
- when the predetermined amount of current is produced, the exposure is terminated
ionization chamber (AEC)
the filament circuit, the incoming voltage is
stepped down
which device is not typically found in the high-voltage section of the x-ray machine
rheostat
the milliamperage in the x-ray circuit is controlled by what device
rheostat
in the x-ray console the exposure switch is placed
between the auto transformer and timer circuit
the autotransformer in the x-ray circuit controls which technical factor?
kvp
on a 3 phase, 12 pulse generator, the photon energy is increased by how much compared to a single phase generator
40%
the incoming voltage of filament circuit is stepped down by how many volts
6-12
US public law 90-602 states that generators must terminate the exposure at ___ mas for exposures above 50 kvp and ___ mas for exposure below 50 kvp
600, 2000
as the wavelength of x-rays increases, the frequency ___
decreases
when an electron is ejected from the inner orbits of the target atoms, x-rays result. this interaction is called
characteristic
when an electron is attracted towards the nucleus and is deflected from its original direction, the electron is slowed down. the kinetic energy lost is emitted in the form of radiation interaction is called
bremsstrahlung
the filament circuit produces amperage. the can be adjusted by a
line monitor & rheostat
the term used to describe the characteristic interaction of an electron from an outer shell dropping into a "hole" of a vacant shell is called
cascade effect
how much more average photon energy does a 3 phase, 12 pulse generator produce when compared to a 3 phase, 6 pulse
5%
- they are non-uniform in energy and wavelength
- a single electron can cause numerous brems interactions
- the amount of energy lost is determined by how close the electron comes to the nucleus
bremsstrahlung interactions
the purpose of the low voltage circuit in the x-ray machine is to
heat up the filament
conditions necessary for x-ray production
- high speed electron undergoes rapid deceleration
- an outer shell electron drops into the inner shell of an electron
- the electron goes from a higher energy state to a lower energy state
what kvp will produce a tube efficiency of .5% (tungsten z = 74)
68
what is the efficiency of an x-ray tube when using a tungsten target at 60 kvp that is converted into heat?
99.56%
what % of characteristic x-rays are produced at 65 kvp
0%
the minimum reaction time for an ionization chamber is
.001 seconds
the purpose of the rectification bridge is to
change AC to DC
which component of the x-ray tube causes the ther moons to jump the gap at very high speeds?
potential difference
what is the efficiency of an x-ray tube using a bismuth target (z=83) at 80 kvp
66%
for a radiograph produced a 90 KVP, what percent of x-rays produced will be from a bremsstrahlung interaction
90%
what does the 90-30 table mean
The table tilts 90° in an upright position and 30° in the opposite direction
the fixed height for most radiographic tables is
30-40"
the detent lock is placed at what location in the room
40" and 72"
The type of x-ray equipment that may be stationary or mobile and has fluoroscopic capability is called
c-arm
a radiographic table too may be
stationary, floating, motor-driven, radiolucent
the most common cause of tube failure is
arcing
The physical area of the focal spot that is projected out of the table towards the patient is called the
effective focal spot
parts of the cathode assembly include
filament, focusing cup, associated wiring, filament circuit
The thickness of most filament wires is
.1 to .2 mm
an anode has an angle of 40°. How will the effective focal spot up here compared to the effective focal spot of an anode with only an angle of 15°?
larger